Excertos do catálogo

Agilent PXA Signal Analyzer N9030A Data Sheet class C certified Available frequency ranges N9030A-503 3 Hz to 3.6 GHz N9030A-508 3 Hz to 8.4 GHz N9030A-513 3 Hz to 13.6 GHz N9030A-526 3 Hz to 26.5 GHz
Abrir o catálogo na página 1
3 This PXA signal analyzer data sheet is a summary of the complete specifications and conditions. The complete PXA Signal Analyzer Specification Guide can be obtained from the web at: www.agilent.com/find/pxa_specifications Defi nitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to temperature ranges 0 to 55 °C for the solid-state drive (Option SSD), or 5 to 50 ºC for the hard disk drive (standard), unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2 ó) of performance tolerances...
Abrir o catálogo na página 3
4 Frequency and Time Specifi cations Frequency range DC coupled AC coupled Option 503 3 Hz to 3.6 GHz 10 MHz to 3.6 GHz Option 508 3 Hz to 8.4 GHz 10 MHz to 8.4 GHz Option 513 3 Hz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 3 Hz to 26.5 GHz 10 MHz to 26.5 GHz Band LO Multiple (N) DC coupled 0 1 3 Hz to 3.6 GHz 1 1 3.5 to 8.4 GHz 2 2 8.3 to 13.6 GHz 3 2 13.5 to 17.1 GHz 4 4 17 to 26.5 GHz Precision frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate ± 1 x 10–7/year ± 1.5 x 10–7/2 years Temperature stability 20 to...
Abrir o catálogo na página 4
5 Frequency and Time Specifi cations (continued) Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept FFT ± (0.1% x span + horizontal resolution) ± (0.1% x span + horizontal resolution) Sweep time and triggering Range Span = 0 Hz Span 10 Hz 1 ìs to 6000 s 1 ms to 4000 s Accuracy Span 10 Hz, swept Span 10 Hz, FFT Span = 0 Hz ± 0.01% nominal ± 40% nominal ± 0.01% nominal Sweep trigger Free run, line, video, external 1, external 2, RF burst, periodic timer Trigger delay Span = 0 Hz or FFT Span 10 Hz, swept...
Abrir o catálogo na página 5
6 Frequency and Time Specifi cations (continued) Analysis bandwidth1 Maximum bandwidth Standard Option B25 Option B40 Option B1X 10 MHz 25 MHz 40 MHz 140 MHz Video bandwidth (VBW) Range 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz and wide open (labeled 50 MHz) Accuracy ± 6% nominal (in swept mode and zero span) Measurement speed 2 Local measurement and display update rate 10 ms (100/s) nominal Remote measurement and LAN transfer rate 10 ms (100/s) nominal Marker peak search 2.5 ms nominal Center frequency tune and transfer (RF) 43 ms nominal Center frequency tune and transfer (ìW) 69 ms...
Abrir o catálogo na página 6
7 Amplitude Accuracy and Range Specifi cations Amplitude range Measurement range Displayed average noise level (DANL) to maximum safe input level Input attenuator range (3 Hz to 26.5 GHz) 0 to 70 dB in 2 dB steps Electronic attenuator (Option EA3) Frequency range 3 Hz to 3.6 GHz Attenuation range Electronic attenuator range Full attenuation range (mechanical + electronic) 0 to 24 dB, 1 dB steps 0 to 94 dB, 1 dB steps Maximum safe input level Average total power +30 dBm (1 W) With or without preamp Peak pulse power 10 ìs pulse width, 1% duty cycle and input attenuation 30 dB +50 dBm (100 W)...
Abrir o catálogo na página 7
8 Amplitude Accuracy and Range Specifi cations (continued) Frequency response (10 dB input attenuation, 20 to 30 °C, preselector centering applied at 3.6 GHz and above) 95th percentile (~2 ó) 3 Hz to 10 MHz ± 0.46 dB ± 0.19 dB 10 MHz to 3.6 GHz ± 0.35 dB ± 0.16 dB 3.5 to 8.4 GHz ± 1.5 dB ± 0.39 dB 8.3 to 13.6 GHz ± 2.0 dB ± 0.45 dB 13.5 to 22.0 GHz ± 2.0 dB ± 0.62 dB 22.0 to 26.5 GHz ± 2.5 dB ± 0.82 dB Preamp on (Option P03, P08, P13, P26) 9 to 100 kHz ± 0.36 dB attenuation 0 dB 100 kHz to 50 MHz ± 0.68 dB ± 0.26 dB 50 MHz to 3.6 GHz ± 0.55 dB ± 0.28 dB 3.5 to 8.4 GHz ± 2.0 dB ± 0.64 dB 8.3...
Abrir o catálogo na página 8
9 Amplitude Accuracy and Range Specifi cations (continued) Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 1.5 MHz RBW ± 0.03 dB 1.6 MHz to 2.7 MHz RBW ± 0.05 dB 3 MHz RBW ± 0.10 dB 4, 5, 6, 8 MHz RBW ± 0.30 dB Reference level Range Log scale Linear scale –170 to +30 dBm in 0.01 dB steps 707 pV to 7.07 V with 0.11% (0.01 dB) resolution Accuracy 0 dB Display scale switching uncertainty Switching between linear and log 0 dB Log scale/div switching 0 dB Display scale fidelity Between –10 dBm and –18 dBm input mixer level ± 0.10 dB ± 0.04 dB typical below –18 dBm...
Abrir o catálogo na página 9
10 Dynamic Range Specifi cations 1 dB gain compression (two-tone) At 1 kHz RBW with 100 kHz tone spacing, 20 to 30 °C Maximum power at input mixer 20 to 40 MHz –3 dBm 0 dBm typical 40 to 200 MHz +1 dBm +3 dBm typical 200 MHz to 3.6 GHz +3 dBm +5 dBm typical 3.6 to 16 GHz +1 dBm +4 dBm typical 16 to 26.5 GHz –1 dBm +2 dBm typical Preamp on (Option P03, P08, P13, P26) 10 MHz to 3.6 GHz –14 dBm nominal 3.6 to 26.5 GHz Tone spacing 100 kHz to 20 MHz – 28 dBm nominal Tone spacing > 70 MHz – 10 dBm nominal Displayed average noise level (DANL) (Input terminated, sample or average detector,...
Abrir o catálogo na página 10
11 Dynamic Range Specifi cations (continued) DANL with Noise Floor Extension (NFE) on 95th Percentile Improvement for noise-like signals Preamp Off Preamp On Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 8.5 dB 4 dB 7.5 dB 7 dB 6 dB 8.5 dB 7 dB 7 dB 7.5 dB 6 dB Examples of effective DANL Frequency 20 to 30 °C Mid-Band 0 (1.8 GHz) Mid-Band 1 (5.95 GHz) Mid-Band 2 (10.95 GHz) Mid-Band 3 (15.3 GHz) Mid-Band 4 (21.75 GHz) Preamp Off –163 dBm –158 dBm –157 dBm –153 dBm –145 dBm Preamp On –172 dBm –172 dBm –170 dBm –166 dBm –162 dBm Residues, images, and spurious responses Residual responses...
Abrir o catálogo na página 11
12 Dynamic Range Specifi cations (continued) Second harmonic distortion (SHI) Source frequency 10 to 100 MHz 0.1 to 1.8 GHz 1.75 to 2.5 GHz 2.5 to 4 GHz 4 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz Mixer level –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm Distortion* –57 dBc/NA –60 dBc/NA –77 dBc/–95 dBc –77 dBc/–101 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc SHI* +42 dBm/NA +45 dBm/NA +62 dBm/+80 dBm +62 dBm/+86 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm Preamp on (Option P03, P08, P13, P26) Preamp level Distortion SHI 10 MHz to 1.8 GHz 1.8 to 13.25 GHz –45 dBm...
Abrir o catálogo na página 12Todos os catálogos e folhetos técnicos KEYSIGHT TECHNOLOGIES
-
Network Visibility Products
24 Páginas
-
RF Products
20 Páginas
-
Signal Generation Solutions Catalog
27 Páginas
-
PathWave Design and Test Software
23 Páginas
-
Basic Instruments
34 Páginas
-
InfiniiVision 2000 X-Series
31 Páginas
-
X-Series Measurement Applications
24 Páginas
-
Signal Analyzers X-Series
21 Páginas
-
KeysightCare
8 Páginas
-
Automotive Power Testing Solutions
22 Páginas
-
FieldFox and Nemo Handy
4 Páginas
-
Impedance and Network Analysis
42 Páginas
-
E4990A Impedance Analyzer
12 Páginas
-
Keysight N6780 Series
10 Páginas
-
E5071C ENA Network Analyzers
18 Páginas
-
M9485A PXIe Vector Network Analyzer
12 Páginas
-
W2361EP/ET Ptolemy and VTB Engine
2 Páginas
-
Keysight Process Analysis
7 Páginas
-
BenchVue Software
2 Páginas
-
Keysight Process Analysis
7 Páginas
-
PXI RF Switch Modules
2 Páginas
-
I/O Hardware
18 Páginas
-
GPIB and Instrument Control Products
2 Páginas
-
Digital Multimeters
28 Páginas
-
Data Acquisition - DAQ
32 Páginas
-
PCIe Digitizers
8 Páginas
-
2015 Optical Component Test
40 Páginas
-
Power Products Catalog
215 Páginas
-
N9000A CXA Signal Analyzer
2 Páginas
-
The Agilent Remote Link Solution
9 Páginas
-
Agilent 89600 VSA Software
8 Páginas
-
X-Series
23 Páginas
-
CXA X-Series
8 Páginas
-
SOLUTION BROCHURE
6 Páginas
-
Agilent E4990A Impedance Analyzer
12 Páginas
-
Agilent WaferPro Express Software
11 Páginas
-
Agilent InfiniiVision 3000 X-Series
2 Páginas
-
Waveguide Power Sensors
2 Páginas
-
Agilent Calibration Refresh Modules
2 Páginas
-
Power Meters and Power Sensors
34 Páginas
-
E7515A UXM Wireless Test Set
2 Páginas
-
E5063A Network Analyzer
8 Páginas
-
N9320B RF Spectrum Analyzer
12 Páginas
-
Digital Design & Interconnect Standards
16 Páginas
-
N9330B Cable and Antenna Tester
2 Páginas
-
E5052B Signal Source Analyzer
21 Páginas
-
N9038A MXE EMI Receiver
22 Páginas
-
RDX Test Solutions for DigRF
17 Páginas
-
U4611A/B USB 2.0/3.0 Protocol Analyzer
13 Páginas
-
M9252A DigRF Host Adapter
2 Páginas
-
U4431A MIPI? M-PHY Protocol Analyzer
8 Páginas
-
M9502A and M9505A
12 Páginas
-
Infiniium 8000 Series Oscilloscopes
33 Páginas
-
16800 Series Portable Logic Analyzers
34 Páginas
-
Spectrum Analyzers (Signal Analyzers)
17 Páginas
-
Oscilloscopes
16 Páginas
-
U1600 Series Handheld Oscilloscopes
40 Páginas
-
N7700A Photonic Application Suite
12 Páginas
-
Agilent Power Products
31 Páginas
-
LIGHTWAVE CATALOG VOLUME I 2013
40 Páginas
-
Enclosures Solutions Product Catalog
46 Páginas
-
Agilent Test & Measurement Catalog 2011/2012
368 Páginas
-
N4391A Optical Modulation Analyzer
24 Páginas
-
N7744A Optical Multiport Power Meter
6 Páginas
-
E1333A 3-Channel Universal Counter
4 Páginas
-
U2653A 64 Output USB Modular Digital I/O
75 Páginas
-
M9182A Digital Multimeter, 6½ digit, PXI
14 Páginas
-
U4301A PCIe Analyzer
7 Páginas
-
E8257D PSG Analog Signal Generator
28 Páginas
-
34401A Digital Multimeter, 6½ Digit
4 Páginas
-
N5241A PNA-X Microwave Network Analyzer
117 Páginas
Catálogos arquivados
-
Agilent Technologies - Catalog
40 Páginas