Overview
The ZIROX Electrolysis Device SGM5EL utilizes zirconia solid electrolyte technology for the dosing or removal of oxygen from protective or reactive gases. It is designed to produce gases with specific oxygen concentrations or reduction forces.

Components and Functionality
The device comprises a cell module and a control unit. The cell module includes an electrolysis cell and a potentiometric measuring cell, which ensures precise oxygen concentration measurement. An electrical heater maintains the necessary cell temperature. The device can also perform solid electrolyte coulometry, where the cell voltage is kept constant, and the current-time-integral determines the oxygen amount quantitatively.

Applications
  • Production of defined test gases for sensor tests
  • Examination of oxygen permeation in polymers
  • Controlled outgassing of reactors
  • Calibration and monitoring of TA-Systems
  • Examination of decomposition gases in thermal analyses

Technical Specifications
  • Measuring Range: 20.64*104 to 0.5 vol-ppm O2, up to 10-24 vol-ppm possible
  • Carrier Gas: Protective or inert gases (N2, Ar, CO2)
  • Accuracy: Relative error <5%
  • Test Gas Flow: 5 to 10 l/h
  • Max Test Gas Pressure: 100 mbar over pressure
  • Max Test Gas Temperature: 80 °C at gas input
  • Pressure Drop: Approx. 1 kPa at 10 l/h
  • Dimensions: Electronic unit: 320 x 240 x 90 mm; Cell module: 262 x 119 x 119 mm
  • Weight: Electronic unit: 2.5 kg; Cell module: 1.5 kg
  • Gas Connection: 3 mm Swagelok®
  • Protection Degree: Electronic unit: IP 30; Cell module: IP 20
  • Operating Conditions: 10 to 40 °C, max. 80% relative humidity
  • Storing Conditions: -20 to 60 °C, max. 95% relative humidity
  • Interface: RS232
  • Power Supply: 110 to 230 V/50 to 60 Hz
  • Power Consumption: 50 VA
  • Keypad: 6 keys
  • Display: LCD dot-matrix
See more

Catalog excerpts

SGM5EL-1

ZIROX Electrolysis Device SGM5EL Properties The electrolysis device SGM5EL is based on the approved zirconia solid electrolyte technology. It serves the convenient dosing or removal of oxygen from protective gases (N2, Ar) or reactive gas flows (e.g. inert gas N2/H2) either to produce gases with defined oxygen concentration or gases with a requested reduction force. The SGM5EL is a combination of a complex cell module with a control unit. Main parts of the cell module are an electrolysis cell and a potentiometric measuring cell. The potentiometric cell enables the exact measurement of the oxygen concentration adjusted by the electrolysis cell. The required cell temperature is guaranteed by means of an electrical heater. Besides its use as a test gas generator, the SGM5EL can be used for the solid electrolyte coulometry. In this case the cell voltage of the potentiometric cell is kept constant as a set point of a control circuit. The resulting current of the electrolysis cell is proportional to the oxygen transposition with the gas flow. With the current-time-integral the substituted amount of oxygen can be determined quantitatively. Production of defined test gases, e.g. for sensor tests Examination of oxygen permeation of polymers (e.g. foils) Controlled outgassing of reactors Calibration and monitoring of TA-Systems (TG, DTA, TMA, DSC) Examination of the decomposition gases in thermal analyses ZIROX Elektrolysis Device

 Open the catalog to page 1

All ZIROX Sensoren & Elektronik catalogs and technical brochures

  1. SGM7

    2  Pages

  2. SGM5T

    2  Pages

  3. E713

    2  Pages

  4. Vacuum Probe

    2  Pages

  5. E2000

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.