1. Catalogs
  2. Zentro-Elektrik
  3. Laboratory Power Supplies

Laboratory Power Supplies

Laboratory Power Supplies

Laboratory Power Supplies

Product catalog summary
Specifications:
The laboratory power supplies operate with an input voltage of 230 VAC ±10% at 50-60 Hz. The output voltage and current are specified in a table, with line and load regulation for voltage at ≤ 0.01% UA + 1 mV and current at ≤ 0.1% IA + 2 mA. Dynamic response is ≤ 50 µs, and ripple and noise voltage vary based on current levels.

Protection and Safety:
The units include overload and short circuit protection, with optional overvoltage protection. They comply with safety standard EN 61010-1 and have specified isolation levels between input-output, input-ground, and output-ground.

Environmental Conditions:
Operating temperature ranges from 0 to +50°C with cooling via free air convection.

Operation and Control:
Features include programming voltage and current with a 0-10 V reference, master-slave operation, and parallel operation without special action. Remote sensing is available up to 0.25 V per wire. Instruments for voltage and current are LED digital with specified accuracy.

Physical Specifications:
The units are designed for 19" rack installation with options for installed interfaces such as IEEE 488.2 (GPIB)/RS232 and USB1.1. Dimensions and case types are detailed in the document.

Options and Accessories:
Options include a 19" front-panel, 10-turn potentiometer, overvoltage protection, and tracking. External interfaces and analog instruments are available upon request.

Output Specifications:
The document provides a detailed table of output voltage/current combinations, case types, and model numbers, indicating the versatility and customization available for different laboratory needs.
See more

Catalog excerpts

Laboratory Power Supplies-1

Laboratory Power Supplies Linear regulated with high performance Voltage and current regulated All outputs with galvanic isolation Series and parallel operation Master - slave - operation Programming and monitoring current - voltage Intended for 19" rack Option: Installed Interface Euro Card IEEE 488.2 (GPIB)/RS232 with Lab View Driver INT2E series Integrated USB1.1 interface with driver software External CAN Open Interface (on request) Input: Input voltage (AC) Output: Output voltage (DC) Output current Regulation: Line regulation (AC line change ±10%) Load regulation (0 − 100% output load change) Dynamic response Ripple and noise voltage ≤ 0,01 % U A + 1 mV current ≤ ,1% I A + 2 mA 0 voltage ≤ 0,01% U A + 1 mV current ≤ ,1 % I A + 2 mA 0 ≤ 50 µ s Temperature coefficient voltage: I nom ≤ 10 A:0,01% U out + 1 mV pk-pk I nom > 10 A: ,08% U out + 1 mV pk-pk 0 current: ≤ 0,05% I out + 5 mA pk-pk ≤ 0,05% / K Protection: Overload protection Overvoltage protection short circuit protection option Environmental Conditons: Operating temperature Cooling Safety: Safety standard Isolation Input - output: Operation and Control: Programming Master - slave - operation Parallel operation Remote Sense Instruments Adjustment voltage and current (reference - U out) with 0 − 10 V = 0 − 100 % U out, I out and master - slave - operation same units same units without special action consult factory up to 0,25 V per wire for voltage and current LED digital, 3½digits accuracy: V-meter 0,2% ± 1 d A-meter 0,5% ± 1 d voltage and current each 2 potentiometers, coarse/fine Connectors: Input Output Euro - plug 4 mm banana jack, 2 jacks for + and 2 for - , programming: 25 pole sub D plug, rear side Physical Specifications: Dimensions U out ≤ 60 V: 2,3 kVrms U out > 60 V − 150 V: 3,7 kVrms 1,35 kVrms U out ≤ 100 V: 500 Vrms U out > 100 V − 150 V: 820 Vrms EMC: Input EMI filter Input immunity Input - ground: Output - ground: zentro-elektrik GmbH KG | Tel.: ++49 (0) 7232-31310 | Fax: ++49 (0) 7232-3131399 | E-mail: [email protected] Units for Laboratory and Test LD Series Single- and Dual Output

 Open the catalog to page 1
Laboratory Power Supplies-2

Model Number Options: 19" front-panel for cases D and F 10 - turn potentiometer with scale Overvoltage protection (max. 75 V) Tracking Installed IEEE 488.2 Interface Euro-Card INT2E Input cable for external Interface INT2 Analog instruments for voltage and current, 1.5 % of full scale Please order: Last letter of the Model Number: means case type (see following pages) zentro-elektrik GmbH KG | Tel.: ++49 (0) 7232-31310 | Fax: ++49 (0) 7232-3131399 | E-mail: [email protected]

 Open the catalog to page 2
Laboratory Power Supplies-3

Laboratory Power Supplies Units for Laboratory and Test Single output: zentro-elektrik GmbH KG | Tel.: ++49 (0) 7232-31310 | Fax: ++49 (0) 7232-3131399 | E-mail: [email protected]

 Open the catalog to page 3
Laboratory Power Supplies-4

Dual output: zentro-elektrik GmbH KG I Tel.:++49 (0) 7232-31310 I Fax:++49 (0) 7232-3131399 I E-mail: [email protected]

 Open the catalog to page 4

All Zentro-Elektrik catalogs and technical brochures

  1. ZAS Series

    3  Pages

  2. ZPS Series

    2  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.