  3. User Manual for DX2PS Two-dimensional Magnetic Field Probe Station
video corpo

User Manual for DX2PS Two-dimensional Magnetic Field Probe Station

User Manual for DX2PS Two-dimensional Magnetic Field Probe Station
1 / 41 PagesView full catalog

User Manual for DX2PS Two-dimensional Magnetic Field Probe Station

Product catalog summary
Introduction
This document serves as a user manual for the Two-dimensional Magnetic Field Probe Station Model DX2PS by Xiamen Dexing Magnet Tech. Co., Ltd. It provides comprehensive instructions on the operation, installation, and components of the device, which is used for testing the electrical and magnetic properties of various materials and devices.
Specifications
The device can generate an in-plane magnetic field up to 800mT and a vertical magnetic field up to 100mT. It is suitable for high-precision DC/RF measurements, although RF testing requires additional equipment.
Components
  • Probe Station Main Body: Features manual displacement tables for X, Y, Z, T, and θ axes, a two-dimensional three-pole electromagnet, and a ceramic sample table.
  • Probe Holder and Fixture: Allows installation and adjustment of up to four probes.
  • Microscope and Display: Used for observing samples and needle insertion, with images displayed on a 13-inch screen.
  • System Control and Data Acquisition Module: Includes a control computer for real-time control and data reading.
  • Air Flotation Platform: Isolates the device from vibrations (provided by the user).
  • Vacuum Pump: Ensures stable adhesion of the probe holder.
  • Recirculating Cooler: Cools the electromagnet for stable operation.
  • Excitation Power Supply: Includes 1000W and 400W power supplies for horizontal and vertical magnetic fields, respectively.
  • 6221 Current Source and 2182A Nanovoltmeter: Required for operation (provided by the user).
  • Magnetic Field Controller: Manages the magnetic field settings.
Installation Procedures
Installation involves setting up the air flotation platform, placing the probe station, and connecting various components such as the microscope, cooler, and control computer. Detailed cable and pipeline connections are provided for integrating the system.
Operation Method
  • Excitation Power Supply: Operated via front panel buttons.
  • Recirculating Cooler: Set to 18°C with a control accuracy of ±1°C. Ensure coolant levels are maintained.
  • Microscope: Use camera software to display images on the screen.
Recommendations
Users are advised to place power supplies and controllers in separate cabinets to avoid damage and ensure proper operation. Correct connections are crucial to prevent irreversible damage to the electromagnet.
Microscope Adjustment
Manually adjust the lifting and magnification knobs to ensure the microscope is at an appropriate height and magnification for clear image display.
Sample Stage Operation
The sample stage is manually driven. Adjust the X, Y, Z, T, and θ axes to position the sample correctly, avoiding collisions with the electromagnet.
Test Sample Placement
Place the sample on the ceramic table and adjust the displacement tables to align the sample with the microscope's field of view.
Needle Insertion Operation
Position the sample image centrally on the microscope display, adjust magnification, and insert probes sequentially to contact the device under test.
System Startup Process
Sequentially turn on the water chiller, electromagnet power supply, source meter, voltmeter, control computer, and testing program.
Sample Needle Insertion Testing Process
Steps include placing the sample, adjusting the microscope and displacement table, positioning the probe, and starting the test.
System Shutdown Process
Close the testing program, turn off the control computer, source meter, voltmeter, electromagnet power supply, and water chiller.
Microscope Position and Correction
Align the microscope lens with the magnetic field center. If misalignment occurs, recalibrate using the specified steps.
Software Usage
Ensure all instruments are on before software use. The main interface includes a menu bar, display interface, and monitoring interface. Set communication ports and perform tests like resistance and magnetoresistance measurements.
Resistance Measurement
Set parameters such as current and voltage limits. The data display shows the V-I relationship.
Magnetoresistance Measurement
Configure parameters for scanning sequences and sampling points. The data display includes R-H and V-H diagrams.
Pulse Resistance Measurement
Set parameters for pulse width and delay. The test minimizes thermal effects and errors. Data display includes R-I and V-H diagrams.
Variable Field Second Harmonic Measurement
Set parameters for sensitivity and harmonic frequency. The data display includes test and fitting pages.
Equipment and Specifications
The document outlines the equipment required for measuring the second harmonic of variable fields and currents, including the Keithley 6221 current source, Stanford SR830 lock-in amplifiers, magnet power supplies, and a Gaussian meter. It specifies the need for accurate hardware wiring and communication checks before operation.
Measurement Procedures
  • Second Harmonic Measurement: Steps include checking hardware, setting file paths, selecting measurement interfaces, and starting measurements to obtain relationship curves and fitting parameters.
  • Resistance (R-I) Measurement: Involves setting current parameters, file paths, and starting measurements to obtain curve graphs.
  • Magnetoresistance (R-H) Measurement: Similar to R-I measurement but includes magnetic field sequence settings.
  • Pulse Resistance Measurement: Requires switching the current source to pulse mode and setting pulse parameters.
Electromagnet Power Supply Settings
The document details settings for controlling magnetic fields, including parameter settings for in-plane and vertical magnetic fields, error tolerance, stabilization time, and demagnetization procedures.
Configuration File Description
The configuration file for the 2D magnetic field probe station software is automatically loaded and saved, containing instrument addresses and data preservation paths.
Precautions
Includes guidelines for operating temperatures, coolant usage, power supply management, and equipment handling to prevent damage and ensure safety.
Warranty Terms
The product has a one-year warranty covering malfunctions under normal use. Exclusions include improper use, self-repair, and external damage. The company disclaims liability for special, incidental, or indirect damages.
See more

Catalog excerpts

User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-1

User Manual Two-dimensional Magnetic Field Probe Station Model DX2PS Xiamen Dexing Magnet Tech. Co., Ltd. Add: Unit 405, 10th Factory Building of Lianfa, No.34, Yuehua Road, Huli District, Xiamen, China 361006 Tel: (86) 0592 5237772 Mobile Phone: (86) 18030236818 Website: www.dexinmag.com Email: [email protected]

 Open the catalog to page 1
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-2

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. (1) Equipment Introduction 1. To use this device, certain operating procedures must be followed. Please read this text carefully before learning how to use it; 2. This device is a magnetic field probe station; 3. This device can provide an in-plane magnetic field not exceeding 800mT and a vertical magnetic field not exceeding 100mT; 4. This device is mainly used for testing the electrical and magnetic properties of semiconductor materials, micro nano devices, magnetic materials, spintronics devices, and related technical fields. It can provide...

 Open the catalog to page 2
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-3

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. Among them, the two-dimensional triode electromagnet and the X-Y-Z-T -θfive axis adjustable sample stage are fixedly installed in the main structure of the probe station of this equipment. 7、The complete installation picture of this device is as follows: Note: It is recommended that the end users place the magnet power supply, magnetic field controller, computer, etc. in separate cabinets when using, and only place the probe station body on the air flotation platform. (2) Introduction to Each Module a、Probe Station Main Body:  The main structure...

 Open the catalog to page 3
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-4

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd.  X-axis manual displacement table: with a stroke of ± 12.5mm, it can be manually operated to move the sample to be tested and keep it within the range of the microscope and magnetic field;  Y-axis manual displacement table: with a stroke of ± 12.5mm, it can be manually operated to move the sample to be tested and keep it within the range of the microscope and magnetic field;  Z-axis electric displacement table: with a stroke of 0-10mm, it can be manually operated to move the sample to be tested and keep it within the range of the microscope...

 Open the catalog to page 4
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-5

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. The probe is installed on the probe fixture, which is mounted on the probe holder. By adjusting the adjustment knob on the probe holder, the probe tip can be moved to the sample to be tested. This device can simultaneously pin four probes (if more probes are needed, they need to be purchased separately). The diagram is as follows: c、Microscope and its bracket, display screen: The microscope is used to observe the sample to be tested and the needle insertion effect. By connecting the microscope camera and computer through a USB data cable, the...

 Open the catalog to page 5
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-6

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. d、System control and data acquisition module (control computer): The system control and data acquisition module includes a control computer and corresponding control programs. Real time control and data reading of the 6221 and 2182A can be achieved by connecting the control computer, 6221 current source, and 2182A nanovoltmeter through a dedicated USB to GPIB data connection cable; By connecting the control computer and excitation power supply through a dedicated USB to RS232 data connection cable, automatic control and reading of the size and...

 Open the catalog to page 6
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-7

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. The air flotation platform is used to isolate the vibration source and avoid external vibration interference during testing. The diagram is as follows: f、Vacuum Pump: A vacuum pump is used to generate negative pressure, allowing the probe holder to stably adhere to the probe station host. The diagram is as follows: Add: 10th Factory Building of Lianfa, No.34, Yuehua Road, Huli District, Xiamen, China 361006 Tel: 86-592-5237772 Fax: 86-592-5237901 Web: h

 Open the catalog to page 7
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-8

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. g、Recirculating Cooler (Chiller): A Recirculating Cooler (Chiller) is used to cool the electromagnet so that it can be stably excited for a long time. The diagram is as follows: h、Excitation Power Supply for Electromagnet: This device is equipped with one magnet power supply with output power of 400W and one magnet power supply with output power of 1000W. The Diagram is as follows: Add: 10th Factory Building of Lianfa, No.34, Yuehua Road, Huli District, Xiamen, China 361006 Tel: 86-592-5237772 Fax: 86-592-5237901 Web:

 Open the catalog to page 8
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-9

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. i、6221 current source and 2182A nanovoltmeter (provided by the end user): The diagram is as follows: j、Magnetic field Controller: The diagram is as follows: (3) Installation of equipment 1、Installation and leveling of air flotation platform 2、Place the probe station body on the air flotation table 3、Probe holder Installation Add: 10th Factory Building of Lianfa, No.34, Yuehua Road, Huli District, Xiamen, China 361006 Tel: 86-592-5237772 Fax: 86-592-5237901 We

 Open the catalog to page 9
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-10

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. 4、Microscope Installation 5、Recirculating Cooler Installation 6、Control Computer Installation Please refer to the equipment installation video provided by DEXINMAG for the above section. 7、6221 source meter and 2182 voltmeter (purchased by the end user) Installation (4) Connection of equipment cables/pipelines 1、Connection between probe station body and water chiller 1.1、Two pcs of 8mm diameter PU transparent tubes are led out from behind the main body of the probe station, one of which is connected to the "liquid outlet" interface of the water...

 Open the catalog to page 10
User Manual for DX2PS Two-dimensional Magnetic Field Probe Station-11

厦门盈德兴磁电科技有限公司 Xiamen Dexing Magnet Tech. Co., Ltd. 2、Connection between the probe station body and the excitation power supply: 2.1、The in-plane magnet power cord has one end connected to a circular socket at the back of the probe station body, and the other two ends connected to the output port of the in-plane magnet power supply; The diagram is as follows: 2.2. The vertical magnet power cord has one end connected to a circular socket at the back of the probe station body, and the other two ends connected to the output port of the in-plane magnet power supply; The diagram is as follows: Add:...