User’s Manual Model DX-180 Digital Gaussmeter Dexing Magnet Tech. Co.,Ltd. Add: Unit 405,10th Factory Building of Lianfa, No.34, Yuehua Road, Huli District, Xiamen, China 361006 Tel: (86) 0592 5237772 Mobile Phone: (86) 18030236818 Website: www.dexinmag.com Email: [email protected] Methods and apparatus disclosed and described herein have been developed solely on company funds of Dexing Magnet Tech.Co.,Ltd. No government or other contractual support or relationship whatsoever has existed which in any way affects or mitigates proprietary rights of Dexing Magnet Tech.Co.,Ltd. in these developments. Dexing Magnet Tech.Co.,Ltd. reserves the right to add, improve, modify, or withdraw functions, design modifications, or products at any time without notice. Dexing Magnet Tech.Co.,Ltd. shall not be liable for errors contained herein or for incidental or consequential damages in connection with furnishing, performance, or use of this material.
Open the catalog to page 1LIMITED WARRANTY STATEMENT Dexing warrants that this Dexing product (the “Product”) will be free from defects in materials and workmanship for the Warranty Period specified as 12 months from the date of delivery of the Product. If Dexing receives notice of any such defects during the Warranty Period and the Product is shipped freight prepaid, Dexing will, at its option, either repair or replace the Product if it is so defective without charge to the owner for parts, service labor or associated customary return shipping cost. Replacement or repaired parts will be warranted for only the unexpired...
Open the catalog to page 2Dexing DX-180 User’s Manual
Open the catalog to page 3Dexing DX-180 User’s Manual
Open the catalog to page 4This chapter provides an introduction to the Dexing Model 180 Gauss/Tesla meter. The Model DX-180 was designed and manufactured in BeiJing of China by Dexing Magnet Tech.Co.,Ltd. The Model DX-180 is a extremely highly accurate and high resulation intellectualized gaussmeter well suited for field work. It features: Resolution to 5¾ Digits VGA Basic accuracy of measurement: readings×±0.05% ± range×±0.005% A maximum / minimum value/ peak values/ gorge value hold / screen locked Basic resolution: 0.00001mT Data storage (Auto / Manual) / stored data read Auto-zero, automatic, manual selecting range...
Open the catalog to page 5Figure 1-1 DX-180Gauss/Tesla meter Front Panel 1.1 PRODUCT DESCRIPTION The Fully Digital Model 180 Gauss/Tesla meter is well suited to meet these requirements with extremely high accuracy and very high resolution in the magnet testing application. It integrates varieties of the latest high-tech digital technology, with measuring range from DC to the AC magnetic field 50KHZ. it can auto zero, and auto range. Hold Mode is selectable from maximum value, minimum value, and peak and valley values. Instrument can simultaneously measure and display six kinds of parameters. Units of measure is selectable...
Open the catalog to page 6automatic / manual adjust range, threshold setting and alarm-reporting, show temperature and time and brightness self-setting, display unit optional: Gauss/Tesla, A/M or Oe. Automation The DX-180 has a variety of interface features that are compatible with automated test configurations. The RS-232C Serial Interface can perform nearly every function of the instrument front panel. Two analog voltages and an alarm relay facilitate automation without a computer. At the same time, DX-180’s monitoring analog output cases to provide supplementary automation capabilities and an optional remote control...
Open the catalog to page 7DC Accuracy: ±1% of reading ±0. 5% of range Frequency Measurement Range: 2Hz-50 kHz Frequency Dsiplay Range: 2Hz-2.5kHz DC Range: 0-3T (30kG) DC Temp. Coefficient: ±0.03% of reading /°C DC Measurement Resolution: 0. 01G (0.001mT) CAUTION: measuring in the maximal or minimal-scale needs to replace new probe!!! 1.2.4 Front Panel Display Type: 320 * 240 Color LCD Screen Display Units: Gauss (G), Tesla (T), Oster (Oe), Meter/Amps(M/A) Keypad: 14 key membrane Front Panel Features: keypad direct operation, display prompts Figure 1-2 line 3,46 displays the peak Readings RS-232C Capabilities: Baud: 19200,...
Open the catalog to page 8Ambient Temperature: 15 – 35 °C at rated accuracy. 0 °C – 40 °C with reduced accuracy. Power Requirement: 220 VAC (+5%, -10%), 50 or 60 Hz, 20 W Size: W×H×D= 320mm × 110 mm × 287 mm, (12.5 × 4.3 × 11.3 inches) Weight: 4kg Part Number Description Instrument DX-180 the Model DX-180 Gauss/Tesla meter Accessories Included: Model HCHD801F one regular digital slim transverse hall probe with range of 0-10T (0-10kG) 220-10 one 220VAC power wire Man-DX-180 one user manual for DX-180 gauss/tesla meter RS232-DCE9 one 9 pin RS232 direct communication cable Calibration certification one Warranty one Accessories...
Open the catalog to page 9Dexing DX-180 User’s Manual metal shell, measuring fixture, or any place of any instruments that connect to DX-180. The Model DX-180 protects the operator and surrounding area from electric shock or burn, mechanical hazards, excessive temperature, and spread of fire from the instrument. Environmental conditions outside of the conditions below may pose a hazard to the operator and surrounding area. • Maximum Relative Humidity: 80% for temperatures below 31 °C. • Power Supply Voltage Fluctuations: under ±10% of the nominal voltage. To minimize shock hazard, connect instrument chassis and cabinet...
Open the catalog to page 102.0 GENERAL This chapter provides general installation instructions for the Model DX-180 Gauss/Tesla meter. Inspection and unpacking instructions are provided in Paragraph 2.1. Repackaging for shipment instructions are provided in Paragraph 2.2. An definition of rear panel controls is provided in Paragraph 2.3-2.7. Finally, an initial setup and system checkout procedure is provided in Paragraph 2.8. 2.1 INSPECTION AND UNPACKING Inspect shipping containers for external damage, and damage (apparent or concealed) or partial loss of shipment. If damage or loss is apparent, please notify the shipping...