Overview: The flow-captor type 4220 S102 + 4020.46 is a compact and precise flow switch designed for liquid media, featuring analog temperature output and display. It operates on the calorimetric principle, allowing for exact flow set-point adjustments and simultaneous flow rate display.
Specifications:
  • Measuring Range: 0—20 cm/s to 0—300 cm/s, continuously adjustable.
  • Set-point Range: Approximately 15%-90% of the measuring range setting.
  • Temperature Range: Medium: -10 °C to +90 °C; Ambient: -20 °C to +70 °C.
  • Pressure: Maximum 100 bar.
  • Response Time: 2 s - 10 s, depending on range setting.
  • Temperature Drift: Less than 0.3% K.
  • Linearity Deviation: Less than 5%.
  • Repeatability: Less than 2%.
  • Hysteresis: Approximately 10%.
Mechanical Data:
  • Protection Class: Sensor head IP67, electronic IP65.
  • Material: Housing made of Macrolon®; Sensor pipe made of stainless steel WN1.4305 (AISI 303) or WN 1.4571 (AISI 313 Ti).
  • Thread: G ½" BSP or ½" - 14 NPT.
Electrical Data:
  • Operating Voltage: 24 VDC ± 10%.
  • Switching Load: ≤ 5 A, 120 VAC / ≤ 3A, 250 VAC / ≤ 5A, 150 W at VDC.
  • Electrical Output: 2 relays with potential-free single pole double throw contact.
  • Initial Operation: Approximately 10 s after power connection.
Features:
  • Precise switching flow monitor for water-based solutions.
  • High accuracy under low flow conditions.
  • Separate adjustments for "range" and "set-point".
  • Analog display of actual flow and set-point value.
  • LED display for output status.
  • Separate, adjustable temperature output.
  • ISO 9001: 2015 certified.
Temperature Switch-Point Adjustment Procedure:
  1. Turn the temperature potentiometer to the left stop (LED red).
  2. Slowly turn the potentiometer clockwise until the LED switches to green, indicating the switch point.
  3. Example: If the medium temperature is 20 °C, the switch-point will be at 20 °C ± 2.5 °C.
See more

Catalog excerpts

Data sheet flow-captor 4220 S102 + 4020.46-1

raptor flow-captor 4220 SI02 + 4020.46 The flow-captor type 4220 S102 + 4020.46 is a family of compact, precise measurement flow switches with analog temperature output and analog display. They operate based on the calorimetric principle. The flow-captor allows to set an exact flow set-point and will display simultaneously the flow rate up to the lowest flow conditions. The 4020.46 has a separate, adjustable temperature output. • precise switching flow monitor for water based solutions • high accuracy also under low flow conditions • separate adjustment for "range"and "set-point" • analog display of actual flow and display of them adjusted set-point value • LED display for output status • separate, adjustable temperature output • ISO 9001: 2015 weber Sensors GmbH ■ Strohdeich 32 ■ D-25377 Kollmar ■ Tel.: +49 4128-591 ■ Fax: -593 eMail: [email protected] Technical data subject to alteration! • Rev. AC /

 Open the catalog to page 1
Data sheet flow-captor 4220 S102 + 4020.46-2

Flow monitor for liquid media flow-captor 4220 S102 + 4020.46 Connection diagram: -O sn. o. contact -o switching contact -o break contact -O sn. o. contact Temperature switch-point adjustment (sensor in contact with medium) left Stopp. (LED red) 2. Turn temp.-potentiometer slowly clockwise until the LED switches to green (Switch point). For example: If the medium temperature is 20 °C the switch-point will be at 20 °C ± 2,5 °C. weber Sensors GmbH ■ Strohdeich 32 ■ D-25377 Kollmar ■ Tel.: +49 4128-591 ■ Fax: -593 eMail: [email protected] Technical data subject to alteration! • Rev. AB / 23.03.17...

 Open the catalog to page 2

All Weber Sensors GmbH catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.