video corpo

WPH1000

WPH1000

Product catalog summary
Overview: The WPH1000 Field Probe is designed for measuring electromagnetic fields in the frequency range of 30 MHz to 1000 MHz. It is suitable for various applications including defense, labs/R&D, industry, and telecommunications.
Technical Specifications:
  • Frequency Range: 30 MHz - 1000 MHz
  • Sensor Type: Isotropic, RMS Diode Technology
  • Measurement Range: 0.020 - 20 A/m
  • Damage Level: Continuous Wave (CW) 35 A/m, Peak 350 A/m for 1 μs
  • Dynamic Range: 60 dB
  • Sensitivity: 0.020 A/m
  • Frequency Response: +0.5 / -1 dB (40 MHz - 700 MHz), +0.5 / -3 dB (30 MHz - 1000 MHz)
  • Linearity: ± 1 dB (0.04 to 4 A/m)
  • Axial Isotropy: ± 1 dB
  • Calibration Period: 24 months (recommended)
  • Temperature Range: -10 ºC to 50 ºC
  • Dimensions: 28.4 cm x 6 cm Ø
  • Weight: 95 g
Applications: The probe is used for assessing military sites, protecting R&D and lab personnel from RF exposure, evaluating industrial processes for worker safety, and certifying telecommunication services.
Compatibility: Compatible with SMP2 and MonitEM devices. Frequency response can be corrected using SMP2 with stored correction factors.
Note: Product specifications and descriptions are subject to change without notice.
See more

Catalog excerpts

WPH1000-1

WPH1000 Field Probe 30 MHz - 1000 MHz • High sensitivity from 0.020 A/m • Isotropic and RMS measurement • Measurements in accordance with International Standards Telecommunications: certification and audit f telecommunication services. o Industry: assessment of industrial processes for workers’ exposure protection. Defence: assessment of military sites and personnel exposure protection. Labs/R&D: RF exposure protection o f R&D and labs personnel. Technical Specifications Frequency range Sensor type Isotropic, RMS Diode Technology Type of frequency response Measurement range Dynamic range Frequency response (*) Axial isotropy Calibration period Temperature range (*) The frequency response can be corrected with the SMP2 by using the correction factors stored in the probe. Compatible with SMP2, MonitEM Product specifications and descriptions in this document subject to change without notice Safety, Quality, Service

 Open the catalog to page 1

All WAVECONTROL catalogs and technical brochures

  1. MonitEM-Lab

    2  Pages

  2. WPH-DC Probe

    1  Page

  3. WP Wifi

    1  Page

  4. WPT

    1  Page

  5. WPH60

    1  Page

  6. WPF40

    1  Page

  7. WPF18

    1  Page

  8. WPF8

    1  Page

  9. WPF6

    1  Page

  10. WPF3

    1  Page

  11. WP400

    2  Pages

  12. WP400-3

    2  Pages

  13. WP50

    1  Page

  14. SMP2

    2  Pages

  15. MapEM

    2  Pages

  16. MonitEM

    2  Pages

  17. cMonitEM

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.