1. Catalogs
  2. Vitrek
  3. V7X SERIES

V7X SERIES

V7X SERIES

V7X SERIES

Product catalog summary
Overview
The V7X Series Hipot Testers are advanced electrical safety testing devices designed to meet stringent UL, CSA, TUV, and IEC requirements. These testers are known for their speed, accuracy, user safety, and reliability, offering a high price/performance ratio. They are manufactured in the USA and come with a two-year warranty.

Key Features
  • Color touch LCD and high-speed DSP technology for enhanced user interface and performance.
  • Available in six models offering AC and DC Hipot up to 5KV, leakage current measurement to 100 nano-amps, insulation resistance to 450GΩ, and ground bond to 30 amps.
  • Interfaces include USB, RS232, and Digital I/O.
  • Compact, lightweight, and fan-free design for ultra-reliability and efficiency.

Specifications
  • AC Hipot: Output voltage from 10 to 5000V RMS, with a maximum load current of 20mArms.
  • DC Hipot: Output voltage from 20 to 5000V, with a maximum load current of 10mA.
  • Insulation Resistance: Test voltage from 20 to 5000VDC, with a maximum IR of 450Gigohm.
  • Ground Bond: Test current from 1 to 30Arms, with a compliance voltage of >4.5Vrms.

Performance and Safety
  • Fast test times with a minimum of 100mS.
  • Safety features include a 150µS safety shutdown and selectable ARC detection from 1-20mA.
  • CE Safety Mark certified to EN61010.

Additional Features
  • Test memory stores up to 999 steps and 60 test sequences.
  • Internal self-test and pre-programmed daily verification test.
  • Optional rear panel terminals and high voltage switching capabilities.

Ordering Information
  • Models range from V70 to V79, each offering different combinations of testing capabilities.
  • Accessories include test leads, software, and additional testing equipment.

Conclusion
The Vitrek V7X Series Hipot Testers offer state-of-the-art performance in a compact and cost-effective package, making them a reliable choice for production line testing.
See more

Catalog excerpts

V7X SERIES-1

V7X Series Hipot Testers Electrical safety testing will never be the same. With color touch LCD and high speed DSP technology, the compact and rugged V7X sets the standard for price/performance ratio. Made in the USA to meet tough UL, CSA, TUV and IEC Hipot requirements—the V7X provides unbeatable speed, accuracy, user safety and reliability. Choose from six low cost models offering AC and DC Hipot to 5KV, leakage current measurement to 100 nano-amps, Insulation Resistance to 450GΩ, Ground Bond to 30 amps and built-in switching. Combine all that with USB, RS232 and Digital I/O interfaces, plus a two year warranty. The V7X is simply unbeatable.

 Open the catalog to page 1
V7X SERIES-2

Vitrek V7X Series – No other Hipot can touch ‘em Vitrek V7X Series Need Multi-point Hipot? For nearly 25 years, an obsessions has driven us to keep improving how hipot testers perform—speed, accuracy, user safety and reliability have been our driving force. Building a better hipot tester is in our DNA— it drove us to successfully engineer the first 4-in-1 safety tester back in 1990 and it has driven us to redefine performance in low cost hipot testers today. For high voltage cable and component testing—choose the V75 with 16 channels of built-in high voltage switching. With the versatile V75,...

 Open the catalog to page 2
V7X SERIES-3

Vitrek V7X Series Performance Specifications AC Hipot Output Voltage 10 to 5000V RMS, 50/60 Hz (2000V max on V75) Accuracy : 1% of setting +5V, No load to full load Resolution : 1V at all levels Max load current : 20mArms Leakage Current Accuracy : 1% of reading +5uA Resolution : 1uA DC Hipot Output Voltage 20 to 5000V (3000V max on V75) Accuracy : 1% of setting+ 5V, No load to full load Resolution : 1V at all levels Max load current : 10mA Leakage Current Accuracy : 1% of reading +1uA Resolution : 0.1uA IR - Insulation Resistance Test Voltage 20 to 5000VDC (3000V max on V75) Accuracy : 2.5%...

 Open the catalog to page 3
V7X SERIES-4

Vitrek V7X Series Performance Specifications (continued) Shutdown Breakdown : within 150us HV Safety : within 1ms User Stop or Interlock opened : within 2ms Resistance/Current Limit : within 100ms Breakdown For Hipot tests, automatically checks for sudden uncontrolled increases of load current throughout test, no min/max leakage required. Arc Detect For Hipot tests, settable to none or adjustable level between 1 and 30mApk, 5MHz bandwidth. Pause Step A timed pause of defined length between 0.1s and 999.9s. Hold Step A user continued hold step with a two line message to be displayed to the user...

 Open the catalog to page 4

All Vitrek catalogs and technical brochures

  1. Vitrek Catalog

    48  Pages

  2. 4700

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.