TC 205

TC 205

TC 205

Product catalog summary
Specifications:
The Laboratory Air Circulated Refrigerator TC 205 is designed in accordance with DIN 13277 standards. It features external dimensions of 750mm (W) x 820mm (D) x 2100mm (H) and internal dimensions of 600mm (W) x 670mm (D) x 1335mm (H). The refrigerator has a capacity of 700 liters and operates within a temperature range of 2°C to 20°C, with an optional range of -5°C to 20°C. The housing is made of galvanized sheet steel with a high-quality white coating, and the interior is designed for easy cleaning and high hygiene standards.

Control System:
The refrigerator is equipped with a touchscreen temperature controller (TC 2015) that allows for precise temperature control with an accuracy of +/- 1.8°C. It includes features such as a PT1000 temperature sensor, digital temperature setting, password protection, and data import/export via USB or Ethernet. The system provides both optical and audible alarms for various conditions, including high/low temperature, door open, power failure, and more.

Refrigerating Unit:
The unit is fully hermetically sealed and features a low-noise, energy-saving compressor. It uses environment-friendly refrigerant R 290 or equivalent. The defrost process is automatic, with time and thermal monitoring.

Electrical Data:
The refrigerator operates on a power supply of 230 V/50 Hz with a power input of 0.25 kW. It includes a 2.0 m power cable with a Schuko plug and a 16 A fuse.

Special Equipment and Accessories:
Optional accessories include a GSM module for alarm notifications, a round chart recorder for temperature logging, electronic door locks, and a wireless data logger for independent temperature recording.

Qualifications:
The refrigerator undergoes various qualification processes, including Design Qualification (DQ), Installation Qualification (IQ), Operational Qualification (OQ), Calibration Qualification (CQ), and Performance Qualification (PQ), all in accordance with DIN 13277:2022-05 standards. These processes ensure that the equipment meets quality and regulatory requirements and performs reliably under specified conditions.
See more

Catalog excerpts

TC 205-1

Laboratory Air circulated Refrigerator also in accordance with DIN 13277 External Dimensions: Internal Dimensions: W = 600 mm (clear door width 540 mm) D = 670 mm H = 1335 mm (lower edge evaporator) Temperature range: Optional: Housing (high size) galvanized sheet steel with high quality white coating, antibacterial coated. With 4 feet, height adjustable. Optional: of stainless steel (Backplane, as well as top and bottom side of the cabinet of galvanised sheet steel) Set of 4 castors, 2 lockable Interior space Inner case of galvanized sheet steel with high quality white coating, antibacterial coated. Cleaning friendly by rounded corners, slippery surface in the interior, meets highest hygiene requirements Optional: of stainless steel Insulation high density foamed-in-place polyurethane, with a thickness of 75 mm. CFC-free Circulation air cooling for rapid and uniform temperature control, reducing temperature gradients to a minimum. When the door is "open" the circulating fan is switched off automatically by a micro switch in order to prevent warmer ambient air being drawn in. Door Solid door, supplied as standard with right hand hinge, also available with left hand hinge at no extra cost. Grease resistant magnetic seal. Optional: with door lock, or electronic door lock with glass door Interior fittings 4 pcs. Grating-type shelves (dim.: 530x650 mm), white coated (max. 13 shelves possible) Optional: - Stainless Steel shelves or - Stainless Steel drawer compartments, with 3 lengthwise dividers (max. 13 drawers possible) with telescopic-type guides on rollers with stops, drawer dimensions: 500x575x77 mm (WxDxH) Usable width between the lengthwise dividers: 108,5 mm (lengthwise dividers easily removable) - Interior lighting with automatic ON/OFF micro switch tritec ® Gesellschaft für Labortechnik und Umweltsimulation mbH Hüttenstraße 9 D-30165 Hannover Homepage www.tritec-klima.com E-mail [email protected] Phone ++49-511/3523508 Fax ++49-511/3521715 technische Änderungen vorbehalten

 Open the catalog to page 1
TC 205-2

Laboratory Air circulated Refrigerator also in accordance with DIN 13277 Control system located above the door Touchscreen temperature controller, TC 2015 Working range from 2°C to 20°C Optional: -5°C to 20°C (Provided that the ambient temperature is around 4°C above the set point) Temperature accuracy +/- 1,8°C after stabilising (with 50% filling) Language in German / English adjustable Touchscreen Display, size 12,5 cm (5“) PT1000 temperature sensor for better temperature control accuracy Digital temperature setting and display with an accuracy of 0.1°C Power supply of 100-240V, 50-60Hz Password...

 Open the catalog to page 2
TC 205-3

Laboratory Air circulated Refrigerator also in accordance with DIN 13277 • Network Connection 2 Ethernet ports 10/100 Mbps available. This interface can be accessed directly on the log data of the controller to import in the software. NO/NC contact Alarm signal time delayed (adjustable) Language: English / German, other languages on request Maintenance message: Notification that maintenance should be carried out. Test facilities for Alarm limits and battery charge status In the housing of the temperature regulator there are all relevant components for using the cabinet. Plug-in connectors of...

 Open the catalog to page 3
TC 205-4

Laboratory Air circulated Refrigerator also in accordance with DIN 13277 Special Equipment and Accessories: GSM Modul Connecting to the potential-free output. In case of an alarm either a message or a call will be sent automatically. Archiving of 1000 phone numbers is possible. The GSM module is equipped with a rechargeable battery. Automatic alert via SMS when the credit has been used on the SIM card. 6 units can be connected per module. The SIM card is not included Round Chart recorder to record temperature, permanently installed in control panel. Comes as standard with battery back-up power...

 Open the catalog to page 4
TC 205-5

Laboratory Air circulated Refrigerator also in accordance with DIN 13277 Qualifications DQ (Design Qualification) Definition: Documented proof that the quality-related, GMP-related requirements has been adequately addressed in the design of equipment, including buildings, premises and auxiliary equipment The user-requirement profiles (specifications) are documented and confirmed by us. On request, a specification can be created by us. IQ (Installation Qualification) Definition: Documented proof that critical equipment and systems have been delivered and installed in accordance with the set requirements...

 Open the catalog to page 5

All Tritec GmbH catalogs and technical brochures

  1. TC 104

    5  Pages

  2. KB 7070 LL

    3  Pages

  3. KB 9400 F

    4  Pages

  4. TC 704

    4  Pages

  5. TC 703

    4  Pages

  6. TC 702

    4  Pages

  7. TC 701

    4  Pages

  8. KB 9207

    4  Pages

  9. KB 9206

    5  Pages

  10. KB 9205

    4  Pages

  11. KB 9204

    4  Pages

  12. KB 9203

    4  Pages

  13. KB 9201

    4  Pages

  14. KB 9182

    4  Pages

  15. KB 9200

    4  Pages

  16. KB 7140 LL

    3  Pages

  17. KB 7070 F-LL

    3  Pages

  18. TC 409

    2  Pages

  19. TC 408

    2  Pages

  20. TC 407

    2  Pages

  21. TC 406

    2  Pages

  22. TC 405

    2  Pages

  23. TC 404

    2  Pages

  24. TC 403

    2  Pages

  25. TC 401

    2  Pages

  26. TC 402

    2  Pages

  27. KB 7140 F-LL

    4  Pages

  28. TS 9430

    4  Pages

  29. TS 9135

    4  Pages

  30. TS 9053

    4  Pages

  31. TS 9026

    4  Pages

  32. B 9420

    5  Pages

  33. B 9130

    5  Pages

  34. B 9051

    5  Pages

  35. B 9025

    5  Pages

  36. TC 1203

    5  Pages

  37. TC 1004-ex

    3  Pages

  38. TC 1207

    5  Pages

  39. TC 1000-TH

    5  Pages

  40. TC 1000-AX

    5  Pages

  41. TC 1206

    5  Pages

  42. TC 1205

    5  Pages

  43. TC 1204

    5  Pages

  44. TC 1202

    5  Pages

  45. TC 605

    6  Pages

  46. TC 604

    6  Pages

  47. TC 603

    6  Pages

  48. TC 602

    6  Pages

  49. TC 601

    6  Pages

  50. TC 506

    5  Pages

  51. TC 511

    5  Pages

  52. TC 505

    5  Pages

  53. TC 510

    5  Pages

  54. TC 504

    5  Pages

  55. TC 560-G

    5  Pages

  56. TC 560

    5  Pages

  57. TC 550-G

    5  Pages

  58. TC 550

    5  Pages

  59. TC 509

    5  Pages

  60. TC 503

    5  Pages

  61. TC 508

    5  Pages

  62. TC 502

    5  Pages

  63. TC 507

    5  Pages

  64. TC 501

    5  Pages

  65. TC 500-V

    5  Pages

  66. TC 500-H

    5  Pages

  67. TC 1014-ex

    5  Pages

  68. TC 208

    5  Pages

  69. TC 1013-ex

    5  Pages

  70. TC 207

    5  Pages

  71. TC 1012-ex

    5  Pages

  72. TC 206

    5  Pages

  73. TC 1011-ex

    5  Pages

  74. TC 204

    5  Pages

  75. TC 1010-ex

    5  Pages

  76. TC 203

    5  Pages

  77. TC 1009-ex

    5  Pages

  78. TC 202

    5  Pages

  79. TC 201

    5  Pages

  80. TC 1001-ex

    3  Pages

  81. TC 200

    6  Pages

  82. TC 114-2

    6  Pages

  83. TC 107

    5  Pages

  84. TC 113-2

    6  Pages

  85. TC 106

    5  Pages

  86. TC 120

    6  Pages

  87. TC 112-2

    6  Pages

  88. TC 105

    5  Pages

  89. TC 111-2

    6  Pages

  90. TC 103

    5  Pages

  91. TC 108-2

    6  Pages

  92. TC 101

    6  Pages

  93. TC 100-V

    6  Pages

  94. TC 100-H

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.