Overview
The ImageMaster® HR is a highly accurate and flexible MTF test station designed for medium-sized sample lenses. It is suitable for R&D, quality assurance, and production environments. Its modular design allows for future upgrades and measurements of various optical parameters.

Key Features
  • Vertical setup for space efficiency and accurate lens positioning.
  • Measures ultra-wide field angles up to ±105° with a precise swinging arm.
  • Upgradeable for finite testing with additional motorized components.
  • High-quality components ensure precise MTF testing.

Measurement Parameters
  • MTF on-axis and off-axis
  • Effective Focal Length (EFL)
  • Distortion
  • Field Curvature
  • Chromatic Aberrations (Lateral and Longitudinal)
  • Astigmatism
  • Relative Transmission and Illumination
  • Field of View
  • Chief Ray Angle
  • Depth of Focus
  • F-Number
  • Relative Flange Focal Length
  • Veiling Glare

Options and Upgrades
  • Motorized sample holder and reticle/filter changer.
  • Finite conjugate stage with object generator.
  • Various collimators and spectral range extensions (NUV, NIR).
  • Additional filters and reticles.

Software
The MTF-Lab software package offers advanced functions for scanning and image positioning. It includes a script language for stage control and data manipulation, with data export capabilities in various formats.

Specifications
  • Optical setup: Infinity conjugates (finite optional)
  • Max. off-axis angle: ±105°
  • Spectral range: 450-750 nm (NIR/NUV optional)
  • Azimuth range: 360°
  • Max. image height: ±46 mm
  • Spatial frequency: 0-500 lp/mm (up to 1000 lp/mm depending on sample)

Accuracy and Repeatability
  • MTF: Accuracy 2%, Repeatability 0.5%
  • Effective Focal Length: Accuracy ±5 µm to ±0.2%
  • Distortion: Accuracy ±0.7% / ±0.25%, Repeatability ±0.3% / ±0.1%
  • Chromatic Aberrations: Lateral ±0.5 µm, Longitudinal 1 µm
  • Chief Ray Angle: Accuracy ±1°, Repeatability ±0.4°
  • Flange Focal Length: Accuracy ±2 µm, Repeatability ±1 µm
  • Astigmatism: Accuracy ±4 µm, Repeatability ±1 µm
  • Field Curvature: Accuracy ±2 µm, Repeatability ±1 µm
  • Relative Illumination: Accuracy ±3%, Repeatability ±2%
See more

Catalog excerpts

ImageMaster-HR-2016-1

Ultra-accurate, Multi-functional

 Open the catalog to page 1
ImageMaster-HR-2016-2

ImageMaster ® HR Outstanding Level of Accuracy and Flexibility The ImageMaster® HR is a fully equipped R&D quality test station for medium sized sample lenses. Its modular and upgradeable design enables the measurement of the image quality (MTF) and a wide range of other optical parameters for today and future needs. The instrument is used in the R&D laboratory as well as in the quality assurance or in production. The unique vertical setup of the ImageMaster® HR is space saving and ensures the most convenient and accurate positioning of the sample lens mounts. For the majority of lenses this...

 Open the catalog to page 2
ImageMaster-HR-2016-3

TRIOPTICS has developed a new forward-look- ing software package called MTF-Lab. Sever- al useful functions are integrated which help the user in scanning and perceiving the cor- rect image position of the sample under test, Changing the measurement mode is easy and time-saving. All important measurements for the imaging properties have been revised and are quickly accessible. The script lan- guage for advanced measurements was im- proved and extended including a simple pro- gramming function for stage controlling, data manipulation, loops and variables. The data export is possible to a variety...

 Open the catalog to page 3
ImageMaster-HR-2016-4

Specification ImageMaster® HR Parameter Infinity conjugates (finite optional) Spectral range 450…750 nm, halogen (NIR/NUV optional) Azimuth range Spatial frequency (in specification) Max. spatial frequency 0 … 500 lp /mm 1000 lp/mm (depending on sample) MTF on-axis and off-axis Accuracy Repeatability Effective Focal Length Accuracy Distortion Accuracy (without / with encoder) Repeatability (without / with encoder) Lateral chromatic aberration Accuracy Repeatability Longitudinal chromatic aberration Accuracy Repeatability Chief Ray Angle Accuracy Repeatability Flange Focal Length (relative) Accuracy...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.