Catalog excerpts
C-Series Replacement Syringes Materials: UHMWPE Seal, Borosilicate Glass Fitting Interface: 1/4 - 28 Stroke: 3 cm *Compatible only with the TriContinent C3000 Model #8657-15 and C24000 Model #8657-11. Materials: PTFE seal, Borosilicate Glass Fitting Interface: 1/4 -28 Stroke: 3 cm
Open the catalog to page 1SYRINGE REPLACEMENT To replace the syringe, prime all fluid from the syringe, and follow the steps below: 1. To replace the syringe, lower the plunger driver. = For the C3000 pump, if the power is on, this can be done by sending the A3000R command or if the pump is powered off, manually lower the plunger by pushing firmly down on the syringe plunger holder. C24000 pump, with the power on, send the A24000R command. You cannot manually lower the plunger on the C24000 pump. 2. Turn the plunger lock screw counterclockwise 3 turns. 3. Turn the syringe counterclockwise and remove from the valve....