Overview: The Trek Model PD04002A is a microprocessor-based charged plate monitor designed for monitoring air ionizers in environments where static charge can affect production, such as semiconductor and electronic assembly processes. It features high sensitivity, requiring minimal ion field current for accuracy and stability.
Specifications:
  • Charged Plate Voltage Range: 0 to ±55 V DC or peak AC.
  • Measurement Accuracy: 0.2% of full scale.
  • Operating Temperature Range: -60° to 160° C.
  • Relative Humidity: Up to 85%, noncondensing.
  • Dimensions: 122 mm W x 43 mm H x 153 mm D.
  • Weight: 0.3 kg (0.6 lb).
Features:
  • Utilizes standard, custom, and high-temperature charged plates.
  • Remote monitoring capability through a rear panel connection.
  • Digital command signals for remote initialization of balance and discharge tests.
  • Voltage Monitor Output: 0 to +5 V DC represents a charged plate voltage of -55 V to +55 V.
  • Compact design with exceptional accuracy and stability.
Performance:
  • Large Signal Bandwidth: DC to greater than 200 Hz.
  • Small Signal Bandwidth: DC to 2.5 kHz.
  • Self-Discharge Rate: Less than 2 V per minute at 55 V.
Installation and Safety:
  • Mounting: Four M4 screws for the monitor, one M4 screw for the charged plate.
  • Power Supply: +24 V ±10% DC at 0.21 amps.
  • Connections: Phoenix Contact Connector for power and digital commands.
  • Caution: Designed for electrostatic voltage measurements only; not for "hard" voltage sources.
Certification: The Model PD04002A is tested and calibrated to specifications traceable to the National Institute of Standards and Technology or consensus standards. It is CE compliant.
See more

Catalog excerpts

PD04002A-1

Trek’s PD04002A versatile microprocessor based charged plate monitor is ideally suited to monitor the performance of air ionizers that are used in critical wide temperature operations associated with semiconductor, LCD, electronic assembly, and other processes where static charge accumulations pose a threat to production yields and/or product quality. • Operating Range -60° to 160° C • Relative Humidity To 85%, noncondensing • Voltage Monitor Output for Remote Monitoring • Exceptional Accuracy and Stability • Compact Design • C€ Compliant Model PD04002ACharged Plate Monitor Controller • Utilizes Standard, Custom and Wide-Temperature Range Charged Plates • High-Temperature Charged Plate (Model PD04002AP) 25 mm x 25 mm (1" x 1") Cable Length: 3 meters Capacitance: 20 pF ±5 pF The Model PD04002A utilizes a new technique which requires less than 0.1 picoampere of ion field current to achieve full accuracy and stability. Other product designs require up to 2000 times more ion current to be drawn from the ion field to achieve stability. The ultra sensitivity of the Model PD04002A makes it the ideal choice in air ionizer monitor applications where small area ion collecting sensors are used to meet the requirement for high accuracy ion balance measurement to levels of less than one volt. The voltage of the charged plate is initialized prior to discharge time testing. Balance tests, positive discharge tests, and negative discharge tests can be initialized from a remote location through digital command signals. The charged plate voltage can be remotely monitored through a rear panel connection. The voltage output at the Output “HI" terminal in reference to the Output “LO" terminal (ground) is a proportional buffered representation of the charged plate voltage at a scale factor of 1 to 22 (other scale factors are available). The Model PD04002A wide temperature range charged plate monitor functions with standard, custom, or high temperature charged-plates. The monitor and high-temperature charged-plate can be used on a table top or securely mounted byfastenerstothe bottom of the unit. Measurement and Power Solutions™ www.trekinc.com

 Open the catalog to page 1
PD04002A-2

Performance Charged Plate Voltage Range 0 to ±55 V DC or peak AC, nominal. Charged Plate Voltage Measurement Accuracy 0.2% of full scale. Large Signal Bandwidth (-3 dB) DC to greater than 200 Hz. Small Signal Bandwidth (-3 dB) DC to 2.5 kHz. Charged Plate Self-Discharge Rate (no incident ion flow) Less than 2 V per minute at 55 V for relative humidity up to 85%. Charged Plate Capacitance 20 pF ± 4 pF. (The capacitance is independent of charged plate connecting cable length.) Features Voltage Monitor Output V 0 to +5 V DC represents a charged plate voltage of -55 V to +55 V (other scale factors...

 Open the catalog to page 2

All TREK, INC. catalogs and technical brochures

  1. 158A

    2  Pages

  2. 157

    2  Pages

  3. 156A

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.