PD04002A - 1

Trek’s PD04002A versatile microprocessor based charged plate monitor is ideally suited to monitor the performance of air ionizers that are used in critical wide temperature operations associated with semiconductor, LCD, electronic assembly, and other processes where static charge accumulations pose a threat to production yields and/or product quality. • Operating Range -60° to 160° C • Relative Humidity To 85%, noncondensing • Voltage Monitor Output for Remote Monitoring • Exceptional Accuracy and Stability • Compact Design • C€ Compliant Model PD04002ACharged Plate Monitor Controller • Utilizes Standard, Custom and Wide-Temperature Range Charged Plates • High-Temperature Charged Plate (Model PD04002AP) 25 mm x 25 mm (1" x 1") Cable Length: 3 meters Capacitance: 20 pF ±5 pF The Model PD04002A utilizes a new technique which requires less than 0.1 picoampere of ion field current to achieve full accuracy and stability. Other product designs require up to 2000 times more ion current to be drawn from the ion field to achieve stability. The ultra sensitivity of the Model PD04002A makes it the ideal choice in air ionizer monitor applications where small area ion collecting sensors are used to meet the requirement for high accuracy ion balance measurement to levels of less than one volt. The voltage of the charged plate is initialized prior to discharge time testing. Balance tests, positive discharge tests, and negative discharge tests can be initialized from a remote location through digital command signals. The charged plate voltage can be remotely monitored through a rear panel connection. The voltage output at the Output “HI" terminal in reference to the Output “LO" terminal (ground) is a proportional buffered representation of the charged plate voltage at a scale factor of 1 to 22 (other scale factors are available). The Model PD04002A wide temperature range charged plate monitor functions with standard, custom, or high temperature charged-plates. The monitor and high-temperature charged-plate can be used on a table top or securely mounted byfastenerstothe bottom of the unit. Measurement and Power Solutions™ www.trekinc.com

Open the catalog to page 1
PD04002A - 2

Performance Charged Plate Voltage Range 0 to ±55 V DC or peak AC, nominal. Charged Plate Voltage Measurement Accuracy 0.2% of full scale. Large Signal Bandwidth (-3 dB) DC to greater than 200 Hz. Small Signal Bandwidth (-3 dB) DC to 2.5 kHz. Charged Plate Self-Discharge Rate (no incident ion flow) Less than 2 V per minute at 55 V for relative humidity up to 85%. Charged Plate Capacitance 20 pF ± 4 pF. (The capacitance is independent of charged plate connecting cable length.) Features Voltage Monitor Output V 0 to +5 V DC represents a charged plate voltage of -55 V to +55 V (other...

Open the catalog to page 2

All TREK catalogs and technical brochures

  1. 158A

    2 Pages

  2. 157

    2 Pages

  3. 156A

    2 Pages