MODEL 9041 EXTENDED SPECIFICATIONS

MODEL 9041 EXTENDED SPECIFICATIONS
1 / 23 PagesView full catalog

MODEL 9041 EXTENDED SPECIFICATIONS

Product catalog summary
General Specifications:
The 9041A Transportable Calibrator requires a warm-up time of up to 20 minutes and features standard USB interfaces with optional RS232 interfaces. It operates within a temperature range of 0°C to 50°C and can be stored between -5°C and +60°C. The device is suitable for altitudes up to 3000m and can be transported up to 12000m. It must be earthed for EMC and safety compliance, supports line voltages of 110V or 230V, and consumes a maximum of 200 Watts. It includes safety sockets for connections and a backlit display with a resolution of 240 x 64 dots.
DC Voltage (DCV) Specifications:
The calibrator provides a range of DC voltage outputs with specific accuracy and stability. It includes safety features like high voltage ramping and automatic standby for voltages over 20V, with a maximum floating voltage of 100V. Specifications are valid within a temperature range of TCal ±5°C.
DC Current (DCI) Specifications:
The DCI specifications highlight the accuracy and stability of current outputs across different ranges, with overload protection and a microprocessor to prevent overheating, especially in the 30A range. Specifications are valid at TCal ±5°C, with adjustments needed outside this range.
AC Voltage (ACV) Specifications:
The ACV section details the accuracy and stability of AC voltage outputs across various frequencies, including high voltage safety features and an external control switch option. Specifications apply from 10% of full scale, with a maximum floating voltage of 100V.
AC Current (ACI) Specifications:
The ACI specifications cover the accuracy and stability of AC current outputs, capable of handling inductive loads up to 1H without additional protection. Specifications apply at TCal ±5°C, with allowances for temperature variations.
Frequency Specifications:
The frequency specifications provide accuracy details for various frequency ranges, maintaining consistent accuracy across different time frames.
Notes:
Important considerations include output resistance, safety features, and the impact of temperature on specifications.
General Specifications:
  • Temperature Range: 25°C (77°F) to 200°C (392°F) at ambient 20°C (68°F).
  • Resolution: 0.1°C.
  • Heating Time: From ambient to maximum (200°C) in 10 minutes.
  • Stability: 0.1°C over 10 minutes.
  • Accuracy: 1°C over 1 year.
  • Probe Insert Diameter Options: 3mm (0.11"), 4mm (0.25"), 5mm (0.19"), 6mm (0.23"), 9mm (0.34").

Note: Specifications may change due to continuous development.

9041A Extended Specifications - Dry Block Well Option:
  • Version: V1.60, Page 21.
  • Specifications are similar to the general specifications with a focus on dry block well applications.
9041A Extended Specifications - Black Body Source Option:
  • Version: V1.60, Page 22.
  • Temperature Range: 25°C (77°F) to 200°C (392°F) at ambient 20°C (68°F).
  • Resolution: 0.1°C.
  • Heating Time: From ambient to maximum (200°C) in 10 minutes.
  • Stability: 0.2°C over 10 minutes.
  • Accuracy: 1°C over 1 year.
  • Target Plate Diameter: 40mm (1.5").

Note: Specifications may change due to continuous development.

See more

Catalog excerpts

MODEL 9041 EXTENDED SPECIFICATIONS-1

MODEL 9041 EXTENDED SPECIFICATIONS

 Open the catalog to page 1
MODEL 9041 EXTENDED SPECIFICATIONS-2

Warm Up Time Standard Interfaces Optional Interfaces Temperature Performance Relative Humidity Altitude EMC & Safety Line Power Power Consumption Low Analogue Isolation Connections Display Information Keyboard Fuses Isolation Dimensions & Weights Warranty Period General Specifications Double the time since last used up to 20 minutes maximum USB RS232 (Serial) Storage : -5°C to +60°C Operation : 0°C to +50°C Operation : <80% to 30°C, <70% to 40°C, <40% to 50°C Storage : <95%, non-condensing Operation : 3000m (10,000ft) Maximum Transit : 12000m (40,000ft) Maximum The calibrator line input plug...

 Open the catalog to page 2
MODEL 9041 EXTENDED SPECIFICATIONS-3

1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Typical Output Stability (Accuracy relative to calibration Standards) Range Notes Note 1 : Allowance must be made for output resistance when driving into a load. Note 2 : Limited by 50 Ohm output impedance. Note 3 : Internally adjustable from 2mA to 30mA - Factory set to 20mA as standard. For safety the trip is controlled by a fail-safe circuit independant of the processor which shuts the high voltage output off in the event of an overload. Note 4 : Typical RMS noise figures at 50% of full scale, bandwidth 1 to 10Hz. High...

 Open the catalog to page 3
MODEL 9041 EXTENDED SPECIFICATIONS-4

1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters 1 Year Total % set uA Compliance Voltage Overload Protection Stability (Accuracy relative to calibration Standards) Range Notes Note 1 : Typical RMS noise figures at 50% of full scale. Note 2 : Power & temperature sensor on 30A range - microprocessor monitors & protects from overheating. Higher resistance loads allow a longer ON period. See graph 1 for details. Note 3 : Specifications apply to loads of less than 10% of the maximum burden voltage. Note 4: Zero or floor allowance. Specifications apply at TCal ± 5°C. Outside this...

 Open the catalog to page 4
MODEL 9041 EXTENDED SPECIFICATIONS-5

Graph 1* : Operating time on 30A range with current into a short circuit at 20°C Continuous current in availiable up to 10A output. * Note Timing is started after a minimum period of 7 minutes at zero output. Shorter periods will reduce the output time availiable. Due to continuous development specifications may be subject to change. 9041A Extended Specifications DCI Specifications : V1.60

 Open the catalog to page 5
MODEL 9041 EXTENDED SPECIFICATIONS-6

1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Typical Output Resistance 50 Ohms 50 Ohms 50 Ohms 50 Ohms 50 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.2 Ohms 0.5 Ohms 0.5 Ohms 0.5 Ohms 0.7 Ohms 0.7 Ohms 0.7 Ohms Stability (Accuracy relative to calibration Standards) Range All specifications apply from 10% of full scale. 5 AC Frequency Accuracy : 30ppm of Setting Notes Note 1: Current limited by 50 ohms output resistance. Note 2 : Internally adjustable from 2mA to 30mA - Factory set to 20mA as standard For safety the trip is controlled...

 Open the catalog to page 6
MODEL 9041 EXTENDED SPECIFICATIONS-7

2 Wire output / Remote sensing not available. Maximum floating voltage : 100V. THD less than .6% Isolation : Floating or grounded selection available as standard. Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. High Voltage Safety High voltage output is ramped to allow instruments under test to auto-range. Standby is automatically activated when setting voltages greater than 20V or 200V from a lower voltage Standby is automatically selected for high voltage (>20V) after 20 minutes on the same setting for frequencies up to 5kHz...

 Open the catalog to page 7
MODEL 9041 EXTENDED SPECIFICATIONS-8

1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Overload Protection Maximum Burden Voltage (Peak) All specifications apply from 10% of full scale. AC Frequency Accuracy : 30ppm Settling Time: For 50% change in output: Less than 3 second from standby to within spec Inductive Loads : Up to 1H may be connected without additional protection providing the frequency / inductance combination does not exceed the maximum burden voltage. Stability (Accuracy relative to calibration Standards) Range Frequency Resolution Due to continuous development specifications may be subject...

 Open the catalog to page 8
MODEL 9041 EXTENDED SPECIFICATIONS-9

Notes Note 1 : Temperature sensor on 30A range - microprocessor monitors & protects from overheating. Higher resistance loads allow a longer ON period. See graph 5 for details. Note 2 : Specifications apply to loads of less than 10% of the maximum burden voltage. Driving Coils and Inductive Loads When driving any load exceeding the maximum compliance voltage will cause the calibrator to trip into standby The maximum compliance voltage on the 10Amp range is specified at a max 2.8V RMS, 7.8V Peak to Peak at 220V supply Slightly higher compliances are available when powered from a 240V supply. When...

 Open the catalog to page 9
MODEL 9041 EXTENDED SPECIFICATIONS-10

Frequency Specifications Total Accuracy - Standard Accuracy Range Total Accuracy - High Accuracy (Option) Range Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. Due to continuous development specifications may be subject to change. 9041A Extended Specifications Frequency Specifications : V1.60

 Open the catalog to page 10
MODEL 9041 EXTENDED SPECIFICATIONS-11

For the highest possible accuracy and dependability of the measured value, regardless of the measurement technique used, the 3000 Series calibrators use passive standard resistors, the calibrated value of which is displayed when selected. 1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Maximum Current Maximum Voltage Display Resolution 1 Year Total Accuracy Stability (Accuracy relative to calibration Standards) Range Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. Due to continuous development specifications...

 Open the catalog to page 11
MODEL 9041 EXTENDED SPECIFICATIONS-12

Capacitance Specifications For the highest possible accuracy and dependability of the measured value, regardless of the measurement technique used, the 3000 Series calibrators use passive standard capacitors, the calibrated value of which is displayed when selected. General Specifications Range Maximum Voltage Display Resolution Specifications apply at 1kHz. Allow 20pF for lead effects. No appreciable variation is noticable at frequencies below 1kHz. Total Accuracy Range Measurement methods Cp up to 1uF Cs above 1uF Capacitance is calibrated as value at the terminals ie. displayed value incorporates...

 Open the catalog to page 12