EXTENDED SPECIFICATIONS

 Open the catalog to page 1
1000 SERIES-2

1000 SERIES EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Due to continuous development specifications may be subject to change. 1000 Series Extended Specifications General Specifications : V1.03

 Open the catalog to page 2
1000 SERIES-3

1000 SERIES EXTENDED SPECIFICATIONS 1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Notes Note 1 : Allowance must be made for output resistance when driving into a load. Note 2: Internally adjustable from 2mA to 15mA - Factory set to 12mA as standard. For safety the trip is controlled by a fail-safe circuit independant of the processor which shuts the high voltage output off in the event of an overload. Note 3: Typical RMS noise figures at 50% of full scale, bandwidth 1Hz to 10Hz 2 Wire output / Remote sensing not available. Specifications apply at TCal ± 5°C Outside...

 Open the catalog to page 3
1000 SERIES-4

1000 SERIES EXTENDED SPECIFICATIONS 1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Notes Note 1 : Power & temperature sensor on 10A range - microprocessor monitors & protects from overheating. Note 2 : Specifications apply to loads of less than 10% of the maximum burden voltage. Note 3 : Zero or floor allowance. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. Due to continuous development specifications may be subject to change. 1000 Series Extended Specifications DCI Specifications : V1.03

 Open the catalog to page 4
1000 SERIES-5

1000 SERIES EXTENDED SPECIFICATIONS ACV Specifications TRANSMILLE LTD All specifications apply from 10% of full scale. AC Frequency Accuracy = 30ppm of Setting Notes Note 1 : Internally adjustable from 2mA to 15mA - Factory set to 12mA as standard Note 2 : For safety the trip is controlled by a fail-safe circuit independant of the processor which shuts the high voltage output off in the event of an overload. Note 3 : Allowance must be made for output resistance when driving into a load. Note 4 : 2 Wire output / Remote sensing not available. Note 5 : THD less than .6% Specifications apply at TCal...

 Open the catalog to page 5
1000 SERIES-6

1000 SERIES EXTENDED SPECIFICATIONS 1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Notes All specifications apply from 10% of full scale. Settling Time: For 50% change in output: Less than 3 second from standby to within specifications Inductive Loads: Up to 1H may be connected without additional protection providing the frequency / inductance combination does not exceed the maximum burden voltage. Temperature sensor on 10A range - microprocessor monitors & protects from overheating. Higher resistance loads allow a longer ON period. Specifications apply to loads of...

 Open the catalog to page 6
1000 SERIES-7

1000 SERIES EXTENDED SPECIFICATIONS Simulated Resistance 1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters Range Notes Minimum terminal voltage = 80mV Maximum input current = 320mA Input measurement current must be a constant DC current, isolated from earth. Current must be stable for a period of 1s. Use manual range on the UUT. The 2-Wire value is measured at the terminals. DC measurement technique used. Use passive resistance for AC component bridges. Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. Due to...

 Open the catalog to page 7
1000 SERIES-8

1000 SERIES EXTENDED SPECIFICATIONS Continuity Resistance TRANSMILLE LTD Notes Minimum terminal voltage = 80mV Maximum input current = 320mA Input measurement current must be a constant DC current, isolated from earth. Current must be stable for a period of 1s. Use manual range on the UUT. The 2-Wire value is measured at the terminals. DC measurement technique used. Use passive resistance for AC component bridges. Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. Due to continuous development specifications may be subject to change....

 Open the catalog to page 8
1000 SERIES-9

1000 SERIES EXTENDED SPECIFICATIONS Passive Resistance For the highest possible accuracy and dependability of the measured value, regardless of the measurement technique used, the 1000 Series calibrators use passive standard resistors, the calibrated value of which is displayed when selected. Notes The 2 Wire value for each resistor is calibrated. The 2-Wire value is measured at the terminals. Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added._ Due to continuous development specifications may be subject to change. 1000 Series Extended...

 Open the catalog to page 9
1000 SERIES-10

1000 SERIES EXTENDED SPECIFICATIONS Capacitance Specifications For the highest possible accuracy and dependability of the measured value, regardless of the measurement technique used, the 1000 Series calibrators use passive standard capacitors, the calibrated value of which is displayed when selected. 1 Year Total Accuracy Specifications at TCal ±5°C Notes Specifications apply at 1kHz. Allow 20pF for lead effects. No appreciable variation is noticable at frequencies below 1kHz. Capacitance is calibrated as value at the terminals, internal wiring is compensated for Specifications apply at TCal...

 Open the catalog to page 10
1000 SERIES-11

1000 SERIES EXTENDED SPECIFICATIONS Simulated PRT TRANSMILLE LTD Notes Minimum terminal voltage = 80mV Maximum input current = 320mA Input measurement current must be a constant DC current, isolated from earth. Current must be stable for a period of 1s. Use manual range on the UUT. The 2-Wire value is measured at the terminals. DC measurement technique used. Use passive resistance for AC component bridges. Specifications apply at TCal ± 5°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. Due to continuous development specifications may be subject to change. 1000...

 Open the catalog to page 11
1000 SERIES-12

1000 SERIES EXTENDED SPECIFICATIONS Thermocouple Output 1 Year Total Accuracy Specifications at TCal ±5°C Thermocouple Type Due to continuous development specifications may be subject to change. 1000 Series Extended Specifications The

 Open the catalog to page 12