1. Catalogs
  2. Topas GmbH
  3. Laser Aerosol Particle Counter Series LAP 340

Laser Aerosol Particle Counter Series LAP 340

Laser Aerosol Particle Counter Series LAP 340

Laser Aerosol Particle Counter Series LAP 340

Product catalog summary
Overview: The Laser Aerosol Particle Counter LAP 340 is a portable evaluation system designed for measuring particle number and size in air streams and gaseous media. It operates within a measuring range of 0.3 to 10 µm across 16 particle size channels at a flow rate of 28.3 l/min, making it suitable for various applications including clean room monitoring and filter efficiency testing.
Key Features:
  • Portable for both stationary and mobile use.
  • Complies with standards such as VDI 2083, DIN 1946, DIN 12950/12980, US Federal Standard 209 E, and DIN EN 14644.
  • Automatic calibration function for easy system checking and recalibration.
  • User-friendly software for measurement and data archiving.
  • Quick setup and operation.
Applications:
  • Validation and monitoring of clean rooms and laminar air flow boxes.
  • Emission measurement, e.g., vacuum cleaner testing.
  • Fractional efficiency tests of filters and separators.
Technical Specifications:
  • Flow generation via vacuum pump.
  • Optical system uses 90° light scattering optics.
  • Graphical LCD display with 19-key operation.
  • Measuring time ranges from 1 second to 99 hours.
  • Single or continuous measurement modes available.
  • Data includes date, time, sample ID, measuring time, and particle size distribution.
  • Programmable alarms for clean room classes.
  • Thermal printer for data output.
  • RS 232 C computer interface.
  • Power supply: 230 V AC, 50/60 Hz, max. 250 W.
  • Dimensions: 410 x 280 x 180 mm; Weight: 13 kg.
Accessories:
  • Aerosol Generator Series ATM 226.
  • Dilution System Series DIL 554/T.
  • Isokinetic Sampling Probe SYS 529.
  • Topas Particle Analysis Software PASWin.
  • Sensors for temperature, humidity, pressure, and air velocity.
  • Remote control for measurements up to 5 m away.
  • Antistatic flexible tube.
Calibration and Accuracy: The sensor is calibrated with certified Latex particles, ensuring a counting efficiency rate of at least 50% for the smallest particle size. The maximum concentration is indicated at a coincidence error of 7.8%.
Quality Assurance: The system is QMS certified to DIN EN ISO 9001.
See more

Catalog excerpts

Laser Aerosol Particle Counter Series LAP 340-1

TOPAS 1® Laser Arosol Particle Counter LAP 340 Advantages Portable particle counter for stationary as well as mobile use 镕 Complies with the following standards: VDI 2083 DIN 1946 DIN 12950/12980 US Federal Standard 209 E DIN EN 14644 Automatic calibrating function enables an easy checking and recalibration of the system Օ User-friendly software for measuring and archiving Quick instrument setup Applications Օ Validation and monitoring of Clean rooms and Laminar air flow boxes Emission measurement (e. g. vacuum cleaner testing) Օ Fractional efficiency tests of filters and separators Laminar Air Flow Box Laser Arosol Particle Counter LAP 340 Laser Arosol Particle Counter LAP 340 The particle counter LAP 340 is a mobile useable 驩valuation system for measuring the number and size of particles in air streams and gaseous media. The measuring range comprises 0.3 ... 10 |_im in 16 particle size channels at a flow rate of 28.3 l/min. This makes it suitable for a variety of measuring tasks. These include monitoring and checking of clean rooms, clean benches, filter facilities in operation rooms or filling facilities in the pharmaceutical industry. Another area of application, especially in connection with the Topas Dilution System Series DIL, are separation efficiency tests of filters. An LCD display allows during the measurement an overview of various data such as particle size channels, date, time and the current measurement. The measuring results are shown in a table. At the end of the measurement, a thermal printer enables a print out of the data together with threshold value exceedings and sample identification. With the in-house developed Topas software package PASWin further processing of the measuring results can be done in programs like MS Excel. The software also allows direct measuring and archiving. Arrangement of Topas instruments Aerosol Generator ATM, Dilution System DIL, Sampling Probe SYS 529 and Laser Aerosol Particle Counter LAP 340 for validation of a laminar air flow box BS

 Open the catalog to page 1
Laser Aerosol Particle Counter Series LAP 340-2

Spcification Measuring Range Technical Data Flow generation Vacuum pump Optical system 90° light scattering optics Display and Background-lighted graphical control LCD display, 19-key- operation Measuring time / 1 (or 0) seconds ... 99 hours pause time Measuring mode Single measurement technique with 1.9 measurements per cycle or continuous measurement Data contents Date, time, sample per data batch identification / number, measuring time, pause time, 16 programmable particle size channels with distributive, cumulative particle number and volume distribution, mean value depending on setting Alarm...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.