1. Catalogs
  2. TIRA GmbH
  3. TV 51010/LS-230
video corpo

TV 51010/LS-230

TV 51010/LS-230

TV 51010/LS-230

Product catalog summary
Technical Parameters of Vibration Exciter S 51010/LS-230:
  • Rated Peak Force: 11,000 Npk RMS for Sine/Random, 33,000 Npk for Shock.
  • Frequency Range: 2 - 3000 Hz.
  • Main Resonance Frequency: Greater than 2300 Hz.
  • Max Displacement: 50.8 mm Peak-Peak.
  • Max Velocity: 2.0 m/s for Sine/Random, 3.0 m/s for Shock.
  • Max Acceleration: 85 g for Sine, 65 g for Random, 220 g for Shock.
  • Suspension Stiffness: 75 N/mm.
  • Effective Moving Mass: 13 kg (±5%).
  • Max Payload: 250 kg.
  • Total Mass: 1100 kg.
  • Magnetic Stray Field: Less than 1.5 mT standard, less than 0.8 mT with low degaussing.
  • Armature Diameter: 230 mm.
  • Compressed Air Supply: Minimum 600 kPa.
  • Interlocks: Field coil temperature, displacement, cooling air, overcurrent, compressed air.
Scope of Delivery:
  • Vibration exciter 11 kN.
  • Trunnion mount.
  • Power amplifier 15 kVA.
  • Cooling blower.
  • Connection and power cables.
  • Blower and compressed-air hoses.
Options:
  • Thread inserts customization.
  • Low degaussing kit.
  • Squeak & Rattle operation.
  • Wheels & Rails.
  • Thermobarrier for temperature extremes.
  • Climatic chamber support.
  • Remote control software.
  • Silencer and acoustic enclosure for noise reduction.
  • Cable extension and factory acceptance test.
Features:
  • Vibration isolation below 6 Hz.
  • Coarse filter unit.
  • Automatic pneumatic load compensation.
  • Automatic armature centering.
  • Degauss kit for stray magnetic field reduction.
  • Made in Germany with service hotline support.
Technical Parameters of Cooling Blower TB 120 FUK:
  • Max Volume Flow Rate: 1500 m³/h.
  • Total Pressure Difference: 10.2 kPa.
  • Motor Output: 5.5 kW.
  • Max Frequency: 100 Hz.
  • Hose Diameter: 140 mm.
  • Max Sound Pressure Level: 102 dB(A).
Technical Parameters of Amplifier A 1 02 11 021 SV:
  • Max Output Power: 15,000 VARMS.
  • Frequency Range: DC - 5 kHz.
  • Voltage and Current: ±212 VRMS, 40-100 ARMS.
  • Signal Input Voltage: 10 VRMS.
  • Total Harmonic Distortion: Less than 0.2% RMS.
  • Signal to Noise Ratio: Greater than 80 dB.
  • Field Voltage and Current: 140-280 V, 6-8 A.
  • Interlocks: Overload, temperature, displacement, cooling air, compressed air, phase monitoring, emergency stop.
See more

Catalog excerpts

TV 51010/LS-230-1

TECHNICAL PARAMETERS Vibration exciter S 51010/LS-230 Rated peak force Sinepk/RandomRMS1/Shockpk2 Frequency range Main resonance frequency Max. displacement Peak-Peak Max. velocity Sine/Random/Shock Max. acceleration Sine/Random/Shock Suspension stiffness Effective moving mass (±5%) Max. payload Total mass Magn. stray field Std./low degaussing3 Armature diameter Min. required compressed air supply Interlocks 11000/11000/33000 N 2 - 3000 Hz >2300 Hz 50.8 mm 2.0/2.0/3.0 m/s 85/65/220 g 75 N/mm 13 kg 250 kg 1100 kg <1.5/<0.8 mT 230 mm 600 kPa Field coil temperature, displacement, cooling air, overcurrent, compressed air 1) Random force according to ISO 5344:2004 2) Theoretical maximum shock value. Depends on payload, amplifier, shock and shock width 3) measured at 150 mm above armature For long-term tests, the load must be reduced to 80 %. Continuous operation at maximum load can cause damage. SCOPE OF DELIVERY, OPTIONS AND FEATURES OF THE SYSTEM Scope of delivery: Vibration exciter 11 kN Trunnion mount Power amplifier 15 kVA Cooling blower Connection cable (5 m) Power cable (5 m) for amplifier (CEE 32 connector) Blower hose ø140 mm (5 m) Compressed-air hose NW 7,2 (Standard) (3 m) Options: Different thread inserts in the armature at customer request Low degaussing kit to further reduce stray magnetic field Squeak&Rattle (Silent operation without blower) Wheels&Rails (incl. 3m rails) Thermobarrier (-40°C to +140°C) Chamber leadthrough Climatic chamber support kit Remote control (Software) Silencer for cooling blower (Noise reduction 3 - 6 dB(A)) Acoustic enclosure for cooling blower (Noise reduction 5 - 23 dB(A)) Cable extension Factory acceptance test Options: Energy Management System Operation with temperature-controlled cooling blower (and optional with variable field strength) ASM-Mode (Auto Shutdown Manager) Features: Vibration isolation < 6 Hz Coarse filter unit Fully automatic pneumatic load compensation Automatic centering of the armature Degauss kit to reduce stray magnetic field Made in Germany Servicehotline GmbH © TIRA GmbH • Version 10 - 2021-09-13 Eisfelder Str. 23/25, 96528 Schalkau, Germany • Tel.: +49 36766 280-0 • Fax: +49 36766 280-99 • Internet: www.tira-gmbh.de • Email: [email protected] Subject to modifications S 51010/LS-230 (Example drawing) Dimensions in mm TIRA Vibration Test Systems Vibration Test System

 Open the catalog to page 1
TV 51010/LS-230-2

TECHNICAL PARAMETERS Cooling blower TB 120 FUK Max. output powerRMS (factory-set) Frequency range VoltageRMS, max. CurrentRMS, max. (factory-set) Signal input voltageRMS Total Harmonic Distortion (at 70ARMS, 200 Hz) Signal to noise ratio Field voltage (factory-set) Field current (factory-set) Total mass Dimensions (WxHxD) Power supply (Standard) Recommended fuse protection Max. power consumption at 400 V (incl. blower) Interlocks: Max. volume flow rate Total pressure difference Motor output Max. frequency Hose diameter Hose length (Std.) Total mass Dimensions (WxHxD) Max. sound pressure level...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.