1. Catalogs
  2. Time Electronics
  3. 7000 RTD Temperature Calibrator Data Sheet
video corpo

7000 RTD Temperature Calibrator Data Sheet

7000 RTD Temperature Calibrator Data Sheet

7000 RTD Temperature Calibrator Data Sheet

Product catalog summary
Description
The Time Electronics 7000 RTD Temperature Calibrator is a portable instrument designed for high accuracy temperature measurements using RTD probes. It combines a precision digital thermometer with an RTD/ohms calibrator, suitable for laboratory, workshop, or field use. It operates on rechargeable batteries or an external mains adaptor and can serve as an external temperature reference.

Modes of Operation
Monitor Mode (4 wire)
- Checks RTD probes by measuring resistance at known temperatures.
- Measures resistance values and indicates temperature when connected to an RTD probe.
- Features include auto re-calibration every 0.6 seconds, temperature stability better than 0.0015% per °C, and automatic logging of max/min values.

Simulator Mode (4 wire)
- Outputs resistance of precise known value and simulates RTD values using internal tables.
- Offers programmable ramp function and fixed step points between user-set minimum and maximum.

Features
- Temperature accuracy: 0.05 °C (0.09 °F), resolution: 0.01 °C (0.02 °F).
- Resistance accuracy: 0.03 Ω, resolution: 0.01 Ω.
- Supports 2, 3, and 4 wire connections and measures/simulates °C, °F, °K, & Ω.
- Compatible with PT100 and 7 other RTD types, user programmable.
- Typical battery life of 24 hours between charges.

Technical Specifications
- Standard RTD types include Pt100, Pt200, Pt500, Pt1000, Ni 120, and Ni 1000 with specified temperature ranges and accuracies.
- Resistance accuracy varies by range, with additional error for excitation currents below 1 mA.

General Specifications
- Operating temperature range: 10 to 50 °C.
- Powered by NiMH rechargeable batteries or external mains adaptor.
- Dimensions: H 165 x W 90 x D 45 mm, Weight: 0.42 kg.
- Optional extras include mains adaptors and calibration certificates.

Ordering Information
- 7000: RTD Temperature Calibrator
- 7633: Mains adapter 230V AC
- 7633: Mains adapter 110V AC
- C183: Traceable calibration certificate (Factory)
- C194: Accredited calibration certificate (ISO 17025)
See more

Catalog excerpts

7000 RTD Temperature Calibrator Data Sheet-1

Time Electronics Description The 7000 is a portable process control test instrument that combines a precision digital thermometer (using RTD probes) with an RTD/ohms calibrator. Compact and easy to use, it solves the problem of making high accuracy temperature measurements without using bulky mains powered instrumentation. Powered from internal long life rechargeable batteries or an external mains adaptor, it is equally valuable in laboratory, workshop or the field. It can also be used as an external temperature reference for dry block and other precision temperature baths. • PT100 plus 7 other RTD types Monitor mode (4 wire) • 24 hours typical use between charges • Temperature accuracy 0.05 °C (0.09 °F) • Temperature resolution 0.01 °C (0.02 °F) • Resistance accuracy 0.03 Ω • 2, 3, and 4 wire connections • Measure and simulate °C, °F, °K, & Ω • Ramp and step • User programmable Typical functions • Check RTD probes by measuring their resistance at known temperatures. • Measure resistance values. • Indicate temperature when connected to an RTD probe. Simulator Mode (4 wire) • Pre-programmed with a particular RTD’s characteristics for very high accuracy. Enhanced performance may be achieved by programming the unit to simulate the characteristic of a particular probe. Five fixed step points (0, 25, 50, 75, 100%) are available between a user set minimum (0 %) and a maximum (100 %). Programmable ramp function is also available. Time Electronics Ltd, Unit 5, T O N Business Park, 2-8 Morley Road, Tonbridge, Kent, TN9 1RA, United Kingdom T: +44 (0) 1732 355993 | F: +44 (0) 1732 350198 | mail@ti

 Open the catalog to page 1
7000 RTD Temperature Calibrator Data Sheet-2

Time Electronics7000 RTD Temperature CalibratorTechnical SpecificationsStandard RTD types (non standard RTD types user programmable)Element Alpha coefficient Temp range °C Accuracy °C Temp range °F Accuracy °F Pt100 DIN Resistance accuracyRange (n) Monitor (n) Generator (n) 20 to 400 plus additional error of 0.05 % of output value. If excitation current is less than 1 mA. 10 to 50 °C. NiMH re-chargeable. External mains adaptor. > 24hrs typical use. Impact resistant ABS. H 165 x W 90 x D 45 mm. 0.42 kg. Mains adaptor 230 V AC, Mains adaptor 110 V AC. Calibration certificates: Traceable (Factory)...

 Open the catalog to page 2

All Time Electronics catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.