Overview
The Time Electronics 5018 is a high-accuracy programmable DC/AC voltage and current calibrator designed for production environments and R&D. It can function as a dedicated DC voltage source or an advanced AC/DC calibrator, controlled via PC or laptop, and is suitable for complex ATE test rigs.

Key Features
  • DC Voltage: 0 to 22 V, with optional ranges up to 220 V DC and 1 kV DC.
  • AC Voltage: Optional 0 to 22 V AC.
  • Current: Optional 0 to 220 mA AC/DC.
  • Accuracy: 15 ppm, Resolution: 0.5 ppm.
  • Interfaces: GPIB, RS-232, USB for PC connectivity.
  • Ramping feature for testing analogue gauges.
  • EasyCal software for automated calibration and traceable certificates.

Technical Specifications
  • DC Voltage: Ranges from 22 mV to 22 V with varying accuracies and resolutions.
  • DC High Voltage Options: 220 V and 1 kV ranges with specific accuracies and resolutions.
  • DC Current Option: Ranges from 220 μA to 220 mA with compliance voltage of 11 V.
  • AC Voltage Option: Frequency range from 10 Hz to 20 kHz with specific accuracies.
  • AC Current Option: Frequency range from 20 Hz to 1 kHz with compliance voltage of 5 V.

General Specifications
  • Warm-up time: 1 hour to full accuracy.
  • Settling time: Less than 5 seconds.
  • Operating temperature: 10 to 40 °C, full spec at 23 °C ± 5 °C.
  • Power: 100 to 230 V AC, 50/60 Hz, 60 W typical consumption.
  • Dimensions: 450 x 272 x 152 mm, Weight: 8.2 kg.

Ordering Information
  • Various internal options for current, AC, and high voltage.
  • Accessories include rack mount kits, test lead sets, and calibration certificates.

Note: Specifications are subject to change without notice and apply at 23 °C ± 5 °C after a 1-hour warm-up.
See more

Catalog excerpts

5018 Benchtop Voltage and Current Calibrator Data Sheet-1

Time Electronics 5018 Programmable DC-AC V-I Calibrator High Accuracy Programmable DC/AC Source The 5018 is a versatile instrument capable of covering a wide range of testing applications. It can be configured as a dedicated single function DC voltage source, or as an advanced AC/DC voltage and current calibrator. Primarily designed for production environments and R&D, the 5018 can be controlled via PC or laptop, programmed to perform any number of tasks as part of a complex ATE test rig. Ramping Feature The 5018 includes an internal ramping feature. A ramp rate per range for each function can be set via a PC. The option to ramp the output can be turned on or off via the front panel. This is particularly useful in aviation and automotive industries for testing analogue gauges. Simple Operation Front panel operation allows the user to quickly set the function and output required. Using the jog / shuttle dial deviation the user can finely adjust the output value as a percentage (± 99.99 %). All this information is shown on a clear, easy to read LED display. GPIB, RS-232, and USB Interfaces These interfaces allow the 5018 to be connected to a PC and controlled by an external program such as EasyCal calibration software. The 5018 uses a SCPI command structure for programming. This makes writing control programs in Visual Basic, C and Labview a simple task. Features • 0 to 22 V DC • Optional 0 to 22 V AC, 220 V DC and 1 kV DC • Optional 0 to 220 mA AC/DC • 15 ppm accuracy, 0.5 ppm resolution • 1999999 full scale +10 % over-range • Deviation control –9.999 % to +9.999 % • Ramping feature • Ideal for ATE applications in production and R&D • Rack mount kit option • RS-232, USB, GPIB interfaces EasyCal Calibration Software The 5018 can be controlled via Time Electronics EasyCal software to automate the calibration process. This provides increased speed of calibration and consistency of results. Produce traceable calibration certificates and test reports for quality standards with additional uncertainty information for ISO 17025 conformance. EasyCal software is primarily used for calibration work, but it is also a powerful automation tool to control and readback from a wide range of programmable instruments. So it can be utilised with comprehensive ATE systems and configured to control a variety of applications. The readback capabilities enable closed loop testing, and conditional tests can be configured to set up automated decision making. Time Electronics Ltd, Unit 5, T O N Business Park, 2-8 Morley Road, Tonbridge, Kent, TN9 1RA, United Kingdom T: +44 (0) 1732 355993 | F: +44 (0) 1732 350198 | [email protected] | www

 Open the catalog to page 1
5018 Benchtop Voltage and Current Calibrator Data Sheet-2

Time Electronics 5018 Programmable DC-AC V-I Calibrator Technical Specifications Accuracy specifications are shown as ppm (or %) of output + floor and apply for settings between 10 % and 100 % of range. Specifications apply at an ambient temperature of 23 °C ± 5 °C after the calibrator has warmed up for at least 1 hour. DC Voltage Accuracy 1 year Output resistance DC High Voltage options (9721: 220 V range only / 9720: Both ranges) Range Output resistance DC Current option 9718 Range Compliance voltage AC Voltage option 9719: 10 Hz to 20 kHz (sine 0.01 % crystal controlled) Range RMS Output resistance...

 Open the catalog to page 2

All Time Electronics catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.