video corpo

5011 Resistance and Temperature Calibrator Data Sheet
2Pages

{{requestButtons}}

Catalog excerpts

5011 Resistance and Temperature Calibrator Data Sheet - 1

Time Electronics 5011 Resistance and Temperature Calibrator Description A versatile, high accuracy calibrator that can be used in a wide range of applications across various industries. Primarily used as a programmable resistance and RTD source, the 5011 provides precision resistance with a best accuracy of 0.01 % and 1 mΩ resolution (50 Ω to 1 kΩ). Internal options for increased capabilities can be fitted as per customer requirement. These include DC voltage and thermocouple simulation, DC current, and frequency. The 5011 can be used to cover a wide workload as a laboratory calibrator or be incorporated into an automated test system. A rack mount kit option is also available. Simple operation and clear visual display Front panel operation allows the user to quickly set the function and output required. Using the jog / shuttle dial deviation the user can finely adjust the output value as a percentage (± 9.99 %). All this information is shown on a clear, easy to read LED display. As standard the 5011 uses a SCPI command structure for programming. It also supports the Time Electronics 9811/19/20 command set making it an ideal replacement for these legacy models. Flexible options A range of internally fitted options are available for the 5011, providing added functionality and features. The DCV / Thermocouple option provides a DC voltage source (± 20 V) and simulates thermocouple types K, J, T, R, S, N, E and B. With the DC current option fitted the 5011 can source up to 220 mA. This makes the instrument ideal for accurate process control calibration. Also available is a 0.1Hz to 10MHz frequency option (Period 100 ns to 10 s). The output is variable with a best resolution setting of 0.1 Hz. Features • 1 Ω to 120 MΩ • 100 ppm basic accuracy • RTD simulation • Optional Thermocouple Simulation • DCV and DCI options • 10 MHz frequency option • RS-232, GPIB, USB interfaces • Front panel operation • PC/laptop control via EasyCal software EasyCal Calibration Software The 5011 can be controlled via Time Electronics EasyCal software to automate the calibration process. This provides increased speed of calibration and consistency of results.Produce traceable calibration certificates and test reports for quality standards with additional uncertainty information for ISO 17025 conformance. Time Electronics Ltd, Unit 5, T O N Business Park, 2-8 Morley Road, Tonbridge, Kent, TN9 1RA, United Kingdom T: +44 (0) 1732 355993 | F: +44 (0) 1732 350198 | mail@timeelectronics.co.uk | w

Open the catalog to page 1
5011 Resistance and Temperature Calibrator Data Sheet - 2

Time Electronics 5011 Resistance and Temperature Calibrator Technical Specifications (Accuracies quoted are for 1 year at 22 ºC ± 3 ºC) Standard features Thermocouple simulation Resolution Cold Junction Compensation ± 0.5 °C (applies to ambient changes of less than ± 1 °C at 23 °C). The accuracy of the thermocouple simulation is determined by the accuracy of the 5011’s DC voltage function and the accuracy of the standard thermocouple tables (BS EN 60584-1) published by the British Standards Institute. The 5011 uses precise digital interpretation of the tables to output voltage levels that...

Open the catalog to page 2

All Time Electronics catalogs and technical brochures

Archived catalogs