VSR - Smartline Vacuum Transducer Piezo / Pirani Absolute Pressure 1200 to 1 x 10 -4 mbar Overview VSR vacuum transducer n Wide measuring range due to combination sensor piezo/Pirani High accuracy and optimal resolution over the whole measuring range Excellent reproducibility and long term stability Stable measurement due to optimized temperature compensation Measurements independent of gas type above 20 mbar and adjustable gas type correction factor for Pirani sensor Short response time of up to 40 ms Integrated metal screen for protection against contamination such as oil and solvent vapors Overview Smartline vacuum transducers n Digital RS485 interfaces plus 0-10 V output signal, EtherCAT or PROFINET, Bluetooth adapter SLKBT for wireless communication PROFINET transducers support MRP Exchangeable sensor heads with stored calibration data Large LCD display (exc. EtherCAT, PROFINET) Readjustment of zero and atmosphere by pushbutton or interface LEDs for device status and switch points Two independent, potentialfree relay switch points (exc. EtherCAT, PROFINET) Protection class up to IP54 0-10 V output signal can be scaled according to the required output characteristics. Exchange of existing vacuum gauges easily, independent of the manufacturer, without programming effort Intelligent sensors for simplified integration according to Industry 4.0 standards Metal sealed stainless steel sensor cell, suitable for UHV (He leackage rate <5e-10 mbar l / s) Expansion with 2 channel controller VD12 or 4 channel controller VD14 possible VacuGraphTM Windows software for visualisation, analysis and storage of measuring data

Open the catalog to page 1
VSR Vacuum Transducer - 2

VSR - Smartline Vacuum Transducer Piezo / Pirani Absolute Pressure 1200 to 1 x 10 -4 mbar Technical Data Measurement Principle Piezo resistive / heat conduction Pirani (Pirani, dep. on gas type) Measurement Range Materials with Vac. Contact Stainless steel 1.4307, tungsten, nickel, glass, gold, silicon oxide Reaction Time Operating Temperature Storage Temperature Max. 150 °C at the flange (voltage supply switched-off) Voltage Supply Power Consumption Max. 2.5 W, add. 0.8 W for EtherCAT / relays /LCD, add. 1.6 W for Profinet Output Signal 0-10 VDC, min. 10 kΩ, measuring range 1.5 to 8.58...

Open the catalog to page 2
VSR Vacuum Transducer - 3

VSR - Smartline Vacuum Transducer Piezo / Pirani Absolute Pressure 1200 to 1 x 10 -4 mbar Model designations VSR53DL DN16 ISO-KF, 0-10 V and Rs485, with LCD display VSR54DL DN16 CF-F, 0-10 V and RS485, with LCD display VSR53E VS´ 54E R DN16 ISO-KF, EtherCAT and RS485 DN16 CF-F, EtherCAT and RS485 VSR53PN DN16 ISO-KF, Profinet and RS485 VSR54PN DN16 CF-F, Profinet and RS485 VD1253P: Set including a 2 channel controller VD12, VSR53D, measuring cable 2 m Accessories n n Replacement sensor heads: B_VSR53, B_VSR54 SLCASE Smartline protective plastic case , SLN4 plug-in power supply, SLKUSB...

Open the catalog to page 3

All Thyracont Vacuum Instruments GmbH catalogs and technical brochures

  1. VSL

    3 Pages

  2. VD81

    3 Pages