Product catalog summary
Overview: The document provides technical specifications and features of the Analogline and VCP vacuum transducers, designed for industrial applications. These devices are used for precise pressure measurement in vacuum environments, particularly suitable for corrosive media.
Specifications:
  • Adjustment and Functionality: The transducers offer digital adjustment to zero pressure and atmosphere with a button press. They are designed with minimal functionalities to optimize cost-performance.
  • Measurement Stability: Equipped with a microprocessor controller and temperature compensation for stable and reproducible measurements.
  • Construction: Robust metal housings with IP54 or IP40 protection, suitable for rugged environments. Compact design for space-constrained applications.
  • Output Signal: Available in linear and logarithmic outputs (4-20 mA or 0-10 V), easily integrated with system controls (PLC).
Analogline Vacuum Transducers:
  • High reliability and excellent reproducibility.
  • Patented impulse Pirani with extended measuring range and high resolution in rough vacuum range.
  • Suitable for aggressive media due to platinum rhodium filament and integrated metal facing for contamination protection.
  • Compatible with 2 channel controller VD12 and UHV suitable due to metal-sealed stainless steel measuring cell.
VCP Vacuum Transducer:
  • Measurement Principle: Heat conduction (Impulse Pirani), dependent on gas type.
  • Measurement Range: 1000 - 5e-4 mbar (750 - 5e-4 Torr).
  • Materials: Stainless steel, nickel, glass, platinum-rhodium filament.
  • Overpressure Stability: Up to 16 bar absolute with CERT31P.
  • Accuracy and Repeatability: Varies across measurement ranges, with response time < 200 ms.
  • Power Supply: 15 - 30 VDC, with max power consumption of 1.5 W.
  • Temperature Range: Operating: +5 to +60°C, Storage: -20 to +70°C, Bake Out: Max 150°C.
  • Output Signal: VCP63MA4: 4 - 20 mA, logarithmic; VCP63MV: 0 - 10 V, logarithmic.
  • Protection Class: IP40 (VCP63MV), IP54 (VCP63MA4).
Accessories:
  • Various measuring cables and mating plugs for different models.
  • Overpressure stability certificate (CERT31P) available.
Dimensions: Detailed dimensions provided for VCP63MV and VCP63MA4 models.
See more

Catalog excerpts

VCP63 Vacuum Transducer-1

VCP - Analogline Vacuum Transducer Pirani for Corrosive Media Absolute Pressure 1000 to 5 x 10-4 mbar Overview VCP vacuum transducer n n Durable, elastic Pirani helix filament Suitable for demanding applications with aggressive media due to platinum rhodium filament Patented impulse Pirani with extended measuring range; very high resolution also in the rough vacuum range High reliability Integrated metal facing for protection against contamination like oil vapours or solvent vapours Overpressure stability up to 16 bar UHV suitable due to metal sealed stainless steel measuring cell (He leakage rate <5e-10 mbar l/s) Suitable for 2 channel controller VD12 Excellent reproducibility Overview Analogline vacuum transducers Stable, reproducible measurement values due to microprocessor controller and individual temperature compensation Precise, digital adjustment to zero pressure and atmosphere at the push of a button. The vacuum transducer automatically recognizes the adjustment point Functionalities of the Analogline are reduced to minimum to achieve an optimal costperformance-ratio Robust metal housings and plugs with protection class IP54 (4-20 mA versions) or IP40 (0-10 V versions), ideal for rugged industry environments Compact construction, ideal for industrial applications with reduced space Linear and logarithmic output signal of the transducer (optionally 4-20 mA or 0-10 V) can easily read out with a system control (PLC)

 Open the catalog to page 1
VCP63 Vacuum Transducer-2

VCP - Analogline Vacuum Transducer Pirani for Corrosive Media Absolute Pressure 1000 to 5 x 10-4 mbar Technical Data Measuring Principle Heat conduction (Impulse Pirani), depending on gas type Measurement Range Materials with Vac. Contact Stainless steel 1.4307, nickel, glass, filament: platinum-rhodium 10 bar absolute, optional: 16 bar abs. (with CERT31P) 1000 - 10 mbar: approx. 30% f. r. 10 - 1e-2 mbar: ±10% f. r. Response Time Power Supply Electrical Connection Hirschmann, 6 pole, male, lockable (VCP63MV) M12 A, 5pole, male, lockable (VCP63MA4) Power Consumption Max. 1.5 W with 24 VDC supply...

 Open the catalog to page 2
VCP63 Vacuum Transducer-3

VCP - Analogline Vacuum Transducer Pirani for Corrosive Media Absolute Pressure 1000 to 5 x 10-4 mbar Model designations DN 16 ISO-KF connection, output 0-10 V logarithmic VCP63MA4 DN 16 ISO-KF connection, output 4-20 mA logarithmic XB0600002 Mating plug, 6pole, for VCP63MV XB0500004 Mating plug, 5 pole, for VCP63MA4 W0606002 Measuring cable, for VCP63MV, shielded, 2 m Overpressure stability 16 bar, incl. testing certificate 3.1 W0509002 Measuring cable, for VCP63MA4, shielded, 2 m Further accessories as well as detailed information about our product family can be found in our Analogline brochure....

 Open the catalog to page 3
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.