2900-System-Specs-2019-01
4Pages

{{requestButtons}}

Catalog excerpts

2900-System-Specs-2019-01 - 1

Thunder Scientific Model 2900 Automated “Two-Pressure” Humidity Generation Syste

Open the catalog to page 1
2900-System-Specs-2019-01 - 2

Model 2900 Automated “Two-Pressure” Humidity Generation System FEATURES • Traceable to SI 1 • 0.5% of Reading RH Uncertainty 5 • High Flow Capability of 50 L/min • Based on the NIST “Two-Pressure” Principle • ControLog® Embedded Automation Software • Automatically Applies Enhancement Factors • HumiCalc® with Uncertainty Mathematical Engine • Generate Multipoint Pro iles • Externally Driven Chamber Fan • Multi-point Touch LCD • Fluid Jacketed Chamber Door • Window Chamber Door Option (WDA) DESCRIPTION The 2900 Humidity Generation System is a self-contained system capable of producing...

Open the catalog to page 2
2900-System-Specs-2019-01 - 3

via the multi-point touch LCD which allows the user to select settings like, pressure, temperature, and flow units. Other settings are °C, °F, atm, Pa, hPa, kPa, L/min, L/ hr, cfm and cfh. Humidity is calculated and displayed in percent relative humidity (%RH). The Parameter Tab is the primary interface for the user and is divided in two sections, Control and Calculated Humidity. Within each tab there are two different tiles, the Setpoint Tile is for the user to enter the desired setpoints and a Value Tile displays the Actual Value. Temperature Control: The 2900 humidity generation system...

Open the catalog to page 3
2900-System-Specs-2019-01 - 4

RESERVOIR FILL PORT 22.00” Automated Humidity Generation System SPECIFICATIONS LEFT BACK ACS REMOTE OPERATION PLUG CHAMBER EXHAUST PORT PRESSURE ADJUSTMENT AIR SUPPLY INLET (1/4” FPT) PRESSURE GAUGE MAIN POWER SWITCH AC POWER Dimensions are excluding feet, handle, and other protrusions. Other custom options are available. 1. Traceable to the International System of Units (SI) through NIST-maintained standards. 2. Using glycol/water as the temperature heat transfer fluid from -10 to 72 °C. Note: the 2900 can only operate for a finite amount of time at or below 0 °C. 3. Temperature Control...

Open the catalog to page 4