![Thames Side Sensors - logo](https://img.directindustry.com/images_di/logo-p/L16671.gif)
Group: Tecnicas
Catalog excerpts
![LCT-U Load Cell Tester Data Sheet - 1](https://img.directindustry.com/pdf/repository_di/16671/lct-u-load-cell-tester-data-sheet-642333_1m.jpg)
LCT-U ADVANCED DIAGNOSTIC LOAD CELL TESTER The advanced Load Cell Tester Model LCT-U is a stand-alone hand-held device specially designed for fast, comprehensive troubleshooting of strain gauge load cells. The LCT-U can quickly diagnose problems with all common types of load cell available today, including 6-wire load cells (with sense wires). The LCT-U provides the operator with all the essential data about the load-cell being tested, including the bridge resistance and integrity, physical distortion (possibly caused by overload, shock load or metal fatigue) and insulation resistance (which can highlight moisture or chemical ingress). ■ Fast, accurate testing of load cells and scales – full test complete within 5 seconds ■ Continuous signal reading allows for checking of linearity and repeatability ■ Very user friendly – pin assignment is printed on front of tester ■ Can check for physical distortion of a load cell ■ Coloured LEDs highlight suspicious readings ■ Can check alignment of multiple load cell scales ■ Operated by standard batteries for portability ■ Works with 4-wire or 6-wire load cells ■ Rugged ABS enclosure and rubber cover ■ Lightweight ■ Ideal tool for installation/commissioning engineers
Open the catalog to page 1![LCT-U Load Cell Tester Data Sheet - 2](https://img.directindustry.com/pdf/repository_di/16671/lct-u-load-cell-tester-data-sheet-642333_2m.jpg)
technical specification... LCT-U Advanced Load Cell Tester Specifications 1.25 VDC Additional Data High resistance testing voltage Sense resistance (for 6 wire load cells) Resistance accuracy Alphanumeric: 16 characters, 2 lines 150 mm x 80 mm x 28 mm Gain adjustment Load cell output in percentage of full scale (input resistance > 175 Ω) 4 x Type AA batteries (not supplied) High resistance accuracy Power source Insulation resistance 8-pin with screw terminals (supplied with the unit) Bridge input and output resistance Bridge testing voltage For further information, including a video user...
Open the catalog to page 2All Thames Side Sensors catalogs and technical brochures
-
VC3500 Load cell Data sheet
4 Pages
-
T85-T Load cell Data sheet
2 Pages
-
T85-N Load cell Data sheet
2 Pages
-
T85 Load cell Data sheet
4 Pages
-
T95 Load cell Data sheet
2 Pages
-
T68 Load cell Data sheet
2 Pages
-
T66 Load cell Data sheet
2 Pages
-
T65 Load cell Data sheet
2 Pages
-
T64 Load cell Data sheet
2 Pages
-
T61 Load cell Data sheet
2 Pages
-
T60 Load cell Data sheet
2 Pages
-
T38 Load cell Data sheet
2 Pages
-
T35D load cell data sheet
2 Pages
-
T34D Load cell Data sheet
2 Pages
-
T35 Load cell Data sheet
2 Pages
-
T30 Load cell Data sheet
2 Pages
-
T20 Load cell Data sheet
2 Pages
-
T16 Load cell Data sheet
2 Pages
-
T12A Load cell Data sheet
2 Pages
-
T12 Load cell Data sheet
2 Pages
-
T6 Load cell Data sheet
2 Pages
-
T5 Load cell Data sheet
2 Pages
-
T4 Load cell Data sheet
2 Pages
-
T2 Load Cell Datasheet
2 Pages
-
650 Load Cell Data Sheet
2 Pages
-
T11 Load Cell
2 Pages
-
LA66 Weighing Assembly
2 Pages