Open the catalog to page 1What is Calibration The comparison of a measuring instrument against an accurate standard to determine any deviation. The device with known assigned correctness is called the Standard. The second device is the Unit Under Test (UUT). Accurate Temperature Calibration Temperature Calibration has many facets. It can be carried out thermally in the case of probes or electrically in the case of Instrument and it can be performed directly with certified equipment, or indirectly with traceable standards. Thermal (Temperature) calibration is achieved by elevating (or depressing) the temperature...
Open the catalog to page 2Division of contact type calibrators according to the temperature range. ^ 1600°C Master Sensor Stable Reference Unit TYPE * Reference temperature could be change according to customer requirement * To be specified at time of ordering
Open the catalog to page 3Division of non-contact type calibrators according to the temperature range. Master Sensor Stable Reference Unit TYPE Reference temperature could be change according to customer requirement * To be specified at time of ordering
Open the catalog to page 4188(D) mm • Insert Construction : 33mm Dia x 220mm long for dry
Open the catalog to page 5O -196 to -80oC portable rtd calibrator CALsys -196/-80 offers ultra low temperature range from -196 to -80°C. It is a highly stable standard source for calibrating Thermocouples / RTD. It has been designed for low temperature range calibration and find application in the glass, pharma, electrical power, automotive & material processing industries. The comparison volume is a metallic block of aluminium having dimension 25mm diameter with 300 mm long which fixed into the liquid nitrogen container for calibration of RTD & thermocouple. The temperature of the calibrator is set and controlled...
Open the catalog to page 6X -80 to 50oC high stability bath CALsys -80/50 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction. CALsys -80/50 offers the temperature range of -80 to 50°C. It has large tank which is filled with liquid Methanol. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid. This CALsys -80/50 is built using CFC free refrigerants. • Reference Standard PRT...Part No. TPRT-A-300 NABL accredited calibration certificate - 3 point. • Software...
Open the catalog to page 79MEMPSENSX -40 to 50oC high stability bath CALsys -40/50 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction. CALsys -40/50 offers the temperature range of -40 to 50°C. It has large tank which is filled with liquid Methanol. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid. This CALsys -40/50 is built using CFC free refrigerants. 500(D) mm (Approx.) Weight : 60 Kg (without • Reference Standard PRT...Part No. TPRT-A-300 NABL...
Open the catalog to page 8high stability bath CALsys -35/200 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction. CALsys -35/200 offers the temperature range of -35 to 200°C. It has large tank which is filled with liquid Methanol/Silicon Oil. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid. This CALsys -35/200 is built using CFC free refrigerants. S High Accuracy S High Stability and Uniformity S Wide Temperature Range S PC Interfacing S Simple to Use...
Open the catalog to page 99MEMPSENScalsys -30/110 low temperature dry block calibrator Calsys -30/110 is low temperature dry block furnace. The dry block furnace based on thermoelectric cooling circuitry. This model provides an special design isothermal enclosure (Aluminium block) in which the thermocouple/RTD can be calibrated against the temperature of the calibrator. The calibrator controller uses a precision RTD as a controlling sensor and controls the well temperature with thermoelectric cooler. To obtain and maintain a required temperature the controller varies the power to the heater via solid-state relay....
Open the catalog to page 10-15 to 110oC peltier based temperature calibrator CALsys -15 / 110 offers medium temperature range from -15 to 110°C. It is a highly stable standard temperature source for calibrating RTD /thermocouples. It has peltier elements to generate stable temperature in the well. The comparison volume is a metallic block of aluminum, which has diameter of 24mm and 120mm length. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment. Controlling Sensor : Precision PRT Method of Control : Digital self tuned PID s Wide Operating Range...
Open the catalog to page 11X 0 to 120oC peltier based temperature calibrator CALsys 120SP offers medium temperature range from 0 to 120°C. It is a highly stable standard temperature source for calibrating RTD /thermocouples. It has peltier elements to generate stable temperature in the well. Use as a dry block & liquid bath using magnetic stirrer in one model. The comparison volume is a metallic block of aluminum, which has diameter of 24mm and 100mm length. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment. specifications Temperature Range...
Open the catalog to page 12HIGH ACCURACY LIQUID CALIBRATION BATH GENERAL CALsys 300SP calibration bath is a highly stable standard stirred liquid bath for calibrating RTDs/Thermocouples and other temperature sensors in the laboratory. The temperature of the bath is set and controlled by a self tuned PID controller with automatic super fine adjustment. The unique features is portability with very high stability and uniformity. The medium used in the bath is silicon oil. Temperature Range : 50 to 250°C Temperature Resolution : 0.1°C Stability :±0.05°C at 50°C ±0.07°C at 150°C ±0.1°C at 250°C Uniformity :±0.1°C at 50°C...
Open the catalog to page 13All TEMPSENS catalogs and technical brochures
-
Nickel & Thermocouple Alloy
26 Pages
-
Thermal Profiling System
10 Pages
-
Laboratory Furnace
13 Pages
-
Tempsens Product Catalogue
60 Pages
-
Thermowells and accessories
12 Pages
-
Petrochemical Industry
22 Pages
-
Cables & Wires
16 Pages
-
Petro-Chemicals Industry
12 Pages
-
Miniature Thermocouple
1 Pages
Archived catalogs
-
Tempsens Products
6 Pages
-
Product overview
2 Pages