1. Catalogs
  2. Teledyne Tekmar
  3. LSS Boat - Lotix Solids Sampler

LSS Boat - Lotix Solids Sampler

LSS Boat - Lotix Solids Sampler

LSS Boat - Lotix Solids Sampler

Product catalog summary
Overview: The document describes the High Temperature Combustion TOC Analyzer designed for analyzing solids. It highlights the automated sample delivery system, reverse flow technique, and the ability to handle both liquids and solids efficiently.
Key Features:
  • Automated sample delivery into the furnace controlled via software.
  • Reverse flow technique eliminates the need for pre-peak analysis.
  • Single calibration curve for 0.1 mg - 20 mg Carbon.
  • Handles up to 1.5 grams of sample size, depending on the matrix.
  • Real-time data viewing and sample report printing.
  • Compact design with a width of only 10 inches, optimizing bench space.
  • Lotix software includes user permissions and maintenance reminders.
Specifications:
  • Required Instrumentation: Lotix High Temperature Combustion Analyzer.
  • Chemistry: Oxidation by combustion at 680 °C - 1000 °C.
  • Analytical Method: TC, TOC, NPOC.
  • Detector: Non-Dispersive Infrared (NDIR) Detector.
  • Boat Material: Quartz.
  • Controller: PC interface through Windows 7 or 10.
  • TOC Analytical Standard Deviation: +/- 2% typical of a mid-range standard.
  • Maximum Sample Amount: Up to 1.5 grams.
  • Measuring Range: 0.1 mg to 20 mg carbon.
  • TOC Analysis Time: Approximately 5-7 minutes.
  • Sample Introduction: Automated delivery via software.
  • Principal Applications: Soils, sludges, sediments, particulate-laden liquids, hard-to-oxidize samples.
  • Certification: CE certified, CSA site certified if required.
  • Utility Requirements: Voltage 100/115/230 VAC, Frequency 50/60 Hz.
  • Dimensions: 10” W x 24” D x 15.8” H.
  • Weight: 32 lbs (14.5 kg).
  • Gas Supply: Hydrocarbon and CO2 free air with TOC content <1 ppm or UHP O2.
  • Gas Inlet Flow: 200 mL/min from Lotix +/- 10%.
Applications and Benefits: The LSS Boat Module is an ideal add-on for the Lotix TOC analyzer, providing accurate carbon content measurement in various matrices such as soils, sludges, and sediments. It is compact, simple, and cost-effective, meeting global user demands for simplicity, performance, and durability.
See more

Catalog excerpts

LSS Boat - Lotix Solids Sampler-1

LSS Boat High Temperature Combustion TOC Analyzer for the Analysis of Solids • Automated sample delivery into furnace via software control • Uses reverse flow technique eliminating the need for pre-peak analysis • 0.1 mg - 20 mg Carbon in a single calibration curve • Up to 1.5 gram sample size giving you a wide analytical range (depending on matrix) • Ability to switch back and forth between liquids and solids in minutes • Real-time viewing of analytical data and printing of sample reports • Only 10" (25.4 cm) wide, maximizing bench space • Upgraded features in Lotix software to include user permissions and maintenance reminders TELEDYNE TEKMAR Everywhereyoulook” www.tekmar.com TOC users from around the globe requested simplicity, performance, and durability for their analyzers. The Teledyne Tekmar LSS Boat Module is an ideal add-on to the Lotix Total Organic Carbon (TOC) analyzer. It uses high-temperature combustion to accurately measure carbon content in a variety of matrices: soils, sludges, sediments, particulate-laden liquids and hard-to-oxidize samples. Compact and simple, the LSS Boat is designed to work with the Lotix making it the ultimate solution for all of your TOC needs with an economical price to fit any budget.