Catalog excerpts
TELEDYNE TEKMAR Automated Static and Dynamic Headspace Vial Sampler
Open the catalog to page 1To be successful in today’s laboratory, you have to do more with less – less time, less bench space and less budget for new instrumentation. Teledyne Tekmar can help you do just that with the HT3 Automated Static and Dynamic Headspace System. The HT3 combines Static and Dynamic Headspace analysis techniques into one easy-to-use unit, saving you time, bench space and money. Built on proven static headspace technology, the HT3 provides the following added benefits: • Increased sensitivity from 50 to 100 times with the Dynamic Headspace option (dependant on compound) • Accurate and precise...
Open the catalog to page 2Advantages of the HT3 • Standard integrated 60-position autosampler with 10-position platen heater provides true walk away automation • Increased sensitivity from 50 to 100 times with the Dynamic Headspace option (compound dependent) • Removable sample path for trouble-free maintenance • High temperature capability to 300 °C expands range of applications • Inert sample pathway including transfer line, sample needle and loop provide superior analytical results by eliminating adsorption and reducing carryover • Automated Leak Check and Benchmark for quick troubleshooting • Automated method...
Open the catalog to page 3Automated Productivity How It Works The HT3 incorporates traditional static headspace with the option to perform dynamic headspace. In the static set up, a sample is placed in a vial and then delivered to the autosampler. Once in the autosampler, the vial is loaded into a platen for heating. Upon reaching the final heat time it is then mixed for a set period of time. Using an electronic Mass Flow Controller the static vial pressure is recorded and the sample is pressurized to a user-defined set point. Next, the sample is passed through a fixed volume loop to another user-defined final...
Open the catalog to page 4A. Autosampler - The HT3 comes with a standard integrated, 60-position autosampler, providing true walk away automation. Vials are automatically queued into the platen heater ensuring reproducible dwell times. Vials may also be added during a run to increase productivity. (Optional 9 or 12 mL vial inserts sold separately.) B. Valve/Loop System - The valve and loop sample method has proven to be the industry’s most reliable technique for headspace sampling. The entire sample path is SilcoNert® 1000 and can be heated up to 300 oC thus minimizing analyte carryover. Different loop sizes ranging...
Open the catalog to page 5VOC TekLink™ Fully Optimized User Interface VOC TekLink™ software allows the user to enter all analysis parameters and once activated, it will continuously monitor the system ensuring operating limits are not exceeded. VOC TekLink™ software is capable of performing useful diagnostics such as leak and benchmark tests for instrument validation. All instrument parameters, method scheduling and editing can be programmed. VOC TekLink™ provides pre-developed methods, allowing startup with little or no modifications and also contains optional 21 CFR Part 11 tools. Current conditions of the unit...
Open the catalog to page 6HT3 Specifications Autosampler Capacity Platen Heater Up to 10 vials simultaneously heated up to 300 °C via resistance heater settable in 1 °C increments with a +/- 0.1 °C uniformity Vial Size 9, 12, 22 mL vials accepted (9 mL and 12 mL vials require inserts) Sample Loop Ships with a 1 mL SilcoNert® 1000 coated loop. Other optional loop sizes include 100, 250, 300, and 500 µL as well as 2, 3, and 5 mL Sample Mixing Optimix system allows variable power settings from 1 to 10 Trap Heater Controlled temperature range up to 300 °C (Trapping Module only) Trap Size Sample Pathway SilcoNert® 1000...
Open the catalog to page 7TELEDYNE TEKMAR