Overview: The Atomx Automated VOC Sample Prep System integrates an autosampler and purge and trap into a single instrument for analyzing volatile organic compounds (VOCs) in soils and waters. It uniquely automates methanol extraction for high-level soils, aligning with USEPA Method 5035, and features a patent-pending Extractasol clean-up step to minimize carryover and cross-contamination.
Key Features:
  • Automation: Automates sample preparation and purge and trap steps, including vial handling, sample volume measurement, standard injections, dilutions, rinsing, purging, desorption, and baking.
  • Sample Handling: Supports both low-level and high-level solid samples, with methanol extraction and dilution capabilities.
  • Capacity: Features an 80-vial carousel for high throughput and reduced downtime.
  • Software: VOC TekLink™ software provides a user-friendly interface for entering analysis parameters, monitoring system operations, and performing diagnostics.
Applications and Industries: The system is applicable in environmental, food and beverage, pharmaceutical, flavor, fragrance, packaging, and petrochemical industries.
Methods Supported: Compatible with multiple USEPA methods (e.g., 502.1, 5035) and ASTM standards.
Technical Specifications:
  • Sample Capacity: 80 positions for 40 mL VOA vials.
  • Liquid Handling: Sample syringe dispenses 1-25 mL with <1% RSD precision.
  • Gas Handling: Electronic mass flow controller for flow rates between 5 mL/min to 500 mL/min.
  • Standard Injection: Three systems with capacities of 1, 2, 5, 10, and 20 µL increments.
Additional Features:
  • Automated Methanol Extraction: Reduces labor and matches manual extraction results.
  • Mass Flow Controller: Allows programmable flow control for optimized performance.
  • Diagnostics: Includes electronic leak check, benchmark test, and email alerts for schedule completion.
System Control and Compliance: Operates with VOC TekLink™ software, supporting 21 CFR Part 11 compliance tools.
Environmental Specifications: Operates between 10° to 30°C with corrosion-resistant components.
Accessories: Optional cryofocusing module, sparge vessel heater, vial cooling, and foam eliminator available.
See more

Catalog excerpts

Atomx Automated VOC Sample Prep System-1

Automated VOC Sample Prep System

 Open the catalog to page 1
Atomx Automated VOC Sample Prep System-2

The Atomx combines an Autosampler and Purge and Trap into a single instrument for the analysis of VOCs in soils and waters. This is the first of its kind and the only system that employs a unique methanol extraction automation feature for high level soils in accordance with USEPA Method 5035. Extractasol clean-up step (patent pending) in the Atomx reduces carryover or cross contamination normally associated with high-level water or soil analysis. Three independent injection systems for internal standards, surrogates, and matrix spikes provide the ability to select variable volumes without generating...

 Open the catalog to page 2
Atomx Automated VOC Sample Prep System-3

How It Works Atomx is a Purge and Trap (P&T) system, combining automation and concentration into one system for the analysis of solid and liquid samples. The system prepares P&T samples, including drinking water, wastewater, soils, and sludges. Atomx completely automates sample preparation and P&T steps including vial handling, sample volume measurement, standard injections, dilutions, rinsing, purging, desorption and baking. The Atomx single needle design transfers liquid sample aliquots from the vial to the sparger. Low-level solid samples are purged directly in the vial, transferring analytes...

 Open the catalog to page 3
Atomx Automated VOC Sample Prep System-4

80-Position Carousel Design for Optimal Throughput The carousel drive is an electronically controlled mechanism that positions sample vials for sampling. The carousel positions the sample vials by rotating the carousel and indexing. The carousel tray is removable from the drive assembly for easy vial loading. Mixer Assembly This assembly allows for a magnetic stir bar to be added to the sample vial of a solid sample. The Atomx has the ability to spin or agitate the vial when it is in the sampling station to mix the soil during the purge process, or to mix a soil for Methanol Extraction. Vial...

 Open the catalog to page 4
Atomx Automated VOC Sample Prep System-5

Sample Transfer Line A patented 3-stage sample needle is used to displace the sample from the vial, purge in the vial and complete Methanol extractions. Heated Sample Mount and Path The heated sample mount on the Atomx is programmable up to 100 oC and the sample path is programmable up to 250 oC. The entire gaseous sample pathway uses SilcoNert® 2000-coated surfaces and Inertium® tubing. Glassware 5 mL or 25 mL fritted, or fritless glassware available. Unit ships with 5 mL fritted glassware. Extractasol (patent pending)– Potential carryover can be reduced by as much as 50% with the dedicated...

 Open the catalog to page 5
Atomx Automated VOC Sample Prep System-6

VOC TekLink™ Fully Optimized User Interface VOC TekLink™ software allows the user to enter all analysis parameters and once activated, it will continuously monitor the system ensuring operating limits are not exceeded. VOC TekLink™ software is capable of performing useful diagnostics such as leak and benchmark tests for instrument validation. All instrument parameters, method scheduling and editing can be programmed. VOC TekLink™ provides pre-developed methods, allowing startup with little or no modifications and also contains an optional 21 CFR Part 11 data audit trail package. The Status section...

 Open the catalog to page 6
Atomx Automated VOC Sample Prep System-7

Atomx Specifications Automation Sample Capacity: Vial Size: Nominal 40 mL capacity, single hole cap with Teflon®-faced silicone septum, per EPA specifications; 3 3/4“ high without cap and septum; 1 1/16” OD; 24 mm ID cap for water sampling. Liquid Handling Sample Liquid Handling: Sample syringe (27 mL) dispenses variable volumes of water from 1 - 25 mL in 1 mL increments. Sample Precision: < 1% RSD (n=7 @ 5 mL delivery volume measured by weight) Glass, PEEK™ Inertium® and Teflon® for syringe handling. 1/16” OD PEEK™ tubing for liquid transfer The entire liquid pathway can be rinsed using a combination...

 Open the catalog to page 7
Atomx Automated VOC Sample Prep System-8

System Control Instrument Control: VOC TekLink™ software in a Windows® XP or greater environment. Via RS-232 or USB converter (optional). Method Scheduling: All method types can be run from any position in the sample sequence. Up to three standards can be added to any user-specified position. Multiple runs can be made from the same vial (not recommended). System History: The system records a complete history of all sample, schedule and method information. 21 CFR Part 11 Compliance Tools: VOC TekLink™ can be configured to allow for full 21 CFR Part 11 compliance tools to be available to the end...