Mercury Analyzers comparison

Mercury Analyzers comparison

Mercury Analyzers comparison

Product catalog summary
Overview of Analytical Techniques:
1. Chemical Reduction followed by Cold Vapor Atomic Absorption (CVAA): Suitable for aqueous samples with mercury (Hg) measured from low ng/L to mg/L levels.
2. Chemical Reduction with or without Hg Pre-concentration followed by Cold Vapor Atomic Fluorescence (CVAF): Ideal for aqueous samples requiring the lowest detection levels, measuring Hg from sub ng/L to high µg/L levels.
3. Thermal Decomposition followed by Amalgamation and Atomic Absorption: Used for solid and semi-solid matrices, allowing analysis without sample preparation.
Instrument Detection Limits:
- CVAA: ≤ 5 ng/L
- CVAF with Pre-concentration: ≤ 0.5 ng/L
- Thermal Decomposition: ≤ 0.2 ng
Short Term Precision:
- CVAA: ≤ 3.3% RSD at 20 ng/L
- CVAF with Pre-concentration: ≤ 1.2% RSD at 20 ng/L
- Thermal Decomposition: ≤ 4% RSD at 10 ng
Usable Range:
- CVAA: 5 ng/L – 1 mg/L
- CVAF: <0.05 ng/L – 400 µg/L
- Thermal Decomposition: 0.5 ng – 1500 ng; <25,000 ng with extended range option
Key Capabilities of Mercury Analyzers:
- Hydra IIAA, QuickTrace M-7600, QuickTrace M-8000, Hydra IIC: High-performance gas/liquid separator, priority samples, built-in high concentration protection system, flow-through rinse to minimize carryover, counter-flow Nafion® membrane dryer.
- Hydra IIC: Can be reconfigured for liquids analysis, ultra-trace analysis mode with gold amalgamation for lower detection levels.
Method Selection Guide:
- Various EPA, ASTM, EN, and ISO methods are supported by different analyzers, with specific methods applicable to each model.
See more

Catalog excerpts

Mercury Analyzers comparison-1

Mercury Analayzers At A Glance OVERVIEW Hydra IIAA Principle Analytical Chemical reduction Technique followed by cold vapor atomic absorption (CVAA) Chemical reduction followed by cold vapor atomic absorption (CVAA) Chemical reduction with or without Hg pre-concentration followed by cold vapor atomic fluorescence (CVAF) Thermal decomposition followed by amalgamation and atomic absorption Sample Matrix • Aqueous samples • Aqueous samples • Hg measured from • Hg measured from low ng/L to mg/L levels sub ng/L to high µg/L levels • Aqueous samples • When the lowest detection levels are required • Solid, semi-solid matrices • Analysis without sample preparation Instrument Detection Limit Short Term Precision (% RSD @ 95% Confidence, N=5) Usable Range 0.5 ng – 1500 ng; <25,000 ng with extended range option

 Open the catalog to page 1
Mercury Analyzers comparison-2

Dual cell detection dynamic range protection system Flow through ultra-trace levels Nafion® membrane detector cells liquids analysis solids analysis Ultra-trace analysis mode with gold lower detection Manual Operation ^Bg^TELEDYNE LEEMAN LABS