1. Catalogs
  2. Teledyne LeCroy Europe
  3. T3DSO1000 Data Sheet Oscilloscopes

T3DSO1000 Data Sheet Oscilloscopes
1 /16Pages

T3DSO1000 Data Sheet Oscilloscopes

T3DSO1000 Data Sheet Oscilloscopes
1 /16Pages

Catalog excerpts

T3DSO1000 Data Sheet Oscilloscopes-1

TELEDYNE TEST TOOLS Front panel of the four channel series Tools for Improved Debugging • Long Capture - 7 Mpts/Ch and 14 Mpts interleaved. • Math and Measure - 7 basic math functions plus FFT and 38 automatic measurement parameters. • Connectivity - USB for mass storage, printing and PC control, plus LAN for fast data transfer. • Serial Bus Trigger and Decode - I2C, SPI, UART, RS232, CAN, LIN. • Waveform Sequence Recorder - record and play back up to 80,000 waveforms. • Optional MSO - 16 Digital Channels (4 channel series only). © Capture more time and show more waveform detail. © Extract results from waveforms and measurements. © Save data for external analysis and screen images for reports. © Debug serial buses directly in your Oscilloscope. © Replay the changing waveform history. © Add mixed signal debugging to your Oscilloscope. Memory up to 7 Mpts/Ch (14 Mpts interleaved) Sample Rate up to 500 MS/s / 1 GS/s interleaved Connectivity USB Host, USB Device, LAN

 Open the catalog to page 1
T3DSO1000 Data Sheet Oscilloscopes-2

PRODUCT OVERVIEW T3DSO1102: 2 Channel 100 MHz T3DSO1104: 4 Channel 100 MHz T3DSO1204: 4 Channel 200 MHz Teledyne Test Tools new T3DSO1000 Oscilloscopes feature two channel and four channel models. The two channel model is available with 100 MHz analog bandwidth, a single ADC with a 1 GSa/s maximum sample rate, and a single memory module with 14 Mpts of sample memory. The four channel scope is available in 100 and 200 MHz models and incorporates two 1 GSa/s ADCs and two 14 Mpts memory modules. When all channels are enabled, each channel has sample rate of 500 MSa/s and a standard record length...

 Open the catalog to page 2
T3DSO1000 Data Sheet Oscilloscopes-3

• Bode plot (four channel series only) • Search and navigate (four channel series only) • USB AWG module (four channel series only, option) • USB WIFI adapter (four channel series only, option) • Web Browser based control (four channel series only) • Large 7 inch TFT-LCD display with 800 1 480 resolution • Multiple interface types: USB Host, USB Device (USB-TMC), LAN, Trigger Out • Supports SCPI remote control commands • Supports Multi-language display and embedded online help Automatic measurement function for 38 parameters as well as Measurement Statistics, Zoom, Gating, Math, History and Reference...

 Open the catalog to page 3
T3DSO1000 Data Sheet Oscilloscopes-4

FUNCTION & CHARACTERISTICS 7 inch TFT-LCD display and 10 one-button menus When all channels are enabled, each channel has a maximum sample rate of 500 MSa/s. When a single channel per pair is active, that channel has sample rate of 1 GSa/s Front panel of the four channel series Front panel of the two channel series ● ● 7-inch TFT-LCD display with 800 * 480 resolution Most commonly used functions are accessible using 10 different one-button operation keys: Auto Setup, Default, Cursor, Measure, Roll, History, Persist, Clear Sweep, Zoom, Print Record Length of Up to 14 Mpts (single channel/pair...

 Open the catalog to page 4
T3DSO1000 Data Sheet Oscilloscopes-5

256-Level Intensity Grading and Color Temperature Display History Waveforms (History) Mode and Segmented Acquisition (Sequence) New display technology provides for fast refresh rates. The resulting intensity-graded trace is brighter for events that occur with more frequency and dims when the events occur with less frequency. Playback the latest triggered events using the history function. Segmented memory collection will store trigger events into multiple (Up to 80,000) memory segments, each segment will store triggered waveforms and timestamp each frame. True measurement to 14 M points The color...

 Open the catalog to page 5
T3DSO1000 Data Sheet Oscilloscopes-6

FUNCTION & CHARACTERISTICS 1 M points FFT 16 Digital Channels/MSO (four channel series only, option) The new math co-processor enables FFT analysis of incoming signals using up to 1 M samples per waveform. This provides high frequency resolution with a fast refresh rate. The FFT function also supports a variety of window functions so that it can adapt to different spectrum measurement needs. 16 digital channels enables users to acquire and trigger on the waveforms then analyze the pattern, simultaneously with one instrument. Color coded logic levels clearly differentiate high and low states....

 Open the catalog to page 6
T3DSO1000 Data Sheet Oscilloscopes-7

Bode Plot (four channel series only) USB 25 MHz AWG Module (four channel series only, option) The four channel series supports an optional USB 25 MHz function/arbitrary waveform generator that is operated from the USB host connection. Functions include Sine, Square, Ramp, Pulse, Noise, DC and 45 built-in waveforms. The arbitrary waveforms can be accessed and edited by the Teledyne Test Tools PC software. Complete Connectivity T3DSO1000 can control the USB AWG module, control an independent Teledyne Test Tools T3AFG instrument, scan an object’s amplitude and phase frequency response, and display...

 Open the catalog to page 7
T3DSO1000 Data Sheet Oscilloscopes-8

SPECIFICATIONS Acquire System Sampling Rate Memory Depth Peak Detect Average Eres Waveform interpolation 1 GSa/s (single channel/pair), 500 MSa/s (two channels/pair) Max 14 Mpts/Ch (single channel/pair), 7 Mpts/Ch (two channels/pair) 2 nsec (Four channel series) 4 nsec (Two channel series) Averages: 4, 16, 32, 64, 128, 256, 512, 1024 Enhance bits: 0.5, 1.5, 2, 2.5, 3; Selectable Sin(x)/x, Linear Input Channels Coupling Impedance Max.Input voltage CH to CH Isolation Probe attenuation 4 (Four channel series) 2+EXT (Two channel series) DC, AC, GND DC: (1 MΩ ± 2 %) || (15 pF ± 2 pF) (Four channel...

 Open the catalog to page 8
T3DSO1000 Data Sheet Oscilloscopes-9

Trigger System Trigger Mode Trigger Level Holdoff Range Trigger Coupling Coupling Frequency Response Coupling Frequency Response components below 10 KHz Trigger Sensitivity Trigger Jitter Trigger Displacement Auto, Normal, Single Internal: ± 4.5 div from the center of the screen EXT: ± 0.6 V (Two channel series) EXT/5: ± 3 V (Two channel series) 80 ns – 1.5 s AC DC LFRJ HFRJ Noise RJ DC: Passes all components of the signal AC: Blocks DC components and attenuates signals below 8 Hz LFRJ: Blocks the DC component and attenuates the low-frequency components below 2 MHz HFRJ: Attenuates the high-frequency...

 Open the catalog to page 9
T3DSO1000 Data Sheet Oscilloscopes-10

SPECIFICATIONS Window Trigger Window Type Source Absolute, Relative All channels Interval Trigger Slope Limit Range Source Time Range Resolution Rising, Falling <,>,<>,>< All channels 2 ns ~ 4.2 s 1 ns Dropout Trigger Timeout Type Source Slope Time Range Resolution Edge, State All channels Rising, Falling 2 ns ~ 4.2 s 1 ns Runt Trigger Polarity Limit Range Source Time Range Resolution +wid , -wid <,>,<>,>< All channels 2 ns ~ 4.2 s 1 ns Pattern Trigger Pattern Setting Logic Source Limit Range Time Range Resolution Invalid, Low, High AND, OR, NAND, NOR All channels <,>,<>,>< 2 ns ~ 4.2 s 1 ns...

 Open the catalog to page 10

All Teledyne LeCroy Europe catalogs and technical brochures

  1. T3DSO4000L-HD

    19  Pages

  2. HDO6000B

    20  Pages

  3. T3DAQ1-16

    16  Pages

  4. WavePro HD

    20  Pages

  5. WaveRunner 9000

    20  Pages

  6. DDR Test Suite

    12  Pages

  7. WaveRunner 8000

    24  Pages

  8. wavesurfer 510

    20  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.