SPARQ
1 /12Pages

SPARQ

SPARQ
1 /12Pages

Catalog excerpts

SPARQ-1

SPARQ™ Signal Integrity Network Analyzer High-bandwidth, Multi-port S-parameters

 Open the catalog to page 1
SPARQ-2

SPARQ: S-Para meters Quick SPARQ: S-Para meters Quick 2 Key Features • Provides S-parameter measurements up to four-ports • M easures from DC to 40 GHz • One-button-press internal OSLT calibration • Analyzes in both frequency and time domain • Produces mixed-mode and single-ended simulation-ready S-parameters • Includes advanced de-embedding features • Inherent TDR/TDT capability and preview modes for quick checks and debugging • Available at a fraction of the cost of other network analyzer solutions High-bandwidth, Multi-port S-parameters for the Masses S-parameter measurements are most often...

 Open the catalog to page 2
SPARQ-3

S-parameters, Quick VNA measurements begin with the unpleasant and complex task of calibration. This involves multiple connections that can produce misleading results due to operator error. The SPARQ provides calibrated measurements with a single connection to the DUT and offers simple setup choices. Start and complete the entire measurement with a single button press. I nternal Calibration SPARQ takes a revolutionary approach to calibration by building in calibration standards. This enables measurements to be made without multiple connection steps and removes the need for additional electronic...

 Open the catalog to page 3
SPARQ-4

THE SIGNAL THE SIGNAL IINNTTEEGGRRIITTYY TTOOOOLLSS yoyo uu eexxppeecctt 1. Differential- and common-mode step response at input and output ports 2. Mixed-mode return loss to 40 GHz 3. TDR traces shown during measurement 4. Differential- and common-mode insertion loss to 40 GHz 5. Mode conversion step responses 6. Differential- and common-mode impedance vs. electrical length 7. Rise time normalization for all time domain results 8. Up to 16 measurements can be displayed simultaneously 9. Independent zoom control over each trace 10. Smith chart display alone, or with individual S-parameter plots...

 Open the catalog to page 4
SPARQ-5

5 I ncludes the Tools That You Expect A signal integrity network analyzer should include well-integrated tools for providing measurement and analysis in both time and frequency domains. Signal integrity requires more than just S-parameters; the time domain offers important insight as it shows the performance of the S-parameter models in simulation. SPARQ includes standard all of the hardware and software tools needed to make signal integrity measurements right out of the box. These tools include capabilities that cost extra on most instruments. Mixed-mode S-parameter conversion and port renumbering,...

 Open the catalog to page 5
SPARQ-6

6 Streamlined Setup The simple setup shown in the main setup screen above is all that is needed to configure a SPARQ measurement. You provide the frequencies and number of ports and then go. Helper information like time length assist in frequency spacing choices, and DUT length mode choices control the pulser repetition rate for faster measurements. Various measurement sequence control modes allow for trade-offs between precision and speed, and helper information provides an estimate of the measurement time. All measurements proceed automatically without user intervention. Advanced screens are...

 Open the catalog to page 6
SPARQ-7

7 From Measurement Directly to Simulation S-parameters present many difficulties for time domain simulators. These difficulties come from the two ends of the frequency spectrum. Lack of a DC point and truncation of the high frequency content causes simulation problems. Since it is based in the time domain, SPARQ provides a DC measurement point and 40 GHz frequency content so that simulators come up with the right answer. SPARQ provides enforcements of passivity, causality and reciprocity to ensure physical measurement results and provides time domain views so that time domain behavior is verified...

 Open the catalog to page 7
SPARQ-8

ME ASUREME MEASUREMENNTTSS RRIIGGHHTT TTHHEE FFIIRRSSTT TTIMEIME 8 Advanced Features that Prevent Mistakes and Wasted Time A frustrating situation is to find that after spending the time to calibrate and take S-parameter measurements, something is wrong either because of a mistake or a poor connection. Sometimes it is hours or days before the problem is detected; that is hours or days of suspect data in use. Because SPARQ is TDR/TDT based, it can be used to provide basic troubleshooting before you get too far into the measurement. By driving the SPARQ in its native TDR/TDT mode, engineers can...

 Open the catalog to page 8
SPARQ-9

Design Val Design Valiidadattiioonn ttoo CCoommplpliiananccee tteessttiingng 9 SPARQ can perform all tests currently made with TDR or VNA instruments—only easier. Some tests that SPARQ performs include: • Impedance • Return Loss • Impedance Imbalance • Insertion Loss • Crosstalk (near- and far-end) • Differential- to common-mode conversion • Common- to differential-mode conversion • Intra-pair skew • Voltage transfer functions All measurements can be made in differential-mode, common-mode or single-ended, as applicable. Printed Circuit Board Testing Specifications for printed circuit boards are...

 Open the catalog to page 9
SPARQ-10

sp ecifications specifications Model Specific Specifications Model 4004E 4002E 4002M Ports 4 2 2 Calibration Internal, Automatic Internal, Automatic Manual Ask about 8 and 12 port instruments, available soon Operating Frequency DC to 40 GHz S-parameter Measurements Single-ended and mixed mode (calculated) Calibration Method OSLT Connector Type 2.92 mm Standard Measurement Capability Frequency Domain Displays Magnitude, Phase, Real and Imaginary Time Domain Displays Impulse Response, Step Response, Rho, Z – normalized to specified rise time Result Displays Up to 16 measurements displayed simultaneously...

 Open the catalog to page 10
SPARQ-11

OORRDDEERRIINNGG IINNFFOORRMMAATITIOONN Product Description Product Code Signal Integrity Network Analyzers 40 GHz, 4-port, Internal Calibration, SPARQ-4004E Signal Integrity Network Analyzer 40 GHz, 2-port, Internal Calibration, SPARQ-4002E Signal Integrity Network Analyzer 40 GHz, 2-port, Manual Calibration, SPARQ-4002M Signal Integrity Network Analyzer Options and Accessories 2 x 40 GHz Cables SPARQ-C402 4 x 40 GHz Cables SPARQ-C404 Manual Calibration Kit SPARQ-OSLT Soft Carrying Case SPARQ-SFTC Included with Standard Configuration 2 or 4 Color-coded, serialized, calibrated cables (depending...

 Open the catalog to page 11

All Teledyne LeCroy Europe catalogs and technical brochures

  1. T3DSO4000L-HD

    19  Pages

  2. HDO6000B

    20  Pages

  3. T3DAQ1-16

    16  Pages

  4. WavePro HD

    20  Pages

  5. WaveRunner 9000

    20  Pages

  6. DDR Test Suite

    12  Pages

  7. WaveRunner 8000

    24  Pages

  8. wavesurfer 510

    20  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.