1. Catalogs
  2. Teledyne Isco
  3. 260HL Hazardous Location Syringe Pump

260HL Hazardous Location Syringe Pump

260HL Hazardous Location Syringe Pump

260HL Hazardous Location Syringe Pump

Product catalog summary
Standard Controller Features
  • Operating Modes: Supports constant or continuous flow/pressure with up to three pumps.
  • External Interface: Includes RS232 serial interface, analog voltage inputs, and digital inputs/outputs.
Chemical Compatibility

The wetted materials are compatible with various fluids including ethane, butane, diesel fuel, isopropanol, benzene, and more.

260HL Hazardous Location Syringe Pump

Approved for use in hazardous environments (Class I, Div 2 or Zone 2). Features brushless DC motors to eliminate the need for additional safety devices. Compatible with a wide range of liquids and gases.

Applications
  • Used in metering and dispensing in explosive conditions.
  • Precision fluid addition in research and manufacturing.
  • Reactant feed in chemical process development.
260 HL Options & Accessories
  • Corrosion-resistant cylinders (Hastelloy 276 or Titanium).
  • Temperature control jacket for liquefied gas.
  • High-temperature and high-precision pressure transducers.
  • LabVIEW drivers and 4-20 mA input for control.
Specifications
  • Capacity: 266 mL
  • Flow Range: 0.001 – 64 ml/min
  • Flow Accuracy: ± 0.5% of setpoint
  • Pressure Range: 517.1 bar (10 - 7,500 psi)
  • Standard Pressure Accuracy: 0.5% FS
  • Optional Pressure Accuracy: 0.1% FS
  • Wetted Materials: Nitronic 50, Graphite filled PTFE, TFE, Hastelloy C-276, inert polymers, Titanium, Gold
  • Operating Temperature: 5 - 40° C Ambient
  • Power Required: 100 Vac, 117 Vac, 234 Vac, 50/60 Hz
  • Dimensions: 101cm x 27cm x 47cm
  • Weight: Pump unit – 32.8 kg, Controller – 2.96 kg
  • Standards Conformity: UL 1604, CSA-213, EN50021, EN61326:1998, EN61010-1, UL6101A-1, CSA1010
See more

Catalog excerpts

260HL Hazardous Location Syringe Pump-1

Standard Controller Features Operating Modes . Constant flow or pressure with up to three pumps . Continuous flow or pressure with dual pump External Interface . RS232 serial interface . Analog voltage inputs . Digital inputs and outputs Chemical Compatibility Wetted materials have been tested for compatibility with the following fluids. . ethane . butane . diesel fuel . isopropanol . benzene . ethelyne . butadiene . jet fuel . acetone . ethylbenzene . propane . hexane . ammonia . acetonitrile . propylene . gasoline . methanol . heptane 260HL Hazardous Location Syringe Pump NEW! APPROVED BY UL/DEMKO FOR CLASS I, DIV 2 OR ZONE 2 ENVIRONMENTS Teledyne Isco HL-Series pumps give you the same accurate, predictable flow and pressure control as our standard D-Series, while conforming to safety standards for use in UL Class I, Division 2, Groups A B C & D, T3B or Zone 2, II 3G EEx nCIIC T3 environments. The Hazardous location rating is achieved through internal design modifications including the use of brushless DC motors. This approach eliminates the need for purge boxes or other additional safety devices. The HL controller has a keypad and LCD, as well as built-in and optional interfaces for computer control and other devices. Programming is easy and flexible, with instant access to menu screens even when the pump is running. This allows you to change operating parameters on the fly. Wetted materials are compatible with most aqueous and organic liquids, corrosive solutions, heated fluids, liquefied gases, viscous fluids, or slurries and pastes. If needed, optional materials such as Hastelloy C-276 are available on request. Applications . Metering and dispensing in experiments and pilot plants where explosive conditions may occur . Precision fluid addition in research and manufacturing processes . Reactant feed in chemical process development, catalyst evaluation, plastic formulation . Accurate metering of liquefied gases PRODUCT DATA Telledyne IIsco Syrriinge Pumps Teledyne Isco, Inc., 4700 Superior Street, Lincoln, NE 68504 USA Toll free: (800) 228-4373 (USA and Canada) • Phone: (402) 464-0231 • Fax: (402) 465-3022 e-mail: [email protected] • www.isco.com

 Open the catalog to page 1

All Teledyne Isco catalogs and technical brochures

  1. ReaXus 6010R

    2  Pages

  2. L1150 AQ700

    2  Pages

  3. ELSD

    4  Pages

  4. Combiflash

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.