1000HLf  Hazardous Location Syringe Pump - 1

Teledyne Isco Syringe Pumps Product Data Sheet 1000HLf Hazardous Location Syringe Pump Approved by UL for Class I, Div 2 Environments The Teledyne Isco HLf-Series pumps give you the same accurate, predictable flow and pressure control as our standard D-Series, while conforming to safety standards for use in UL Class I, Division 2, Groups A B C & D, T4 environments. The hazardous location rating is achieved through internal design modifications including the use of brushless DC motors. This approach eliminates the need for purge boxes or other additional safety devices. Wetted materials are compatible with most aqueous and organic liquids, corrosive solutions, heated fluids, liquefied gases, viscous fluids, or slurries and pastes. If needed, optional materials such as Hastelloy C-276 are available on request. The HLf controller has a keypad and LCD, as well as built-in and optional interfaces for computer control and other devices. Programming is easy and flexible, with instant access to menu screens even when the pump is running. This allows you to change operating parameters on the fly. Applications hhMetering and dispensing in experiments and pilot plants where explosive conditions may occur hhPrecision fluid addition in research and manufacturing processes hhReactant feed in chemical process development, catalyst evaluation, plastic formulation hhAccurate metering of liquefied gases Standard Controller Features Operating Modes hhConstant flow or pressure with up to four pumps hhContinuous flow or pressure with dual pump hhFlow or pressure programming with single pump hhDispense mode External Interface hhRS232/485 serial interface hhAnalog voltage inputs hhDigital inputs and outputs hhEthernet/USB Chemical Compatibility Wetted materials have been tested for compatibility with the following fluids. hhethane hhbutane hhdiesel fuel hhisopropanol hhbenzene hhethelyne hhbutadiene hhjet fuel hhacetone hhethylbenzene hhpropane hhhexane hhammonia hhacetonitrile hhpropylene hhgasoline hhmethanol hhheptane hhisobutylene hhmethane 4700 Superior Street, Lincoln, NE 68504 USA • Phone: (800) 228-4373 / (402) 464-0231 • iscoinfo@teledyne.com / www.isco.co

Open the catalog to page 1
1000HLf  Hazardous Location Syringe Pump - 2

1000HLf Specifications Pump Module: Capacity: Flow Range: Flow Accuracy: ±0.5% of setpoint, (Maximum 1.5 μL/min seal leakage) Displacement Resolution: Motor Stability: Pressure Range: Standard Pressure Accuracy: Optional Pressure Accuracy: Wetted Materials (Standard): Nitronic 50, Graphite filled PTFE, TFE, Hastelloy C-276, Inert Polymers, Titanium, Gold Plumbing Ports: Operating Temperature: Power Required: Pump unit - 103 x 27 x 47 cm (40.6 x 10.7 x 18.5 inches) Pump unit - 36.5 kg, Controller - 2.96 kg Standards Conformity: Teledyne Isco is continually improving its products and reserves...

Open the catalog to page 2

All Teledyne Isco catalogs and technical brochures

  1. ReaXus 6010R

    2 Pages

  2. L1150 AQ700

    2 Pages

  3. ELSD

    4 Pages

  4. Combiflash

    2 Pages

Archived catalogs