Features
  • Provides a ±15 Volt, 250 mA transducer supply.
  • Includes analog signals available on the back panel and RS232/485 interface.
  • Labview drivers are available for integration.
  • Offers high accuracy with ±0.2% FS ± 1 digit and CE compliance.
  • Equipped with high and low alarms programmable from the front panel.
  • Features a fully integrated 4.5 digit display with a resolution of 1 part out of ±99,999.
  • Supports independent channel programmability and radio control.
  • Comes with user-selectable AC input and an anodized aluminum case for bench top use or panel mounting.
  • Analog I/O options include 0-5 VDC, 0-10 VDC, and 4-20 mA.
Applications
  • Secondary calibration reference.
  • Fuel cell testing.
  • Pharmaceutical production.
  • Gas generators and thin film processes.
  • Gas blending.
Description

The THPS-400 is a digital 4-channel power supply designed to replace older analog models while maintaining a half-rack profile for easy replacement. It includes features such as an integrated totalizer, user-selected filtering, RS-232/485 interfaces, and simultaneous display of all four channels. The unit is housed in an all-metal enclosure with a bright vacuum fluorescent display.

The digital design allows users to set minimum and maximum display values for specific voltage or current inputs, eliminating the need to access hard-to-reach transducers for re-zeroing. This feature is particularly beneficial for OEM manufacturers.

Specifications
  • Accuracy: ±0.2% of F.S.
  • Repeatability: ±0.1% of F.S.
  • Input: 90-125/180-250 VAC 50/60 Hz.
  • Output: ±15 VDC, 250 mAmps Max.
  • Operating Temperature: -20° to 70°C.
  • Zero Temperature Coefficient: < 0.05% /°C of F.S.
  • Standard Operating Ranges: 0-5 VDC, 0-10 VDC, 250 mA max. per channel.
  • Connector Types: 15-pin sub-D for transducers, 9-pin sub-D for serial, 26-pin HD-D for alarm, 9-pin HD-D for analog out.
  • Weight: 5 Lb.

Includes a NIST Traceable Calibration Certificate and offers both analog and digital output options. The device supports six standard gases, with custom gases available upon request.

See more

Catalog excerpts

THPS-400-1

TELEDYNE HASTINGS INSTRUMENTS PowerPod-400 ± 15 Volt, 250 mA Transducer Supply Analog signals available on back panel RS232/485 Interface Labview Drivers available High Accuracy High and Low alarms programmed from front panel Display can be zeroed from front panel or remotely Fully Integrated 4.5 digit display Resolution: 1 part out of ± 99,999 Four channels of simultaneous display Independent Channel Programmability Radio Control Totalizer Set Points User Selectable AC Input Anodized aluminum case for bench top use or panel mounting Secondary calibration reference Fuel cell testing Pharmaceutical production Gas generators Thin Film processes Gas Blending DESCRIPTION The THPS-400 is Hastings Instruments’ newest power supply. This digital 4-channel instrument replaces the old analog version of the Model 400 and 200 but maintains the half-rack profile for simple drop-in replacement of older units. The new THPS-400 incorporates, as standard, features that were optional or non-existent on the previous models. These new THPS-400 features include integrated totalizer with count-up or countdown options, user selected filtering of readings, RS-232, RS-485, simultaneous display of all four channels or selective blanking of unused channels, and ratio control with analog outputs for stacking multiple power supplies. There is also an easy-to-follow, menu-driven calibration and setup. The unit is packaged in an all-metal enclosure with a bright vacuum fluorescent display. The digital design of the THPS-400 allows the user to set both the minimum and maximum display values corresponding to specific voltage or current inputs. One advantage of this approach is that it negates the need to access hard to reach transducers to re-zero them. Should the analog signal from the transducer change due to zero shift, the digital counts seen by the THPS-400 can be changed to display zero either manually from the front panel or via serial communication with the unit. This should be especially attractive to OEM manufacturers.

 Open the catalog to page 1
THPS-400-2

Specifications and Standards DESCRIPTION (Cont) Full scale display values are also easy to change to accommodate transducers with various ranges. Changing the full scale display is so easy, in fact, that it almost makes a scaling factor unnecessary. Still, at the request of many of our customers, a scaling factor feature has been included and can be applied to each individual channel. SPECIFICATIONS Accuracy Operating Temperature Zero Temperature Coefficient < 0.05% /°C of F.S. Standard Operating Ranges Connector Types 15-pin sub-D, Transducers 9-pin sub-D, Serial 26-pin HD- D, Alarm 9-pin HD-D,...

 Open the catalog to page 2
THPS-400-4

Cable PINOUT Edge Connecto "H" Pinout 15 pin D connector 15 pin D connector 15 pin D connector 15 pin D connector Edge Connector 20 Position 20 Position 20 Position 20 Position *Factory Settings Input Voltage Display Zero Display 100% High Set Point Low Set Point A/D Conversion Rate Multiplication Factor Analog I/O VAC SCCM SCCM High flow alarm Low flow alarm Hz Applied to reading VDC *Specify custom set-up when ordering. Telephone: (757) 723-6531 Toll Free: (800) 950-2468 Fax: (757) 723-3925 World Wide Web: http://www.teledyne-hi.com E-mail: [email protected] P.O. Box 1436 Hampton,...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.