The Teledyne Hastings NALL-Series Mass Flowmeters are designed for precise gas flow measurement, utilizing a unique thermal principle for linear response. These flowmeters are ideal for high-temperature and adverse environmental conditions due to their all-metal construction and separation of the transducer from electronic circuitry.
Design Features
  • Original model with over 50 years of customer satisfaction.
  • Operates on a thermal principle for linear flow response.
  • Constructed entirely of metal, with no O-rings.
  • Separate transducer and electronic circuitry.
  • Bright digital display in cabinet or panel mountable case.
Specifications
  • Accuracy: ±1% of range for 20% variation in pressure and temperature.
  • Linearity: ±1% of range.
  • Repeatability: ±0.25% of full scale.
  • Working Pressure: 500 psig max; options for 1000 psig.
  • Temperature: Gas temperature from -20°C to 100°C; ambient temperature for transducer and indicator from 0°C to 40°C.
  • Power: 115 volt AC, 50/60 Hz - 15 watt; 230 volt option available.
  • Output: 0-5 volts DC into a load of 2000 ohms or greater.
  • Material: 316SS or monel, gold/nickel braze.
  • Calibration: Direct reading for air; calibration available for other gases.
  • Response Time: Less than 10 seconds to 67% of reading.
Applications
  • Vapor deposition.
  • Leak testing.
  • Gas blending.
  • Semiconductor processes.
  • Measurement of toxic and hazardous gases.
  • High temperature, high pressure, and hermetically sealed uses.
Options
  • OPT-T: High temperature transducer up to 200°C.
  • OPT-P: Working pressures up to 1000 psig.
  • OPT-H: Hermetically sealed transducer.
  • OPT-NIST: NIST traceability.
  • OPT-E: 230 VAC/50 Hz.
Selection Chart
  • Power Supplies: Various models available with different dimensions and outputs.
  • Cables: Available in lengths from 8 to 300 feet, with high-temperature options.
  • Transducers: Stainless steel and monel options with various flow ranges and connections.
  • Laminar Flow Elements: Stainless steel and monel options for different flow ranges and connections.
See more

Catalog excerpts

Nall Mass Flow Meters-1

NALL SERIES Transducer Dimensions All dimensions shown are in inches Your Customer Service Representative .TELEDYNE INSTRUMENTS Hastings Instruments ATeledyne Technologies Company PB 155-5/03 ©Teledyne 2003. All Rights Reserved. World Wide Web: http://www.teledyneinstruments.com E-mail: [email protected] TELEDYNE HASTINGS NALL-SERIES MASS FLOWMETERS Model NALL Mass Flowmeter • 15 Standard Flow Ranges MAXIMUM WORKING For All of Your Gas Flow Measuring Needs: • Vapor Deposition • Leak Testing • Semiconductor Processes Hazardous Gases • High Temperature, High Sealed Uses Model NALL Power Supply DESIGN FEATURES The Hastings NALL-Series Flowmeter is the original model that has generated repeat customers for over 50 years. It operates on a unique thermal principle which permits inherently linear response to flow. The flow passage is constructed entirely of metal and contains no O-rings. The transducer is also separate from any electronic circuitry, mak- ing it ideal for high temperature or other adverse environmental applications. The power supply features a bright digital dis- play and is contained in either a cabinet or panel mountable case. All models feature high accuracy and excel- lent repeatability. ^BB^TELEDYNE INSTRUMENTS A Teledyne Technologies Company

 Open the catalog to page 1
Nall Mass Flow Meters-2

NALL SERIES NALL SERIES SELECTION CHART Stainless Steel Laminar Flow Elements (to be used with Model HS-10S transducer) Each flowmeter requires one Power Supply; one Cable; one Transducer; and (if using a range higher than 0-50 slpm) one Laminar Flow Element. Digital Cabinet Digital Panel Digital Panel, 4.20 mA output Digital Panel Flowmeter/Totalizer Digital Chassis Flow Range Power Supplies *250 psig working pressure max. at 100°F. See ASA B16.5 100 lb. std. Monel Laminar Flow Elements (to be used with Model HS-10M transducer) Cables Stock # Transducer Cable Length 8-foot Flow Range Monel LFEs,...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.