Overview
Teledyne Hastings Instruments (THI) has over 60 years of experience in designing and manufacturing mass flow products. The 300 Series, featuring patented technologies, offers high-precision flowmeters and controllers.

Design Features
The 300 Series meters and controllers measure mass flow accurately without needing corrections for gas pressure and temperature, achieving an accuracy of ±1.0% of full scale. They require no periodic maintenance under normal conditions and can handle moderate overpressures. Calibration is typically done with standard gases, with special calibrations available upon request.

Key Features
  • Inherently linear design without linearization circuitry.
  • Field recalibration is straightforward, requiring only zero and span adjustments.
  • Linear output signal even with large overflows.
  • Removable and replaceable sensor module.
  • Fast settling times.

Models and Specifications
HFM-301
  • Accuracy: ±1.0% of F.S.
  • Settling Time: ≤ 0.4 sec
  • Power Requirement: ± 15 VDC @ ± 55 mA
  • Wetted Materials: Viton®, 316 SS, 302 SS, Nickel 200
  • Weight: 3.5 lb (0-300 slm), 3.4 lb (300-1000 slm)
HFC-303
  • Accuracy: ±1.0% of F.S.
  • Settling Time: ≤ 2.0 sec
  • Power Requirement: ± 15 VDC @ 150 mA
  • Wetted Materials: 302SS, 316L SS, Nickel 200, Viton, Teflon®, Kalrez®
  • Weight: 5.3 lb (0-300 slm), 5.2 lb (300-1000 slm)

Applications
  • Leak Testing
  • Research
  • Vapor Deposition
  • R&D and Process Flows
  • Semiconductor Processes
  • Pollution Monitoring
  • Gas Blending
  • Chromatography

Optional Features and Accessories
  • Fittings: VCR, VCO, Swagelok®
  • High pressure rating (1000 psig)
  • Cleaned for oxygen service
  • Power supplies with integral flow totalizers and alarm set points
  • Interconnecting cables

Ordering Information
Options for circuit board pinout, output type, fittings, seals, pressure, and calibration type are available. Typical configurations are provided for reference.

Contact Information
For customer service, contact Teledyne Hastings Instruments via phone, fax, or email. Visit their website for more information.
See more

Catalog excerpts

HFM-301 / HFC-303​-1

TELEDYNE HASTINGS MEDIUM CAPACITY FLOWMETERS AND CONTROLLERS INSTRUMENTS DESIGN FEATURES Teledyne Hastings Instruments (THI) products represent over 60 years of experience in the design and manufacture of mass flow products. The 300 Series is a culmination of this experience with patented technologies that make these the finest flowmeters and controllers available today. FEATURES • ±1.0% of Full-Scale Accuracy1 • Rapid Settling Times HFM-301 ≤ 0.4 sec HFC-303 ≤ 2.0 sec • Range — 25 to 1000 slm (Air Equivalent) • Operating Pressures to 500 PSI or Higher • NIST Traceable Calibration Leak Testing Research Vapor Deposition R&D and Process Flows Semiconductor Processes Pollution Monitoring Gas Blending Chromatography Power Supplies Available The THI Mass Flow 300 Series meters and controllers are designed to accurately measure mass flow without corrections or compensations for gas pressure and temperature. They are accurate to better than ±1.0% of full scale. THI mass flow instruments do not require any periodic maintenance under normal operating conditions with clean gases. No damage will occur from the use of moderate overpressures (~500 psi) or overflows. Instruments are normally calibrated with the appropriate standard calibration gas (air & N2), then a gas conversion factor (GCF) is used to adjust the output for the intended gas. Special calibrations for other gases, such as oxygen, helium and argon, are available upon special order. These products contain a number of features that set them apart from other available instruments: (1) They are inherently linear; no linearization circuitry is employed. Should recalibration in the field be desired (a calibration standard is required), the customer needs to simply set the zero and span points. (2) The output signal is linear for very large overflows and will not come back on scale when a flow an order of magnitude over the full scale flow rate is measured. (3) The instrument incorporates a removable/ replaceable sensor

 Open the catalog to page 1
HFM-301 / HFC-303​-2

Optional Features Fittings–VCR, VCO and Swagelok® High pressure rating (1000 psig) Cleaned for oxygen service Pressure Coefficient Leak Integrity *Note: After changing components, instruments require recalibration to meet accuracy specifications. High-Pressure Option Power Supplies with integral Flow Totalizers & Alarm Set Points Interconnecting cables Standard Operating Pressure Temperature Coefficient (zero) < 0.085%/°C of F.S. (0-60°C) Temperature Coefficient (span) < 0.11%/°C of reading (15-60°C) Standard Output Optional Output *Attitude Sensitivity of Zero *Attitude Sensitivity of Span SPECIFICATIONS...

 Open the catalog to page 2
HFM-301 / HFC-303​-3

All dimensions shown are in inches [mm].

 Open the catalog to page 3
HFM-301 / HFC-303​-4

MODELS HFM-301, HFC-303 Selection Chart Typical instrument ordering/options number: Model No. Circuit Board 1000 psi (1500 proof) Calibration Type NIST 5 Point (Standard) Circuit Board Range Information Viton (Standard) Flow Units __________________________ Standard Conditions* _________________ **3/4" Swagelok required for flows above 300 slm *Referenced to standard temperature and pressure (0°C and 760 Torr, respectively). Selection Chart Typical instrument ordering/options number: Model No. Circuit Board Calibration Type 1000 psi (1500 proof) Calibration Type Circuit Board NIST 5 Point (Standard)...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.