video corpo

HFM-210/HFC-212

HFM-210/HFC-212

HFM-210/HFC-212

Product catalog summary
Design Features
The Teledyne Hastings Instruments (THI) Model HFM Mass Flowmeter and HFC Mass Flow Controller are designed with over 55 years of experience, offering reliable and high-quality mass flow instruments. These models can measure and control various gases under different pressure conditions. The HFM-210 and HFC-212 models are specifically designed to match the footprint of the Tylan® FC-280 model, featuring a modular design with an insulated thermo electric sensor for enhanced zero stability and flexible installation. Key features include an integral filter, replaceable electronics card, and an externally adjustable valve with replaceable flow orifices.
Optional Features
Available options include different fittings, O-ring seals, enhanced response time, EMF stability, high-pressure rating up to 1000 psig, 4-20 mA converters, and cleaning for oxygen service. Accessories such as power supplies, readouts, and interconnecting cables are also available.
Applications
The HFM-210 and HFC-212 models are suitable for applications like leak testing, medical research, vapor deposition, R&D and process flows, semiconductor processes, pollution monitoring, gas blending, and chromatography.
Specifications
HFM-210: Accuracy of ±1% F.S., repeatability of ±0.05% F.S., standard pressure rating of 500 psig, and a high-pressure option proof tested to 1500 psig. It features a temperature coefficient of zero ±0.08% F.S./°C and span ±0.15% Rdg/°C, with input power of ±15 VDC at ±30 mA. The flow signal is 0-5.00 VDC or 4-20 mA, with wetted materials including 316 SS, Viton®, and 87/13 Au/Ni Braze.
HFC-212: Similar specifications to the HFM-210, with a control valve DP per customer request and input power of ±15 VDC at +60 mA/-185 mA. It also offers command input options of 0-5.00 VDC or 4-20 mA.
Ordering Information
Options for circuit boards, output types, fittings, O-rings, working pressure, and calibration types are available for customization. Standard conditions are referenced to 0°C and 760 Torr.
See more

Catalog excerpts

HFM-210/HFC-212-1

TELEDYNE HASTINGS INSTRUMENTS LOW CAPACITY FLOWMETERS AND CONTROLLERS DESIGN FEATURES The Teledyne Hastings Instruments (THI) Model HFM Mass Flowmeter and HFC Mass Flow Controller represent over 55 years of experience in designing and manufacturing reliable, high quality mass flow instruments. The HFM/HFC Series flow instrument can be set up to measure and control any of a wide variety of gases under many different pressure conditions as required. The Models HFM-210 and HFC-212 were specifically designed to meet the footprint dimensions of the Tylan® FC-280 model. They are based on refinements to the existing Hastings 200 Series flow product line. The HFM/HFC Series of flow instruments is based on a modular design. At the heart of each instrument is an insulated thermo electric sensor which provides enhanced zero stability and allows installation in any orientation plane in most low pressure applications. This sensor is designed to be removable/replaceable in the field to virtually eliminate long down time due to clogging. Additionally, the HFM/HFC design features an integral filter and an easily replace-able closed loop electronics card.* The HFC also features an externally adjustable valve with easily replaceable flow orifices. All of these standard features, when coupled with the instrument’s inherent linear response to flow changes and THI’s long-proven reputation for quality, result in the finest flowmeters and flow controllers available today. Optional Features Fittings O-ring seals Enhanced response time Enhanced EMF stability High pressure rating (1000 psig) 4-20 mA converters Cleaned for oxygen service Accessories Power supplies/readouts Interconnecting cables *Note: After changing components, instruments require recalibration to meet accuracy specifications. Models HFM-210, HFC-212 FEATURES • ±1% of Full-Scale Accuracy1 • Proven Reliability • Range — 10 to 30,000 sccm (N2 Equivalent) • NIST Traceable Calibration • Tylan® FC-280 Footprint APPLICATIONS • Leak Testing • Medical Research • Vapor Deposition • R&D and Process Flows • Semiconductor Processes • Pollution Monitoring • Gas Blending • Chromatography HFM-210 / HFC-212 Tylan® is a registered trademark of the Mykrolis Corporation.

 Open the catalog to page 1
HFM-210/HFC-212-2

MODELS HFM-210, HFC-212 SPECIFICATIONS HFM-210 Accuracy1 and Linearity ±1% F.S. Repeatability ±0.05% F.S. (max.) Standard Pressure Rating 500 psig Pressure Coefficient +0.0067% Rdg/psi (0-1000 psig N2) High-Pressure Option Proof tested to 1500 psig Leak Integrity < 1x10-9 sccs Temperature Coefficient Zero ±0.08% F.S./oC (max.) (0°-50°C) Span ±0.15% Rdg/oC (max.) STP 0°C and 760 Torr Input Power ±15 VDC at ±30 mA (max.) Flow Signal 0-5.00 VDC or 4-20 mA (inherently linear) Wetted Material 316 SS, Viton®, 87/13 Au/Ni Braze Connector 15 pin Subminiature D Fittings Available Swage (1/8”, 1/4” & 3/8”),...

 Open the catalog to page 2
HFM-210/HFC-212-3

Model HFM-210 MODELS HFM-210, HFC-212 Model HFC-212 All dimensions shown are in inches [mm].

 Open the catalog to page 3
HFM-210/HFC-212-4

Your Customer Service Representative Telephone: (757) 723-6531 Toll Free: (800) 950-2468 Fax: (757) 723-3925 World Wide Web: http://www.teledyne-hi.com E-mail: [email protected] P.O. Box 1436 Hampton, VA 23661 PB 162-10/05 © Teledyne 2005 All Rights Reserved. Typical instrument ordering/options number: Model No. Circuit Output Fittings O-Rings Working Calibration Board Pressure Type HFM-210 01 01 01 01 01 01 Order No. Options Circuit Board 01 Standard 02 Fast Response Output 01 0-5 Volts (Standard) 02 4-20mA Fittings 01 1/4” Swagelok (Standard) 02 1/8” Swagelok 03 VCR® 1/4” 04...