video corpo

HFM-201/HFC-203

HFM-201/HFC-203

HFM-201/HFC-203

Product catalog summary
Overview
The document provides detailed information about the Teledyne Hastings Instruments HFM-201 Flow Meter and HFC-203 Flow Controller, which are medium capacity flowmeters and controllers designed for applications such as leak testing, flame spray, and aerospace.

Features
  • Accuracy of ±1% full scale.
  • Input power options: ±15 VDC or +24 VDC.
  • Flow ranges from 0-25 slm to 0-1000 slm (N2 Equivalent).
  • NIST Traceable Calibration Certificate.

Benefits
  • Excellent stability and reliability.
  • Outstanding zero stability.

Description
The HFM/HFC series is based on a modular design with an insulated thermal transfer sensor for enhanced zero stability. The HFC model includes a two-stage, pilot-operated control valve.

Specifications and Standards
  • Accuracy: ±1.0% of full scale.
  • Repeatability: ±0.05% of full scale.
  • Maximum operating pressure: 500 psi.
  • Leak integrity: < 1x10-9 sccs He.
  • Temperature coefficients for zero and span provided.
  • Standard output: 0-5 VDC, optional 4-20 mA.

HFM-201 Specifics
  • Power requirements: ±(14-16) VDC @ ±30mA or (14-32) VDC.
  • Wetted materials include 302 SS, 316 SS, Nickel 200, Viton®.
  • Weight: 3.55 lb (1.61 kg) for 201L, 3.25 lb (1.47 kg) for 201H.
  • Optional features include various fittings and seals.

HFC-203 Specifics
  • Power requirements: ±(14-16) VDC @ +60/-185 mA or (14-32) VDC.
  • Wetted materials include 302 SS, 316 SS, Nickel 200, Viton®, Kalrez® (Valve Seat).
  • Weight: 5.20 lb (2.36 kg) for 203L, 5.15 lb (2.34 kg) for 203H.

Accessories
  • Power supplies with integral flow totalizer, alarm set points, and interconnecting cables.
  • THCD-100 Power Supply & Display available.

Compliance
  • EMC: EN 61326-1.

Contact Information
Teledyne Hastings Instruments, 804 Newcombe Ave., Hampton, VA 23669. Phone: (757) 723-6531, Toll-Free: (800) 950-2468, Fax: (757) 723-3925, Email: [email protected].
See more

Catalog excerpts

HFM-201/HFC-203-1

TELEDYNE HASTINGS INSTRUMENTSEvery whereyou look” HFM-201 Flow Meter HFC-203 Flow Controller MEDIUM CAPACITY FLOWMETERS AND CONTROLLERS FEATURES • ±1% full scale accuracy1 • Input Power +/- 15 VDC or +24 VDC (specify when ordering) • Available flow ranges: 0-25 slm to 0-1000 slm (N2 Equivalent) • NIST Traceable Calibration Certificate APPLICATIONS • Leak testing • Flame Spray • Aerospace • Excellent Stability • Proven Reliability • Outstanding Zero Stability The Teledyne Hastings Instruments (THI) Model HFM Mass Flowmeter and HFC Mass Flow Controller represent a culmination of over 65 years of experience in designing and manufacturing reliable, high quality mass flow instruments. The HFM/HFC Series of flow instruments is based on a modular design. At the heart of each instrument is an insulated thermal transfer sensor which provides enhanced zero stability. The HFC also features a two-stage, pilot-operated control valve. All of these standard features, when coupled with the instrument’s inherent linear response to flow changes and THI’s long-proven reputation for quality, result in the finest flowmeters and flow controllers available today. TELEDYNE HASTINGS INSTRUMENTS Everywhereyoulook"

 Open the catalog to page 1
HFM-201/HFC-203-2

1 See Product Manual for critical information on instrument accuracy and the use of GCFs (gas conversion factors). Stated accuracy is for nitrogen or other gas specific calibration and use with this gas only. Specifications and Standards Optional Features Fittings -VCR® VCO® NPT Swagelok®, Seals -Kalrez® Neoprene Buna-N Enhanced response time 4-20 mA converters Cleaned for Oxygen service Accessories Power Supplies available with: Integral Flow Totalizer Alarm Set Points Interconnecting cables THCD-100 Power Supply & Display Hastings Instruments reserves the right to change or modify the design...

 Open the catalog to page 2
HFM-201/HFC-203-7

www.teledyne-hi.com Model No. HFM-201 HFC-203 Circuit Board Circuit Board Working Pressure Range Information for all Instruments Each calibration will require the following information: Range Flow Units Gas For the HFC Instruments also Upstream Pressure (maximum & minimum) Downstream Pressure (maximum & minimum) Does the downstream pressure change with flowrate? Y/N For volumetric units the standard temperature and pressure of the unit is also required 0°C & 760 Torr will be used when other values are not specified TELEDYNE HASTINGS INSTRUMENTS A Teledyne Technologies Company PB-140B_12/16 ©...

 Open the catalog to page 7
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.