video corpo

HFM-200-LFE

HFM-200-LFE

HFM-200-LFE

Product catalog summary
Overview: The Teledyne Hastings Instruments Model HFM-200 with Laminar Flow Element (LFE) is a high-capacity mass flowmeter designed for accurate measurement of gas flows. It features a modular construction with a removable thermal transfer sensor, ensuring minimal downtime and easy maintenance.
Features:
  • Flow range from 25 to 15,000 standard liters per minute (slm) for nitrogen equivalent.
  • Accuracy of ±1% of full scale for nitrogen.
  • Modular design with NIST traceable calibration.
  • Low pressure drop and linear response to flow changes.
Applications:
  • Bulk gas handling.
  • Furnace fuel measurement.
  • Research and development, and process flows.
  • Leak testing.
Benefits:
  • High capacity and low pressure drop.
  • Linear response to flow changes.
Design Features: The HFM-200 with LFE offers fast and accurate measurement of high-capacity gas flows. It includes laminar flow elements available in various sizes, providing a typical pressure drop of 4.5 inches of water at full-scale flow.
Optional Features:
  • Various fittings including flanges, smooth pipe, and pipe thread.
  • Oxygen cleaning available.
Accessories: Power supplies and readouts with totalizer and alarms, and cables of various lengths.
Technical Specifications:
  • Accuracy: ±1% of full scale.
  • Repeatability: ±0.05% of full scale.
  • Pressure coefficient: 0.0067%/psi for nitrogen.
  • Leak integrity: < 1x10-7 sccs He.
  • Operating temperature: 0 - 60°C.
  • Temperature coefficient: Zero ±0.035% of full scale, Span ±0.05% of reading.
  • Power requirements: ±15 VDC or +24 VDC.
  • Available outputs: 0 - 5 VDC, 4 - 20 mA, 0 - 20 mA.
  • Wetted materials: 316SS, Viton®, 87/13 Au/Ni Braze.
  • Standard working pressure: 150 psig for flanged LFEs, 500 psig for threaded or smooth LFEs.
Notes: The accuracy is specific to nitrogen or other gas-specific calibration. Reduced accuracy is noted above 12,000 slm (N2 equivalent).
See more

Catalog excerpts

HFM-200-LFE-1

TELEDYNE HASTINGS INSTRUMENTS MODELS Range — 25 to 15,000 slm (N2 Equivalent) Accuracy in Nitrogen ±1% of Full scale Modular Construction NIST Traceable Calibration Bulk Gas Handling Furnace Fuel Measurement R&D and Process Flows Leak Testing High Capacity Low Pressure Drop Linear Response Model HFM-200 with LFE (Laminar Flow Element) DESCRIPTION The Teledyne Hastings Instruments (THI) Model HFM Mass Flowmeter represents a culmination of years of experience in designing and manufacturing reliable, high quality mass flow instruments. The HFM Series of flow instruments is based on a modular design. At the heart of each instrument is an insulated thermal transfer sensor which provides enhanced zero stability. This sensor is designed to be removable/ replaceable in the field to virtually eliminate long down time due to clogging. Additionally, the HFM design features an integral filter and an easily replaceable closed loop electronics card.* All of these standard features, when coupled with the instrument’s inherent linear response to flow changes and THIs’ long-proven reputation for quality, result in the finest flow meters and flow controllers available today.

 Open the catalog to page 1
HFM-200-LFE-2

Design Features (cont) Model HFM-200 with LFE THI Model HFM-200 Flow meters with an LFE offer the user fast, accurate measurement of highcapacity gas flows, in ranges from 25 slm up to 15,000 slm. Our laminar flow elements (smoothed, flanged, or threaded) are available in various sizes from 3/8” up to 8”. Used in conjunction with the standard HFM Flowmeter, these units can be combined to measure full-scale flows up to 15,000 slm and feature a typical pressure drop of 4.5” of water at full-scale flow. Fittings – Flanges, Smooth Pipe, Pipe Thread Oxygen Cleaning Accessories Power Supplies/Readouts...

 Open the catalog to page 2
HFM-200-LFE-4

Working Pressure C. Board 01 Standard 02 24 VDC ** Output - 0-20 Ma 01 0-5 Volts (Std) 02 4-20 mA *** Fittings 05 1/4" Elbow O'Rings Viton (std) Kalrez Neoprene Buna-N Working Pressure 01 500 psig (Std) Calibration Type NIST 5 point (Std) NIST 10 Point NIST 20 Point Curve Fit No RF Rejection - 0 -5 volts only Not available with fast response

 Open the catalog to page 4
HFM-200-LFE-5

Laminar-Selection Chart Range Information Each calibration will require the following information: Range: Flow Units: *Std Conditions: * For volumetric units the standard temperature and pressure of the unit is also required 0°C & 760 Torr will be used when other values are not specified. * * * ** No RF Rejection - 0 -5 volts only *** Not available with fast response Fittings 01 NPT Male 02 Smooth 03 Flanges * Telephone: (757) 723-6531 Toll Free: (800) 950-2468 Fax: (757) 723-3925 World Wide Web: http://www.teledyne-hi.com E-mail: [email protected] P.O. Box 1436 Hampton, VA 23661...

 Open the catalog to page 5
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.