Product catalog summary
Features
  • Available flow ranges from 0-50 SLM to 0-1000 SLM.
  • Accuracy in nitrogen: HFC-D-306A/B and HFM-D-306A/B have an accuracy of ±0.5% of full scale.
  • Operating pressure up to 300 PSI.
  • IP-67 version available for 'A' series.
  • NIST traceable calibration.
  • Optional touchscreen display for 'B' series.
  • USB connectivity for 'B' series.
  • Totalized flow available for digital versions only.
Applications
  • Alternate energy
  • Fuel cell R&D
  • Secondary calibration reference
  • Specialty gas delivery
  • Custody transfer
Benefits
  • High accuracy
  • Fast metering response
  • Superior linearity
  • Rapid controller settling time
  • Digital extended range
Description

Teledyne Hastings Instruments products represent over 60 years of experience in the design and manufacture of gas flow instrumentation. The high flow meters and controllers featured in this document are based on the 300 Series line of flow sensors. The 300 Series transducer is a patented thermal-based flow sensor designed to provide exceptional linear response to changing flow rates. The electronics associated with the sensor are precisely tuned to give fast response time.

The flow transducer is combined with a laminar flow element to configure the flow meter for the customer. In many flow meters, the major source of error is the non-linearities that occur in the laminar flow element due to entrance and exit effects. However, the 300 Series laminar flow element is designed such that the flow sensing region is far less susceptible to these effects.

Specifications & Standards
  • European Pressure Equipment Directive (PED) 97/23/EC
  • EN 61010-1 Safety of Electrical Equipment for Measurement, Control, and Laboratory Use
  • EN 61326-1 Radiated Emissions
  • EN 61326-1 Conducted Emissions
  • EN 61000-4-2 Electrostatic Discharge
  • EN 61000-4-3 Radiated RF Immunity
  • EN 61000-4-4 Electrical Fast Transient/Burst
  • EN 61000-4-6 Conducted RF
  • EN 61000-4-8 Magnetic Field
Performance
  • Full Scale Flow Ranges: 0-50 to 0-1000 SLM
  • Microprocessor-based enhanced accuracy and digital features: Yes for HFC-D-306A/B and HFM-D-306A/B, No for HFM-306
  • Accuracy: ±0.5% of full scale for digital versions, ±1% for analog version
  • Repeatability: ±0.2% of full scale
  • Operating Temperature: -10°C to 50°C
  • Warm-up Time: 30 min for optimum accuracy, 15 min for ±1% of full scale
  • Operating Pressure: Maximum 300 PSIG
Electrical
  • Voltage: 15-24 VDC
  • Power: 2.5 W for 'A' Series, 3.5 W for 'B' Series
Physical
  • Weight: Approximately 3.5 lbs (1.6 kg) for HFC-D-306A/B, 4 lbs (1.8 kg) for HFM-D-306A/B
  • Wetted Materials: 316 SS, 303 SS, Nickel 200, Viton (optional Buna-N, Kalrez, PTFE)
See more

Catalog excerpts

Digital 300 Vue Thermal Mass Flow, 306/308-1

TELEDYNE HASTINGS INSTRUMENTSEverywhereyoulook-” HFC-D-308A/B Digital Flow Controller HFM-D-306A/B Digital Flow Meter {/) HFM-306 Analog Flow Meter ^ • Available Flow Ranges: » Operating Pressure to 300 PSI IP-67 Version Available (“A” Series) NIST Traceable Calibration Touchscreen Display Option (“B” Series) RS232/RS485 (Digitals Only) USB (“B” Series) Totalized Flow (Digitals Only) odCO HFC-D-308A Mass Flow Controller APPLICATIONS • Alternate Energy • Fuel Cell R&D • Secondary Calibration Reference • Specialty Gas Delivery • Custody Transfer BENEFITS • High Accuracy • Fast Metering Response • Superior Linearity • Rapid Controller Settling Time • Digital Extended Range HFM - 306 Analog Flow Meter TELEDYNE HASTINGS INSTRUMENTS Eve

 Open the catalog to page 1
Digital 300 Vue Thermal Mass Flow, 306/308-2

Teledyne Hastings Instruments products represent over 70 years of experience in the design and manufacture of gas flow instrumentation. The high flow meters and controllers featured in this document are based on the 300 Series line of flow sensors. The 300 Series transducer is a patented thermal based flow sensor. This sensor is designed to provide exceptional linear response to changing flow rates. In addition, the electronics associated with the sensor are precisely tuned to give fast response time. The flow transducer is combined with a laminar flow element to configure the flow meter for...

 Open the catalog to page 2
Digital 300 Vue Thermal Mass Flow, 306/308-4

Performance Full Scale Flow Ranges (N2) Microprocessor-based Enhanced Accuracy and Digital Features Accuracy Repeatability Operating Temperature Warm up time Temperature Coefficient of Zero Temperature Coefficient of Span Operating Pressure -Maximum Electrical Voltage Power “A” Series Power “B” Series Physical Weight (approx.) Wetted Materials 2,500 to 10,000 SLM YES ± (0.5% Rdg + 0.2% FS) ±0.15% of FS -20° - 70°C 30 min optimum accuracy 2 min for ± 5% of full scale Auto-zero when Valve is Closed Maximum ±0.1%(Rdg)/°C (-20—70°C) 300 psig 11 -36 VDC (±12, ±15 acceptable) 15 VDC min. for 0-20 &...

 Open the catalog to page 4
Digital 300 Vue Thermal Mass Flow, 306/308-7

Outline Drawing HFM-D-306A (with IP-67)

 Open the catalog to page 7
Digital 300 Vue Thermal Mass Flow, 306/308-8

Outline Drawing HFC-D-308A (with IP-67)

 Open the catalog to page 8
Digital 300 Vue Thermal Mass Flow, 306/308-10

Selection Chart HFM-D-306A & HFC-D-308A Model Number Circuit Board Circuit Board 01 Pinout H (Std) 02 IP-67 Enclosure Output 0-5 Volt (Std) 0-10 Volt 4-20 mA 0-20 mA Fittings 01 1” Compression 02 1 1/2” Compression 03 2” Compression (Std) Seals 01 Viton (Std) 04 Buna-N 02 Kalrez (Meter Only) 03 Neoprene (Meter Only) Calibration Type 1 NIST Traceable Calibration Report (Std) 2 NIST Traceable Calibration Reports 3 NIST Traceable Calibration Reports 4 NIST Traceable Calibration Reports 5 NIST Traceable Calibration Reports 6 NIST Traceable Calibration Reports 7 NIST Traceable Calibration Reports...

 Open the catalog to page 10
Digital 300 Vue Thermal Mass Flow, 306/308-12

Model Number Range Information for all Instruments Each calibration will require the following information: Range Flow Units Gas For the HFC Instruments also Upstream Pressure (maximum & minimum):': Downstream Pressure (maximum & minimum) Does the downstream pressure change with flowrate? Y/N For volumetric units the standard temperature and pressure of the unit is also required 0°C & 760 Torr will be used when other values are not specified

 Open the catalog to page 12
Digital 300 Vue Thermal Mass Flow, 306/308-13

THCD-100 Single Channel Power Supply Meter THCD-100 Includes brackets, connectors, and backshells 24 VDC Switching Power Supply 12-01-169 For use with “B' Series or THCD-100 (Please specify AC Input Clip) Connects Hastings Power Supply (15-pin) to 300 “A" Series (15-pin) Other lengths available Connects Hastings Power Supply (15-pin) to 300 “B” Series (9-pin) Other lengths available CB-12PCF-8-LDS (BareLeads) CB-12PCF-8-AM (Hastings ftwerSupply) Other lengths available “B” Series—Serial Communication Cable CB-RS232-TRRS RS232 Cable (9-pin “D” Female to Male TRRS) 6’ Cable (~1.8m) “A” Series—Serial...

 Open the catalog to page 13
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.