Overview: The Model 100E UV Fluorescence SO2 Analyzer utilizes UV fluorescence and advanced microprocessor technology to deliver precise measurements of low-level sulfur dioxide (SO2). It features an optical shutter and reference detector for stability, a hydrocarbon kicker to prevent interference, and multitasking software for real-time monitoring and diagnostics.
Key Features:- Ranges from 0-50 ppb to 0-20 ppm, user-selectable with dual ranges and auto-ranging capabilities.
- Microprocessor-controlled with continuous self-checking and alarms.
- Data logging capabilities with internal memory for storing various parameters.
- Remote operation via dual bi-directional RS-232 ports, with optional RS-485 or Ethernet.
- Temperature and pressure compensation, internal zero and span check (optional).
- EPA, MCERTS, and EN14212 approved.
Specifications:- Zero Noise: < 0.2 ppb (RMS), Span Noise: < 0.5% of reading (RMS) above 50 ppb.
- Lower Detectable Limit: 0.4 ppb, Zero Drift: < 0.5 ppb/24 hours.
- Sample Flow Rate: 650 cm3/min ±10%.
- Operating Temperature: 5 - 40˚C.
- Dimensions: 7” x 17” x 23.5”, Weight: 35 lbs.
- Power: 100V - 240V, 50/60 Hz, 250W.
Ordering Information: The Model 100E includes a standard two-year warranty, selectable voltage, internal pump, and various output options. Calibration and additional options such as rack mount brackets and Ethernet ports are available.
Contact Information: For more details, visit the Teledyne API website or contact via phone or email.
See more