Overview: The Model T108 Total Sulfur Analyzer is designed to measure sulfur impurities, known as Total Sulfides, in air or carbon dioxide gas. It integrates a modified Model T100 SO2 analyzer with specialized software and a Model 501TS SO2 high-temperature thermal oxidizer. The device features NumaView™ premium T Series software, a durable color touchscreen display, and includes lifetime technical support and a standard two-year warranty.
Specifications:
  • Measurement Ranges: Minimum 0-50 ppb, Maximum 0-20,000 ppb (selectable, dual range supported).
  • Measurement Units: ppb, ppm, μg/m3, mg/m3 (selectable).
  • Noise and Drift: Zero Noise <0.2 ppb (RMS), Span Noise 0.5% of reading (RMS) above 50 ppb, Zero Drift <0.5 ppb/24 hours, Span Drift <0.5% of full scale/24 hours.
  • Response Time: Less than 120 seconds to 95%.
  • Linearity and Precision: Linearity 1% of full scale, Precision 0.5% of reading.
  • Sample Flow Rate: 650 cc/min ±10%.
  • Power Requirements: 100V-120V, 220V-240V, 50/60 Hz.
  • Analog Output Ranges: 10V, 5V, 1V, 0.1V (selectable).
  • Included I/O: Ethernet, RS232, USB ports, opto-isolated digital outputs and inputs, analog outputs.
  • Optional I/O: USB com port, RS485, digital alarm outputs, Multidrop RS232, 4-20mA current outputs.
  • Operating Temperature Range: 5 - 40ºC.
  • Dimensions and Weight: Analyzer and Converter both measure 7” x 17” x 23.5” and weigh 45 lbs and 20 lbs respectively.
  • Certifications: US EPA: EQSA-0495-100.
Additional Information: Specifications are subject to change without notice and are based on constant conditions. For more information, contact Teledyne API at their San Diego office or visit their website.
See more

Catalog excerpts

T108-1

The Model T108 Total Sulfur Analyzer ^ADVANCED POLLUTION INSTRUMENTATION The Model T108 Total Sulfur analyzer is designed to measure mixed sulfur impurities, collectively referred to as Total Sulfides, in air or carbon dioxide gas. The model T108 consists of a modified Model T100 SO2 analyzer with special software and a Model 501TS SO2 high temperature thermal oxidizer. — With NumaView™ premium T Series software — Large, vivid, and durable color touchscreen display All other T Series instrument platform features Lifetime technical support by phone and email Standard two-year warranty

 Open the catalog to page 1
T108-2

Specifications subject to change without notice. All specifications are based on constant conditions. TELEDYNE APIEverywhereyoulook" 9970 Carroll Canyon Road ■ San Diego, CA 92131 Ph. 858-657-9800 Fax 858-657-9816 Email [email protected] For more information about the Teledyne API family of monitoring instrumentation products, call us or visit our website at: www.teledyne-api.com © 2019 Teledyne API Printed documents are uncontrolled. SAL000044D (DCN 8120) 07.29.19

 Open the catalog to page 2

All Teledyne API catalogs and technical brochures

  1. T200

    2  Pages

  2. T102

    2  Pages

  3. T101

    2  Pages

  4. T100H

    2  Pages

  5. CAT-O3 Series

    2  Pages

  6. ​SCI-552

    2  Pages

  7. ​Model ODS

    2  Pages

  8. Mini ODS

    2  Pages

  9. Model 651

    2  Pages

  10. T640

    4  Pages

  11. 452

    2  Pages

  12. 465H

    2  Pages

  13. 465M

    2  Pages

  14. Model 702

    2  Pages

  15. Model T204

    2  Pages

  16. Model T700H

    2  Pages

  17. Model 751H

    2  Pages

  18. Model 751

    2  Pages

  19. T701H

    2  Pages

  20. T701

    2  Pages

  21. T750

    2  Pages

  22. T500U

    2  Pages

  23. T750U

    2  Pages

Archived catalogs

  1. T Series

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.