video corpo

T100H

T100H

T100H

Product catalog summary
Overview: The Model T100H High Range UV Fluorescence SO2 Analyzer is designed for accurate and continuous monitoring of high concentration stack gases using UV fluorescence technology. It features advanced electronics and the NumaView™ premium T Series software, providing a user-friendly interface with a large color touchscreen display.
Key Features:
  • Lifetime technical support via phone and email.
  • Standard two-year warranty.
  • Optional internal O2 or CO2 sensor.
Specifications:
  • Measurement Ranges: Minimum 0 - 10 ppm, Maximum 0 - 5,000 ppm (selectable with dual range support).
  • Measurement Units: ppm, mg/m3 (selectable).
  • Noise and Drift: Zero Noise at 0.1 ppm (RMS), Span Noise < 1% of reading (RMS) above 10 ppm, Zero Drift < 1 ppm/24 hours, Span Drift < 0.5% of full scale/24 hours.
  • Response Time: Less than 40 seconds to 95%.
  • Linearity: 1% of full scale.
  • Precision: 0.5% of reading above 10 ppm.
  • Sample Flow Rate: 700 cc/min ±10%.
  • Power Requirements: 100V-120V, 220V-240V, 50/60 Hz.
  • Analog Output Ranges: 10V, 5V, 1V, 0.1V (selectable).
  • Included I/O: Ethernet, RS232, USB ports, opto-isolated digital outputs and inputs, analog outputs.
  • Optional I/O: USB com port, RS485, digital alarm outputs, Multidrop RS232, 4-20mA current outputs.
  • Operating Temperature Range: 5 - 40ºC.
  • Dimensions and Weight: Analyzer: 7” x 17” x 23.5” (178 x 432 x 597 mm), Weight: 31 lbs (16 kg); External pump: 15 lbs (7 kg).
Additional Information: Specifications are subject to change without notice and are based on constant conditions. For more information, contact Teledyne API or visit their website.
See more

Catalog excerpts

T100H-1

The Model T100H High Range UV Fluorescence SO2 Analyzer ADVANCED POLLUTION INSTRUMENTATION model tiooh The Model T100H High Range UV Fluorescence SO2 analyzer uses the proven UV fluorescence principle and advanced electronics to allow accurate, dependable, continuous measurements for high concentration stack gas monitoring and other applications. — With NumaView™ premium T Series software — Large, vivid, and durable color touchscreen display ■ Lifetime technical support by phone and email All other T Series instrument platform features ■ Standard two-year warranty Optional internal O2 or CO2 sensor TELEDYNE APIEverywhereyoulook"

 Open the catalog to page 1
T100H-2

Specifications subject to change without notice. All specifications are based on constant conditions. For more information about the Teledyne API family of monitoring instrumentation products, call us or visit our website at: www.teledyne-api.com © 2019 Teledyne API Printed documents are uncontrolled. SAL000040D (DCN 8120) 07.29.19 9970 Carroll Canyon Road ■ San Diego, CA 92131 Ph. 858-657-9800 Fax 858-657-9816 Email [email protected]

 Open the catalog to page 2

All Teledyne API catalogs and technical brochures

  1. T200

    2  Pages

  2. T108

    2  Pages

  3. T102

    2  Pages

  4. T101

    2  Pages

  5. CAT-O3 Series

    2  Pages

  6. ​SCI-552

    2  Pages

  7. ​Model ODS

    2  Pages

  8. Mini ODS

    2  Pages

  9. Model 651

    2  Pages

  10. T640

    4  Pages

  11. 452

    2  Pages

  12. 465H

    2  Pages

  13. 465M

    2  Pages

  14. Model 702

    2  Pages

  15. Model T204

    2  Pages

  16. Model T700H

    2  Pages

  17. Model 751H

    2  Pages

  18. Model 751

    2  Pages

  19. T701H

    2  Pages

  20. T701

    2  Pages

  21. T750

    2  Pages

  22. T500U

    2  Pages

  23. T750U

    2  Pages

Archived catalogs

  1. T Series

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.