1. Catalogs
  2. Teledyne API
  3. Model 460L Process Ozone Monitor
video corpo

Model 460L Process Ozone Monitor

Model 460L Process Ozone Monitor

Model 460L Process Ozone Monitor

Product catalog summary
Overview: The Model 460L is a microprocessor-based ozone monitor designed for measuring low concentration ozone levels in various applications such as water treatment, food processing, and research. It offers accurate and stable readings with minimal maintenance.
Key Features:
  • Utilizes UV absorption method for measurement.
  • User-selectable ranges from 1ppm to 500ppm.
  • Microprocessor-based with analog and digital I/O for interfacing with other devices.
Specifications:
  • Measurement Principle: UV Absorption (Beer Lambert Law).
  • Accuracy: ± 1% of Full Scale.
  • Response Time: Less than 30 seconds to 95%.
  • Gas Flow Rate: 0.5 – 2.0 LPM.
  • Temperature Range: 5-45°C.
  • Humidity Range: 10-90% RH, Non-Condensing.
  • Dimensions: 16.01" x 13.31" x 8.02".
  • Weight: 14.60lb (6.64kg).
  • Power: 110-240V~, 50/60Hz, 1.1A.
  • Environmental Protection: IP65 (NEMA 4X).
Ordering Information: The Model 460L includes 1/4” SS Compression fittings, pressure/temperature compensation, flowmeter, and needle valve. Standard ranges and interface options such as RS485 and RS232 are available. Analog output options include 0-5V (standard) and 4-20mA (optional).
Additional Information: For more details, contact Teledyne API or visit their website.
See more

Catalog excerpts

Model 460L Process Ozone Monitor-1

The Model 460L is a microprocessor based low concentration ozone monitor for measuring ambient level ozone concentrations in water treatment, food processing, and research applications. The 460L has been designed to give accurate and stable readings over long time periods with little or no maintenance or calibration.The flexibility of the software as well as the analog and digital I/O allow the Model 460L to interface with a broad range of devices for process control and data logging. 8 > UV absorption method 8 > 1ppm to 500ppm ranges - user selectable 8 > Microprocessor based 8 > Analog and Digital I/O size="-1">

 Open the catalog to page 1
Model 460L Process Ozone Monitor-2

Model 460L Process Ozone Monitor includes: Standard ranges (specify): Specify Interface: > Flowmeter Analog Output Type: > 4-20mA (optional) Measurement Principle: UV Absorption (Beer Lambert Law)Ranges: 1ppm to 500ppm; 1000ppb to 10,000ppbMeasurement Units: ppm; ppbAccuracy: α 1% of Full ScaleZero Noise: < .0015 ppm (rms)Span Noise: < .5% of reading (rms) (above 0.3 ppm)Lower Detectable Limit: < .003 ppm (rms)Linearity: Better than 1% full scaleResponse Time: <30 seconds to 95%Repeatability: 0.5% of full scale rangeDisplay Resolution: .001 ppm, 1 ppbGas Flow Rate: 0.5 і 2.0 LPMCompensation:...

 Open the catalog to page 2

All Teledyne API catalogs and technical brochures

  1. T200

    2  Pages

  2. T108

    2  Pages

  3. T102

    2  Pages

  4. T101

    2  Pages

  5. T100H

    2  Pages

  6. CAT-O3 Series

    2  Pages

  7. ​SCI-552

    2  Pages

  8. ​Model ODS

    2  Pages

  9. Mini ODS

    2  Pages

  10. Model 651

    2  Pages

  11. T640

    4  Pages

  12. 452

    2  Pages

  13. 465H

    2  Pages

  14. 465M

    2  Pages

  15. Model 702

    2  Pages

  16. Model T204

    2  Pages

  17. Model T700H

    2  Pages

  18. Model 751H

    2  Pages

  19. Model 751

    2  Pages

  20. T701H

    2  Pages

  21. T701

    2  Pages

  22. T750

    2  Pages

  23. T500U

    2  Pages

  24. T750U

    2  Pages

Archived catalogs

  1. T Series

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.