Overview: The Model 360E is a Gas Filter Correlation CO2 Analyzer designed to measure carbon dioxide (CO2) concentrations using infrared absorption based on the Beer-Lambert law. It features a Gas Filter Wheel for CO2-specific measurement and offers superior stability and sensitivity.
Key Specifications:- Measurement Ranges: 0-2 ppm to 0-2000 ppm, user-selectable with dual ranges and auto-ranging supported.
- Units: ppb, ppm, μg/m3, mg/m3, %.
- Zero Noise: < 0.1 ppm (RMS).
- Span Noise: < 1.0% of reading above 100 ppm (RMS).
- Lower Detectable Limit (LDL): 0.2 ppm.
- Zero Drift: < 0.25 ppm/24 hours, < 0.5 ppm/7 days.
- Span Drift: < 0.5% of reading/24 hours, < 1% of reading/7 days.
- Lag Time: 10 seconds.
- Rise and Fall Time: < 60 seconds to 95%.
- Linearity: 1% of full scale.
- Precision: 0.5% of reading.
- Sample Flow Rate: 800 cc/min ±10%.
- Operating Temperature Range: 5 - 40˚C.
- Dimensions: 7” x 17” x 23.5”.
- Weight: 40 lbs (17 kg).
- Power: 100V - 120V, 220V - 240V, 50/60 Hz, 250W.
Features:- Free customer support via telephone and email for the life of the instrument.
- Standard two-year warranty with a five-year guarantee on the GFC wheel.
- Multi-tasking software for real-time monitoring and automatic alarms.
- Built-in data acquisition with internal memory for logging parameters.
- Bi-directional RS-232 / RS-485 / Ethernet connectivity.
- Digital status outputs and adaptive signal filtering.
- Temperature and pressure compensation.
Optional Equipment and Accessories:- Rack mount brackets with chassis slides.
- 4-20 mA isolated outputs (up to three channels).
- Ethernet port (replaces one serial port).
- Calibration options with stainless steel valves and internal zero air scrubber.
- RS-232 Cable, Expendables Kit, and Spare Parts Kit.
Ordering Information: Specify input AC voltage and frequency, and output DC voltage. Optional equipment and calibration options are available.