Overview: The Teledyne API Model 200EU NOy Analyzer is designed for ozone precursor research, utilizing chemiluminescence to measure Total Reactive Oxides of Nitrogen. It is part of a comprehensive monitoring program for ozone precursors.
Key Features:- Remote catalytic converter converts reactive nitrogen compounds to NO at the sampling point.
- Stainless steel NEMA 4X housing with a flexible metal umbilical connects the converter to the Flow Module.
- Flow Module provides bypass flow and calibration gas tubing purge.
- High sensitivity with a lower detectable limit of 0.05 ppb.
- Compatible with Model 200EU Ultra-Sensitive NOx Analyzer, Model 700E Multigas Calibrator, and Model 701H Ultra-High Purity Zero Air System for comprehensive monitoring.
Specifications:- Ranges: 0-50 ppb to 0-2 ppm, user selectable.
- Zero Noise: < 0.025 ppb; Span Noise: < 0.5% of reading above 5 ppb.
- Zero Drift: < 0.1 ppb/24 hours; Span Drift: < 0.5% of reading/24 hours.
- Operating Temperature Range: 5 - 40°C.
- Power: 100V - 120V, 220V - 240V, 50/60 Hz, 250W.
Additional Features:- Microprocessor controlled with multi-tasking software.
- Automatic purge of calibration lines and adaptive signal filtering.
- Temperature & Pressure compensation and converter efficiency correction software.
- Internal data logging with multiple averages from 1 min to 365 days.
- Free customer support for the life of the instrument.
Ordering Information: Specify input AC voltage & frequency, output DC voltage, and optional equipment such as rack mount brackets, 4-20mA outputs, and Ethernet port.
See more