BDS series

BDS series

BDS series

Product catalog summary
Introduction
Teledyne Analytical Instruments (TAI) addresses the critical need for monitoring oxygen contamination in semiconductor manufacturing processes. Oxygen contamination can significantly affect product yields, necessitating precise monitoring at parts-per-billion (ppb) levels in ultra-high purity (UHP) gases.

Product Overview
TAI introduces the Bi-potentiostat Driven Sensor (BDS), a high-sensitivity oxygen sensor designed for semiconductor applications. The BDS offers a lower detection limit (LDL) of 500 ppt, rapid response, and ease of use, surpassing previous wet-cell oxygen sensors.

Theory of Operation
The BDS is a four-electrode, non-depleting electrochemical cell driven by a bi-potentiostat circuit. It includes a sensing electrode for oxygen detection, a protective electrode to remove dissolved oxygen, a reference electrode for potential control, and a counter electrode for current path.

Design and Features
The BDS design ensures precise control of electrode potentials, fast recovery from air exposure, and operation in open-air conditions. It is suitable for environments with limited acid gases. The sensor is constructed with high-purity materials and leak-checked to prevent atmospheric contamination.

Specifications
- User-programmable ranges: 0-50 ppb, 0-1000 ppb, 0-10 ppm
- Accuracy: ±2% at constant temperature, ±5% over full range
- Operating temperature: 5 to 40°C
- Response time: 90% in less than 60 seconds
- Outputs: 4-20 mADC, 0-1 VDC, RS-232C interface
- Warranty: Two years, with an expected life of 3-5 years

Additional Features
The BDS includes auto-calibration software, self-diagnostics, and system failure alarms. It can be configured for fixed or auto-ranging analysis and is available in compact or rack-mount designs.

Conclusion
Teledyne's BDS offers a reliable, cost-effective solution for detecting ppb levels of oxygen in semiconductor manufacturing, with advanced features and robust design ensuring high accuracy and minimal downtime.
See more

Catalog excerpts

BDS series-1

ANALYTICAL INSTRUMENTS -DOtentiostat Drivei

 Open the catalog to page 1
BDS series-2

s the density of wafers produced within a semiconductor facility increases, so does the stringent requirement to accurately monitor for oxygen contamination at or below parts-per-billion (ppb) levels. Oxygen contamination is routinely monitored in the ultra-high purity (UHP) bulk gases such as nitrogen, argon, helium and hydrogen, which are used as blanket or carrier gases during the vapor deposition, sputtering and annealing processes. The ingress of oxygen at any point during the wafer fabrication process will have a detrimental effect on product yields. Therefore, it is imperative for semiconductor...

 Open the catalog to page 2
BDS series-3

SENSOR BAFFLE MECHANISM Teledyne's unique, patent-applied-for mechanism within the BDS sensor minimizes sensor downtime when water is added. •  Three user-programmable ranges which can be tailored to the application needs in either a fixed or auto-ranging analysis mode •  Auto-calibration software •  RS-232C bi-directional serial interface as well as 4-20 mA and 0-1 VDC concentration output signals •  Self-diagnostics and system failure alarm along with 2x fully adjustable concentration alarm set points •  low inlet sample pressure furnace-type applications, the BDSFor 3000 can also be fitted...

 Open the catalog to page 3
BDS series-4

BDS SERIES OF PPB OXYGEN ANALYZERS SPECIFICATIONS Ranges: Three user programmable ranges Default ranges: Over-range is not user programmable ±2% of full scale at constant temperature ± 5% of full scale over entire operating temperature range Detection method: Bi-potentiostat Driven Sensor (BDS) Technology (replenishable, non-depleting type) Operating temp range: 5 to 40° C Lower detection limit: Response time: 90% in less than 60 seconds 4-20 mADC isolated percent of range output 0-1 VDC percent of range output and range ID output Auto-calibration software: Standard; the frequency of the autocalibration...

 Open the catalog to page 4

All Teledyne Analytical Instruments catalogs and technical brochures

  1. PEM 9004

    4  Pages

  2. INSTA TRACE

    2  Pages

  3. PHOTO-X

    6  Pages

  4. MODEL MAX5

    2  Pages

  5. 9110T Series

    4  Pages

  6. 9060Z

    4  Pages

  7. 8800 series

    6  Pages

  8. 1220

    4  Pages

  9. SERIES 6700

    6  Pages

  10. model 9070

    2  Pages

  11. MODEL 3010TA

    2  Pages

  12. 3000TA-XL

    2  Pages

  13. 3020T

    2  Pages

  14. 3000TB

    2  Pages

  15. R-15

    1  Page

  16. UFO-130-2

    2  Pages

  17. T-7

    1  Page

  18. R-22MED

    1  Page

  19. R-17MED

    1  Page

  20. PEM-9004

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.