Overview: The Tektronix FCA3000 and FCA3100 Series Timer/Counter/Analyzers are designed for precision measurements in demanding environments, offering high frequency and time resolution, deep internal memory, and fast data transfer rates, suitable for R&D and production testing.
Key Specifications:
  • Available models for 300 MHz, 3 GHz, and 20 GHz.
  • Single-shot time resolution: 50 ps (FCA3100) or 100 ps (FCA3000).
  • 12 digit/s frequency resolution and 0.001° phase resolution.
  • Voltage resolution of 3 mV or better.
  • Optional high-stability oven time base.
Measurement Capabilities:
  • Automated measurements: frequency, period, ratio, time interval, pulse width, rise/fall time, phase angle, duty cycle, and voltage parameters.
  • Features: trend plot mode, measurement statistics, histogram mode, and Allan deviation analysis.
  • Zero dead-time frequency/period measurements and continuous data streaming (FCA3100).
Connectivity and Software:
  • USB and GPIB ports for PC connectivity.
  • SCPI-compatible programmability and emulation mode for ATE systems.
  • Includes LabVIEW SignalExpress™ TE Limited Edition software for bench connectivity.
  • Optional TimeView™ software for modulation domain analysis.
Performance and Features:
  • High measurement throughput with up to 250k samples/s to internal memory.
  • Graphical display for real-time trend plots and histograms.
  • Multi-parameter display for auxiliary measurements alongside main measurements.
  • Fast setup with intuitive operation and autoset function.
Warranty and Support: The FCA Series comes with a three-year warranty, ensuring reliability and support.
Conclusion: The Tektronix FCA3000 and FCA3100 Series offer comprehensive measurement capabilities, high performance, and ease of use, making them ideal for complex design testing and analysis.
Specifications: The document outlines specifications such as an impedance of 50 Ω nominal with a VSWR of less than 2.5:1, and a maximum voltage without damage of 12 VRMS, protected by a pin-diode. The frequency range is from 0.25 to 20 GHz, with varying operating input voltage ranges depending on the frequency band. The prescaler factor is 128, and the AM tolerance is greater than 90% within the sensitivity range. Connectors are Type-N Female and Precision-N Female.
Inputs and Outputs: The rear panel includes reference inputs and outputs, arming inputs, and pulse outputs. The reference input supports 1, 5, or 10 MHz frequencies, while the reference output is 10 MHz. The arming input allows for the arming of all measuring functions, with an impedance of approximately 1 kΩ and a frequency range from DC to 80 MHz. The pulse output is programmable through front GPIB/USB, with TTL levels in 50 Ω.
Auxiliary Functions: The device includes trigger holdoff with a time delay range of 20 ns to 2 s, and external start and stop arming modes. Statistics functions include maximum, minimum, mean, standard deviation, and Allan deviation, with display options for numeric, histograms, or trend plots.
Mathematics and Other Functions: Mathematical functions include (K*X+L)/M and (K/X+L)/M, where X is the current reading. Other functions include measuring time, time-base reference options, display hold, and limit alarms with graphical indications.
Calibration and General Specifications: Calibration is menu-controlled with specific frequencies. Environmental specifications include operating temperatures from 0 °C to +50 °C and storage temperatures from –40 °C to +71 °C. The device complies with various safety and EMC directives.
Power and Time-base Options: Power requirements range from 90 to 265 VRMS, with time-base options for standard, medium stability, and high stability. The document details the uncertainty due to aging and temperature variations.
Physical Dimensions: The device measures 90 mm in height, 210 mm in width, and 395 mm in depth, with a net weight of 2.7 kg.
Ordering Information: The document lists various models and options, including medium and high-stability oven time bases, rear-panel connectors, and power plug options. Service options include calibration and repair services, with recommended accessories and software also provided.
See more

Catalog excerpts

FCA3000 and FCA3100 Series-1

Timer/Counter/Analyzers Tektronix FCA3000 and FCA3100 Series Datasheet Available Functions and Features Automated Measurements: Frequency, Period, Ratio, Time Interval, Time Interval Error, Pulse Width, Rise/Fall Time, Phase Angle, Duty Cycle, Maximum Voltage, Minimum Voltage, Peak-to-Peak Voltage Totalize Measurement (FCA3100 Series) Multi-parameter Display Trend Plot Mode Measurement Statistics Mode Histogram Mode Features & Benefits Key Performance Specifications Allan Deviation Zero Dead-time Frequency/Period Measurements Continuous Data Streaming over USB/GPIB Bus during Measurement (FCA3100 Series) 300-400 MHz in Manual Trigger Mode Programmable Pulse Output from 0.5 Hz to 50 MHz Up to 3 Input Channels 50 ps (FCA3100 Series) or 100 ps (FCA3000 Series) Single-shot Time Resolution Connectivity Optional Rear-panel Inputs 12 Digit/s Frequency Resolution USB Device and GPIB Ports on Rear Panel for Quick PC Connectivity GPIB Interface Supports Full SCPI-compatible Programmability and offers an Emulation Mode for Plug-and-Play Replacement in Existing ATE Systems 3 mV or better Voltage Resolution Optional 5×10–8 High-stability Oven Time Base Measurement Throughput 250k Sample/s Data Transfer Rate to Internal Memory (Up to 3.75M samples stored) Up to 15k Sample/s Data Transfer Rate over USB/GPIB Bus (Block mode) Up to 650 Individually Triggered Measurements/s External Arming Input 10 MHz Reference Oscillator Output Includes National Instrument's LabVIEW SignalExpress™ TE Limited Edition Software for Connecting Your Bench Optional TimeView™ Software Available for Modulation Domain Analysis

 Open the catalog to page 1
FCA3000 and FCA3100 Series-2

Feature-rich Tools for Precision Measurements The FCA3000 and FCA3100 Timer/Counter/Analyzer Series pack many different functions into one feature-rich instrument. With industry-leading frequency and time resolution, the FCA Series comes standard with deep internal memory and a fast data transfer rate of 250k Samples/s to internal memory. In addition, the multi-parameter display shows auxiliary measurements alongside your main measurement to provide you with the results you need at a glance. With the industry's most comprehensive analysis modes, including measurement statistics, histograms, and...

 Open the catalog to page 2
FCA3000 and FCA3100 Series-3

Timer/Counter/Analyzers — Tektronix FCA3000 and FCA3100 Series Trend Plot Analysis. Tektronix TimeView™ Software. Measurement Statistics. Histogram Plot. Measurement Trend Plots Depending on your test case, your signal parameters may change from instant to instant. With the Trend Plot Analysis mode, you can graphically plot the trend of a measured value over time. Measurement Statistics With integrated statistics processing, you can calculate the average, standard, and Allan deviation of a measurement, as well as track the minimum and maximum measured values, all with the push of a button. Histogram...

 Open the catalog to page 3
FCA3000 and FCA3100 Series-4

Connect Your Bench for Intelligent Debug Easily capture, save, and analyze measurement results from your FCA Series timer/counter/analyzer with the special Tektronix Edition of National Instruments LabVIEW SignalExpress™ software. Every FCA Series timer/counter/analyzer ships with a free copy of the Limited Edition version of SignalExpress for basic instrument control, data logging, and analysis. The optional Professional Edition offers over 200 built-in functions that provide additional signal processing, advanced analysis, sweeping, limit testing, and user-defined step capabilities. Measuring...

 Open the catalog to page 4
FCA3000 and FCA3100 Series-5

Timer/Counter/Analyzers — Tektronix FCA3000 and FCA3100 Series Time Interval A to B, B to A, A to A, B to B Characteristic Range Resolution Min Pulse Width Smart Calculation Normal calculation: 0 ns to s Smart calculation: –106 s to +106 s 100 ps single (FCA3000 Series) 50 ps single (FCA3100 Series) 1.6 ns Smart Time Interval to determine sign (A before B or A after B) Positive and Negative Pulse Width A, B Characteristic Range Min Pulse Width Aux Parameters Rise and Fall Time A, B Characteristic Range Trigger Levels Min Pulse Width Aux Parameters 1.5 ns to 106 s 10% and 90% of signal amplitude...

 Open the catalog to page 5
FCA3000 and FCA3100 Series-6

Input C – 3 GHz (FCA3003 and FCA3103 Products) Characteristic Operating Input Voltage Range 100 to 300 MHz 20 mVRMS to 12 VRMS 10 mVRMS to 12 VRMS 0.3 to 2.5 GHz 20 mVRMS to 12 VRMS 2.5 to 2.7 GHz 40 mVRMS to 12 VRMS 2.7 to 3.0 GHz Prescaler Factor 16 Impedance 50 Ω nominal, VSWR < 2.5:1 Max Voltage without 12 VRMS, pin-diode protected Damage Type-N Female Connector External Start and Stop Arming Characteristic Modes Start, Stop, Start and Stop Arming Input Channels A, B, or (rear panel) E Max Rep. Rate for Arming Signal 160 MHz Channel A, B 80 MHz Channel E Start-time Delay Range 20 ns to 2...

 Open the catalog to page 6
FCA3000 and FCA3100 Series-7

Timer/Counter/Analyzers — Tektronix FCA3000 and FCA3100 Series Height Width Depth Mode Closed case, menu controlled Calibration Frequencies 0.1, 1, 5, 10, 1.544, and 2.048 MHz Environmental Data Description Class Operating Temp Storage Temp Humidity Ordering Information General Specifications Power Requirements FCA3000 Timer/Counter/Analyzer 300 MHz / 100 ps FCA3003 Timer/Counter/Analyzer 3 GHz / 100 ps FCA3020 Timer/Counter/Analyzer 20 GHz / 100 ps FCA3100 Timer/Counter/Analyzer 300 MHz / 50 ps FCA3103 Timer/Counter/Analyzer 3 GHz / 50 ps FCA3120 Timer/Counter/Analyzer 20 GHz / 50 ps FCA3000/3100...

 Open the catalog to page 7
FCA3000 and FCA3100 Series-8

Contact Tektronix: ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777 Belgium 00800 2255 4835* Service Options Option Calibration Service 3 Years Calibration Service 5 Years Calibration Data Report Repair Service 5 Years Standard Warranty Extended to 5 Years (FCA3000, FCA3003, FCA3100, and FCA3103) Standard Warranty Extended to 5 Years (FCA3020 and FCA3120) Recommended Accessories and Software Rackmount Shelf Kit for 2 Units Hard Carrying Case Soft Carrying Case USB Host to Device Cable, 3 ft. GPIB Cable, Double Shielded...

 Open the catalog to page 8

Archived catalogs

  1. 6 series mso

    71  Pages

  2. 5 Series MSO

    73  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.