SWAM 5A Monitor
2Pages

{{requestButtons}}

Catalog excerpts

SWAM 5A Monitor - 1

Immission Line 2.220.01 Atmospheric PMx particles monitor · Suitable for PM10, PM2.5, PM1sampling inlet · Operating flow rate range 0.8 ÷ 2.5 m3/h · Sampling on 47mm diam. filter membrane for further analysis · Mass measurement using the ß attenuation method · Completely automatic management of sampling and measurement quality controls, with immediate validation of the supplied PMx concentration data · On line monitoring of all parameters related to the sampling process, with relative diagnostic warnings of possible anomalies · Sampling and measurement data storage on internal buffer · Local control with RS232 serial interface · Complete remote control via Modem/GSM SWAM 5A Monitor has been engineered to allow a 100% yearly data coverage (365 data/year). Actually, ordinary management, maintenance and calibration check are carried out without stopping the operating cycle. The instrument guarantees an excellent level of data quality and reliability. SWAM 5A can be supplied with PM10 or PM2.5 sampling inlet, working at 2.3 m3/h. The instrument gives the mass concentration measurement, in compliance with the Dl 155/10, acknowledging the European Directive 99/30/EC. As this system is in compliance with the EN 12341 and EN14907 standards, it can be used as a reference sampler. The instrument can also carry out sampling and measurement processes in compliance with the US EPA regulations, using the relative sampling inlets: • US EPA 40 CFR part 50 sampling inlet, for PM10 sampling, at 1 m3/h volume flow rate. • US EPA 40 CFR part 50 sampling inlet with WINS PM2.5 impactor, for PM2.5 sampling, at 1 m3/h volume flow rate. Filter membranes The instrument can use glass, quartz, Teflon-coated glass, Teflon and other filter membranes. The choice depends on the selected operating flow rate, the environmental conditions where the instrument is running, the compounds searched. Filter cartridges with reduced ß spot area are available, in order to increase the sensitivity in case of low concentrations. The instrument can manage automatically the changing of 36 filter membranes (72 on demand). At the end of every sampling and measurement cycle, the sampled filter will be moved to the similar unloading cylinder. The operator can replace the filters at any time, with no interference with the operating cycle in process. SWAM 5A Monitor uses a coaxial chamber inside the main sampling tube. Through this chamber passes the ambient air sucked by an auxiliary aspiration fan, in order to guarantee a good thermostatic effect on the sampling tube. The temperature near the filter and the external one are constantly monitored. This process allows to minimize volatile materials losses during the sampling process. Ed. 0 11/11 - Technical specifications may change without previous warning Main characteristics Product Data Immission Line 2.220.01 TECORA - Headquarter: 211-215 rue la Fontaine, 94134 Fontenay Sous Bois Cedex - FRANCE. Italian office: Via A. Volta 22, 20094 Corsico (Mi) - ITALY www.tecora.com SWAM 5A Monitor

Open the catalog to page 1
SWAM 5A Monitor - 2

Immission Line 2.220.02 Max pressure drop Flow rate measurement precision Flow rate measurement accuracy Power supply Absorbed electric power Power supply 0 ÷ 50 mg ± 0.3 µg/ m³ ( 24 hours cycle 2,3 m³/h operating flow rate) 1 µg/ m³ ( 24 hours cycle 2,3 m³/h operating flow rate) n° 36 filter cartridges (or 72 on demand) Programmable in the range 0.8 – 2.5 m³/h PM10 cut size (LVS-PM10 model in compliance with EN 1234-1 standard, working at 2.3 m³/h) PM2.5 cut size (LVS-PM2.5 model in compliance with EN 14907 standard, working at 2.3 m³/h)_Other inlet available 40 kPa at 2.3 m³/h ± 1% of the...

Open the catalog to page 2

All TECORA catalogs and technical brochures

  1. EMITEST ISO

    4 Pages

  2. Sentinel PM

    1 Pages

  3. Echo PM

    1 Pages

  4. Charlie HV

    1 Pages

  5. Echo Hi Vol

    2 Pages

  6. Easy Gas

    1 Pages

  7. Isostack G4

    4 Pages

  8. Isofrost 2

    1 Pages

  9. Igloo

    1 Pages

  10. Delta MK2

    3 Pages

  11. Bravo Series

    1 Pages

  12. Flowcal Air

    3 Pages