Overview: This document provides detailed specifications for the MCR-4V and MCR-4TC data loggers, focusing on voltage and temperature measurements respectively. It includes information on measurement channels, input methods, accuracy, recording intervals, and power requirements.
MCR-4V Specifications:
  • Measurement Channels: 4 channels for voltage.
  • Input Method: Scanning method with differential input, each channel isolated.
  • Measurement Range: ±300 mV to ±24 V, with an auto range option.
  • Accuracy: Varies with range, e.g., ±(0.3% + 0.06 mV) for ±300 mV.
  • Recording Interval: Ranges from 2 ms to 60 min, depending on settings.
  • Logging Capacity: Up to 480,000 readings for 1 channel.
  • Power: Operates on AA batteries or USB power.
MCR-4TC Specifications:
  • Measurement Channels: 4 channels for temperature.
  • Compatible Sensors: Thermocouples (Types K, J, T, S, R).
  • Measurement Range: Varies by sensor type, e.g., Type K: -270 to 1370 °C.
  • Accuracy: ±(0.5 °C + 0.3% reading) for Types K, J, T at -100°C or above.
  • Recording Interval: Ranges from 100 ms to 60 min.
  • Logging Capacity: Up to 960,000 readings for 1 channel.
  • Power: Similar to MCR-4V, with battery life varying by recording settings.
Common Features:
  • Group Recording: Up to 16 channels can be recorded simultaneously.
  • Communication Interfaces: USB and SD memory card support for data export.
  • Isolation: Electrical isolation between channels and USB.
  • Operating Environment: Temperature 0 to 50 °C, humidity 90% RH or less.
Additional Features:
  • Software Compatibility: Compatible with Windows OS, requires .NET Framework 4.
  • Graphical Analysis: Data can be viewed and analyzed in graph form, with options for CSV export.
  • Application Examples: Suitable for various industrial and laboratory applications, including temperature management and voltage measurement in electrical circuits.
See more

Catalog excerpts

MCR-4V/4TC-1

MCR-4TC Specifications Measurement Channels Measurement Channels Input Method Compatible Sensors Measurement Units Measurement Range Input Impedance Measurement Resolution Recording Interval Recording Mode Group Recording Communication Interfaces External Memory Input Terminal Operating Environment Scanning Method, Differential Input, Each Channel Isolated Type K : -270 to 1370 °C Type J : -210 to 1200 °C Type T : -270 to 400 °C Thermocouple Measurement ( Sensor inaccuracies not included ) Type K, J, T: ±( 0.5 °C + 0.3 % reading ) at -100°C or above Type S, R : ±( 1.5 °C + 0.3 % reading ) at 100°C or above Cold Junction Compensation ±0.5°C at 10 to 40 °C ±0.8°C other temperatures within the operating environment of the logger When recording 1 channel : up to 960,000 readings/ch When recording 2 channels : up to 480,000 readings/ch When recording 3 channels : up to 320,000 readings/ch When recording 4 channels : up to 240,000 readings/ch Endless ( Overwrite oldest data in the current recording session when capacity is full ) or One Time ( Stop recording when capacity is full ) Up to 4 units ( 16 channels ) can be recorded simultaneously. Coupling of MCR-4TC and MCR-4V is possible. (*3) Measurements, Trend Graph, Battery Level, etc. SD Memory Card, SDHC Memory Card (*4) For Manual or Automatic Data Export AA Alkaline Battery x 2, AA Ni-MH Battery x 2, AC Adaptor AD-05A2 or AD-05C2, USB Bus Power 5V 250mA For 4-Channel Measurement ( AA Alkaline Battery) Screwless Terminals Compatible Wires Single Wire : 0 0.32 to 0 0.65 mm ( AWG 28 - 22 ) Twisted Wire : 0.08 to 0.32 : ( AWG 28 - 22 ), 0 0.12 mm or more in diameter Stripping Length : 9 to 10 mm CH1, CH2, CH3, CH4, and USB are isolated. CH1-CH4 Maximum Applied Voltage : ±50 V Electrical Isolation Resistance : 50 MQ or more ( DC±250 V ) Temperature: 0 to 50 °C Humidity: 90 %RH or less ( no condensation ) AA Alkaline Battery x 2, USB Mini-B Cable US-15C, Software CD-ROM, Card Slot Cover, User's Manual Set ( Warranty Included ) *1: When "Auto" is selected, measurement range will be automatically changed according to the voltage being measured. *2: If the logging capacity is not filled at the end of one recording session, the logger can record up to 30 times. *3: Group Recording may not be started depending on the recording or measurement interval specifications of the connected Master unit. *4: Please check the T&D Website for information on memory cards whose operation has been confirmed. *5: Battery life varies depending upon multiple factors including measurement interval and 50-60 Hz filter setting. All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life. *1: MCR-4TC has superior noise filter, but the measurement may sometimes fluctuate due to strong noise. Especially when the recording interval is set to 200 ms or less, the filtering becomes weaker and hence the fluctuation may become greater. *2: If the logging capacity is not filled at the end of one recording session, the logger can record up to 30 times. *3: Group Recording may not be started depending on the recording or measurement interval specifications of the connected Master unit. *4: Please check the T&D Website for information on memory cards whose operation has been confirmed. *5: Battery life varies depending upon multiple factors including ambient temperature, recording interval, number of measurement channels, and frequency of data export to a memory card. All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life. Software (MCR for Windows) Compatible OS (*1) *1: For installation, it is necessary to have Administrator ( Computer Administrator ) rights. *2: We recommend using an operating system in the same language as the display language. Operation in different languages is not guaranteed. • Colors in the photos in this catalog may be different from real product colors. • The specification and designs of the products in this catalog are true as of Aug 2017. Specifications are subject to change without notice. • Microsoft and Windows are registered trademarks of Microsoft Corporation USA and other countries. • Company names and product names are trademarks or registered trademarks of each company. T&D Corporation 817-1 Shimadachi, Matsumoto, Nagano 390-0852, Japan Please send your inquiries to: E-mail : [email protected] Facsimile : (+81) 263-40-3152

 Open the catalog to page 1
MCR-4V/4TC-2

MCR-4V Fe Preheat Terminal I I I 24V Sensor Output Preheat Function for Saving Battery Power Sensor Power Input Terminal Easy Connection of MCR-4V and MCR-4TC for Synchronous Measurement of Voltage and Temperature Trend Graph for Real-Time Data Check Quick and Precise Measurements Recording Interval as short as 2 msec / Resolution of 10pV Number of channels determines shortest possible recording interval. Two recording methods: instantaneous value or average value. Easy Touch Panel Operation Scale and Unit Conversion for Recording and Viewing Large Capacity Internal Memory ' MCR-4V: Data logging...

 Open the catalog to page 2

All T&D Corporation catalogs and technical brochures

  1. TR7A

    1  Page

  2. RTR500B

    1  Page

  3. TR4A

    1  Page

  4. TR7A2 Series

    8  Pages

  5. T&D TR7A

    4  Pages