Lyssy L80-5000

Lyssy L80-5000

Lyssy L80-5000

Product catalog summary
Overview
The Lyssy L80-5000 is an advanced water vapour permeability tester, part of the well-established L80 series. It is designed for ease of use, reliability, and minimal maintenance, making it suitable for various laboratory environments.
Key Features
  • Complete traceability in test documentation, data logging, and error reports.
  • Automatic equilibrium detection and temperature control.
  • Easy-to-use test sample holders that do not require grease for sealing.
  • Built-in printer for easy test documentation.
  • Low maintenance requirements.
Testing Capabilities
  • Fast and accurate testing with a broad range of permeability measurements (0.03-10,000 g/m2/day).
  • High repeatability and accuracy, with a standard deviation lower than ±5% and reproducibility tolerance as low as 1.5%.
  • Automatic detection of equilibrium and quick alternation between low and high permeability measurements.
Technical Specifications
  • Dimensions: 480H x 400W x 470D mm; Weight: Approximately 26kg.
  • Measuring temperature range: 5-70°C, with practical range 30-70°C using built-in temperature control.
  • Humidity range: 10% or 35% RH in the measuring chamber.
  • Sample requirements: Low permeability samples - 50cm2; High permeability samples - 2.5cm2; Sample thickness up to 12mm.
  • Data logging capacity: 1 record with 1427 measurements or 109 records with 10 measurements each.
Compliance and Standards
  • Conforms to ASTM E398-03, ISO/CD 15106, JIS K 7129, TAPPI T523 om-82, NF H00-044 standards.
Contact Information
  • Systech Illinois has offices in the U.S., UK, Thailand, and China, providing global support and sales services.
See more

Catalog excerpts

Lyssy L80-5000-1

Water Vapour Permeability Tester Easy and reliable testing for water vapour permeability Simplicity in operation due to the high degree of automation - the quality of tests performed is less operator dependent Easy to use operator interface Fast and accurate test result Extremely broad testing range, covering low and high permeability High repeatability of testing results Easy set-up of test parameters and sample data The preparation of a sample for the L80-5000 is accomplished in minutes. No grease or glue is required for a tight seal around the sample in the chamber. The test sample is affixed to the self-adhesive sample card, which is inserted into the L80-5000 test chamber, separating the upper and lower chamber. The Lyssy L80-5000 is able to measure high permeability materials using special sample reduction cards that decrease the surface area of the sample. The L80-5000 is the latest generation of the proven L80 series, which has been used successfully for decades in many labs. With the L80-5000, the testing of water vapour permeability becomes very simple. The L80-5000 is easy to operate, and its reliable testing principle and low degree of maintenance have demonstrated to users worldwide the superiority of the Lyssy testing method. The L80-5000 is easy to set up and operate through the built-in alphanumeric keyboard and digital display. With a built in printer the L80-5000 also enables easy test documentation. Complete traceability in test documentation, data logging and error reports Automatic equilibrium detection Automatic temperature control Easy to use test sample holders - no grease needed for sealing Built-in printer Low degree of maintenance This reduction of the surface prevents the system from getting saturated, and therefore it becomes possible to dry down the upper chamber of the instrument and obtain a permeability measurement.

 Open the catalog to page 1
Lyssy L80-5000-2

Fast, accurate and versatile Technical Specifications General description Dimensions Weight Measuring range Sensor life Voltage Conforms to the following standards 480H x 400W x 470D mm Approximately 26kg 0.03 - 10,000 g/m2/day as standard More than 5 years under normal conditions 230 VAC or 115/100 VAC ASTM E398-03, ISO/CD 15106, JIS K 7129, TAPPI T523 om-82, NF H00-044 Measuring Measuring temperature range Humidity Sample Requirements Measuring area Sample thickness Minimum sample size Data logging Memory 5 - 70°C. Practical range 30-70°C, with built-in temperature control. For measuring below...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.